Initial experience, feasibility and safety of permanent left bundle branch pacing: results from a prospective single-centre study

被引:6
|
作者
Rademakers, L. M. [1 ]
van den Broek, J. L. P. M. [1 ]
't Hof, M. Op [1 ]
Bracke, F. A. [1 ]
机构
[1] Catharina Hosp, Dept Cardiol, Eindhoven, Netherlands
关键词
Left bundle branch pacing; Physiologic pacing; Cardiac resynchronisation therapy; Bradycardia pacing; CARDIAC RESYNCHRONIZATION THERAPY; DUAL-CHAMBER; ACTIVATION; ATRIAL; TERM;
D O I
10.1007/s12471-021-01648-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Left bundle branch (LBB) pacing is a novel pacing technique which may serve as an alternative to both right ventricular pacing for symptomatic bradycardia and cardiac resynchronisation therapy (CRT). A substantial amount of data is reported by relatively few, highly experienced centres. This study describes the first experience of LBB pacing in a high-volume device centre. Methods Success rates (i.e. the ability to achieve LBB pacing), electrophysiological parameters and complications at implant and up to 6 months of follow-up were prospectively assessed in 100 consecutive patients referred for various pacing indications. Results The mean age was 71 +/- 11 years and 65% were male. Primary pacing indication was atrioventricular (AV) block in 40%, CRT in 42%, and sinus node dysfunction or refractory atrial fibrillation prior to AV node ablation in 9% each. Baseline left ventricular ejection fraction was < 50% in 57% of patients, mean baseline QRS duration 145 +/- 34 ms. Overall LBB pacing was successful in 83 of 100 (83%) patients but tended to be lower in patients with CRT pacing indication (69%, p = ns). Mean left ventricular activation time (LVAT) during LBB pacing was 81 ms and paced QRS duration was 120 +/- 19 ms. LBB capture threshold and R-wave sense at implant was 0.74 +/- 0.4 mV at 0.4 ms and 11.9 +/- 5.9 V and remained stable at 6-month follow-up. No complications occurred during implant or follow-up. Conclusion LBB pacing for bradycardia pacing and resynchronisation therapy can be easily adopted by experienced implanters, with favourable success rates and safety profile.
引用
收藏
页码:258 / 266
页数:9
相关论文
共 50 条
  • [1] Initial experience, feasibility and safety of permanent left bundle branch pacing: results from a prospective single-centre study
    L. M. Rademakers
    J. L. P. M. van den Broek
    M. Op ’t Hof
    F. A. Bracke
    [J]. Netherlands Heart Journal, 2022, 30 : 258 - 266
  • [2] Initial Experience, Safety, and Feasibility of Left Bundle Branch Area Pacing A Multicenter Prospective Study
    Padala, Santosh K.
    Master, Vivak M.
    Terricabras, Maria
    Chiocchini, Andrea
    Garg, Aatish
    Kron, Jordana
    Shepard, Richard
    Kalahasty, Gautham
    Azizi, Zahara
    Tsang, Bernice
    Khaykin, Yaariv
    Pantano, Alfredo
    Koneru, Jayanthi N.
    Ellenbogen, Kenneth A.
    Verma, Atul
    [J]. JACC-CLINICAL ELECTROPHYSIOLOGY, 2020, 6 (14) : 1773 - 1782
  • [3] Left bundle branch pacing: 2 year single-centre experience
    Marcantoni, L.
    Pastore, G.
    Baracca, E.
    Bartolomei, M.
    Centioni, M.
    Andreaggi, S.
    Pellegrini, N.
    Galuppi, E.
    Rigatelli, G.
    Roncon, L.
    Zanon, F.
    [J]. EUROPEAN HEART JOURNAL, 2021, 42 : 685 - 685
  • [4] OUR LEFT BUNDLE BRANCH PACING EXPERIENCE: A SINGLE CENTRE STUDY
    Walsh, R.
    Kehoe, F.
    Frohlich, S.
    Murphy, N.
    Lyne, J.
    [J]. HEART, 2021, 107 : A25 - A25
  • [5] Feasibility and safety of His bundle pacing versus left bundle branch area pacing: insights from a 2-year single centre experience
    De Pooter, J.
    Calle, S.
    Timmermans, F.
    Van Heuverswyn, F.
    [J]. ACTA CARDIOLOGICA, 2020, 75 (06) : 573 - 574
  • [6] Feasibility and safety of zero-fluoroscopy left bundle branch pacing: An initial experience
    Ramos-Maqueda, Javier
    Melero-Polo, Jorge
    Montilla-Padilla, Isabel
    Ruiz-Arroyo, Jose Ramon
    Cabrera-Ramos, Mercedes
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (02) : 429 - 437
  • [7] The feasibility and safety of left bundle branch pacing vs. right ventricular pacing after mid-long-term follow-up: a single-centre experience
    Chen, Xueying
    Jin, Qinchun
    Bai, Jin
    Wang, Wei
    Qin, Shengmei
    Wang, Jingfeng
    Liang, Yixiu
    Su, Yangang
    Ge, Junbo
    [J]. EUROPACE, 2020, 22 : 36 - 44
  • [8] His bundle pacing and left bundle branch area pacing: Feasibility and safety
    Saleiro, Carolina
    Sousa, Pedro A.
    Nogueira, Catarina
    Mota, Lidia
    Almeida, Claudia
    Braganca, Gisela
    Paisana, Francisco
    [J]. REVISTA PORTUGUESA DE CARDIOLOGIA, 2023, 42 (08) : 683 - 691
  • [9] Mid-term feasibility, safety and outcomes of left bundle branch pacing–single center experience
    Shunmuga Sundaram Ponnusamy
    Giridhar Muthu
    Mahesh Kumar
    Dasarath Bopanna
    Vijesh Anand
    Surya Kumar
    [J]. Journal of Interventional Cardiac Electrophysiology, 2021, 60 : 337 - 346
  • [10] Feasibility and safety of left bundle branch area pacing in cardiac amyloidosis. A single center experience
    Pham-Trung, Chinh
    Veloza-Urrea, Darwin
    Segura-Dominguez, Melodie
    Rojas, Yuleisy De la Rosa
    Aguilera-Agudo, Cristina
    Garcia-Izquierdo, Eusebio Alejandro
    Garcia-Rodriguez, Daniel
    Jimenez-Sanchez, Diego
    Lorente-Ros, Alvaro
    Mingo-Santos, Susana
    Gonzalez-Lopez, Esther
    Dominguez, Fernando
    Garcia-Pavia, Pablo
    Toquero-Ramos, Jorge
    Fernandez-Lozano, Ignacio
    Castro-Urda, Victor
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2024, 47 (01): : 149 - 155