A Fully Implantable Miniaturized Liquid Crystal Polymer (LCP)-Based Spinal Cord Stimulator for Pain Control

被引:10
|
作者
Yun, Seunghyeon [1 ]
Koh, Chin Su [2 ]
Seo, Jungmin [1 ]
Shim, Shinyong [1 ]
Park, Minkyung [2 ]
Jung, Hyun Ho [2 ]
Eom, Kyungsik [3 ]
Chang, Jin Woo [2 ]
Kim, Sung June [1 ]
机构
[1] Seoul Natl Univ, Coll Engn, Dept Elect & Comp Engn, Seoul 08826, South Korea
[2] Yonsei Univ, Coll Med, Dept Neurosurg, Seoul 03722, South Korea
[3] Pusan Natl Univ, Coll Engn, Dept Elect Engn, Busan 46241, South Korea
基金
新加坡国家研究基金会;
关键词
spinal cord stimulation; liquid crystal polymer; implantable device; pain control; CHRONIC BACK; MODEL; MECHANISMS; MRI;
D O I
10.3390/s22020501
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Spinal cord stimulation is a therapy to treat the severe neuropathic pain by suppressing the pain signal via electrical stimulation of the spinal cord. The conventional metal packaged and battery-operated implantable pulse generator (IPG) produces electrical pulses to stimulate the spinal cord. Despite its stable operation after implantation, the implantation site is limited due to its bulky size and heavy weight. Wireless communications including wireless power charging is also restricted, which is mainly attributed to the electromagnetic shielding of the metal package. To overcome these limitations, here, we developed a fully implantable miniaturized spinal cord stimulator based on a biocompatible liquid crystal polymer (LCP). The fabrication of electrode arrays in the LCP substrate and monolithically encapsulating the circuitries using LCP packaging reduces the weight (0.4 g) and the size (the width, length, and thickness are 25.3, 9.3, and 1.9 mm, respectively). An inductive link was utilized to wirelessly transfer the power and the data to implanted circuitries to generate the stimulus pulse. Prior to implantation of the device, operation of the pulse generator was evaluated, and characteristics of stimulation electrode such as an electrochemical impedance spectroscopy (EIS) were measured. The LCP-based spinal cord stimulator was implanted into the spared nerve injury rat model. The degree of pain suppression upon spinal cord stimulation was assessed via the Von Frey test where the mechanical stimulation threshold was evaluated by monitoring the paw withdrawal responses. With no spinal cord stimulation, the mechanical stimulation threshold was observed as 1.47 +/- 0.623 g, whereas the stimulation threshold was increased to 12.7 +/- 4.00 g after spinal cord stimulation, confirming the efficacy of pain suppression via electrical stimulation of the spinal cord. This LCP-based spinal cord stimulator opens new avenues for the development of a miniaturized but still effective spinal cord stimulator.
引用
收藏
页数:12
相关论文
共 32 条
  • [31] Fully Wireless and Self-Powered Ocean Wave Observation System Empowered by the Friction-Driven Polymer Network Liquid Crystal-Based Smart Reflector
    Wang, Jiaqi
    Chen, Xingwen
    Xie, Shixing
    Bao, Guowei
    Wu, Fan
    Meng, Cuiling
    ACS APPLIED MATERIALS & INTERFACES, 2023, 15 (46) : 53476 - 53487
  • [32] Wideband Microstrip-to-Microstrip Vialess Vertical Transition Based on Multilayer Liquid Crystal Polymer Technology; [基于多层LCP技术宽带微带到微带无孔垂直过渡结构的设计]
    Liu W.
    Guan D.
    Huang Q.
    Chen L.
    Zhang M.
    Journal of Shanghai Jiaotong University (Science), 2025, 30 (2) : 220 - 226