A five-station hip joint simulator

被引:16
|
作者
Smith, SL [1 ]
Unsworth, A [1 ]
机构
[1] Univ Durham, Sch Engn, Ctr Biomed Engn, Sci Labs, Durham DH1 3LE, England
关键词
wear; hip simulators; zirconia ceramic; ultra-high molecular weight polyethylene; (UHMWPE);
D O I
10.1243/0954411011533535
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A five-station hip joint wear simulator was designed and built which featured simplified motion and loading. An elliptical wear path was produced using approximately sinusoidal motion in the flexion/extension and internal/external rotation axes and the dynamic loading approximated to a square wave. Five 28 mm diameter zirconia femoral heads articulated against ultra-high molecular weight polyethylene acetabular cups in 25 per cent bovine serum for 5 x 10(6) cycles. Gravimetric wear measurement was used with moisture absorption compensation using a dynamically loaded soak control. With motion of physiological magnitude, the mean acetabular cup wear rate was 52.2 mm(3)/10(6) cycles which is comparable with a number of clinical studies.
引用
收藏
页码:61 / 64
页数:4