Percutaneous electrode placement for spinal cord stimulation in a patient with spinal fusion: a technical report

被引:2
|
作者
Park, Sung-Chun [2 ]
Kim, Kyung-Hoon [1 ,2 ]
机构
[1] Pusan Natl Univ, Yangsan Hosp, Pain Clin, Yangsan 626770, Kyungsangnamdo, South Korea
[2] Pusan Natl Univ, Sch Med, Dept Anesthesia & Pain Med, Pusan 609735, South Korea
关键词
Electrical analgesia; Electric stimulation; Failed back surgery syndrome; Spinal cord injury; Spinal fusion; DESIGN;
D O I
10.1007/s00540-011-1311-4
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A spinal cord stimulation (SCS) trial was attempted to alleviate left knee pain in a patient with spinal fusion from T12 to L4. Good paresthesia coverage for the knee pain was attained with SCS. However, while removing the needle used for electrode placement, the needle became fixed in the bony supplementary tissue. Moreover, while attempting to remove the needle using Kelly forceps, the hub of the needle became blocked. Without the hub, we had no choice but to use a pneumatic drill for removing the needle. Accordingly, the supplementary bone tissue was drilled under real-time imaging, using a pneumatic drill with a 3.2-mm drill bit, and another epidural needle was inserted through the hole. We consider that, in patients with spinal fusion, making a borehole with a pneumatic drill for introducing the epidural needle for percutaneous SCS electrode placement may be advisable in order to avoid the above-mentioned difficulties.