Implementation of an early rule-out pathway for myocardial infarction using a high-sensitivity cardiac troponin T assay

被引:6
|
作者
Sandeman, Dennis [1 ]
Syed, Maaz B. J. [2 ]
Kimenai, Dorien M. [3 ]
Lee, Kuan Ken [3 ]
Anand, Atul [3 ]
Joshi, Shruti S. [3 ]
Dinnel, Lorraine [1 ]
Wenham, Philip R. [4 ]
Campbell, Ken [4 ]
Jarvie, Mary [4 ]
Galloway, Donna [4 ]
Anderson, Mhairi [4 ]
Roy, Bappa [5 ]
Andrews, Jack P. M. [3 ]
Strachan, Fiona E. [3 ]
Ferry, Amy, V [3 ]
Chapman, Andrew R. [3 ]
Elsby, Sarah [6 ]
Francis, Mark [1 ]
Cargill, Robert [1 ]
Shah, Anoop S., V [7 ]
Mills, Nicholas L. [3 ,8 ]
机构
[1] NHS Fife, Cardiol Dept, Kirkcaldy, Scotland
[2] Univ Edinburgh, Div Clin & Surg Sci, Edinburgh, Midlothian, Scotland
[3] Univ Edinburgh, British Heart Fdn, Ctr Cardiovasc Sci, Edinburgh, Midlothian, Scotland
[4] NHS Fife, Biochem Dept, Kirkcaldy, Scotland
[5] NHS Fife, Emergency Dept, Kirkcaldy, Scotland
[6] NHS Fife, Informat Serv, Dunfermline, Fife, Scotland
[7] London Sch Hyg & Trop Med, London, England
[8] Univ Edinburgh, Usher Inst, Edinburgh, Midlothian, Scotland
来源
OPEN HEART | 2021年 / 8卷 / 02期
基金
英国医学研究理事会; 英国工程与自然科学研究理事会; 英国惠康基金; 英国经济与社会研究理事会;
关键词
EARLY-DIAGNOSIS; ALGORITHM;
D O I
10.1136/openhrt-2021-001769
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives Patients with suspected acute coronary syndrome and high-sensitivity cardiac troponin (hs-cTn) concentrations below the limit of detection at presentation are low risk. We aim to determine whether implementing this approach facilitates the safe early discharge of patients. Methods In a prospective single-centre cohort study, consecutive patients with suspected acute coronary syndrome were included before (standard care) and after (intervention) implementation of an early rule-out pathway. During standard care, myocardial infarction was ruled out if hs-cTnT concentrations were <99th centile (14 ng/L) at presentation and at 6-12 hours after symptom onset. In the intervention, patients were ruled out if hs-cTnT concentrations were <5 ng/L at presentation and symptoms present for >= 3 hours or were >= 5 ng/L and unchanged within the reference range at 3 hours. We compared duration of stay (efficacy) and all-cause death at 1 year (safety) before and after implementation. Results We included 10 315 consecutive patients (64 +/- 16 years, 46% women) with 6642 (64%) and 3673 (36%) in the standard care and intervention groups, respectively. Duration of stay was reduced from 534 (IQR, 220-2279) to 390 (IQR, 218-1910) min (p<0.001) after implementation. At 1 year, all-cause death occurred in 10.9% (721 of 6642) and 10.4% (381 of 3673) of patients in the standard care group (referent) and intervention group, respectively (adjusted OR 1.02, 95% CI 0.88 to 1.18). Conclusion In patients with suspected acute coronary syndrome, implementing an early rule-out pathway using hs-cTnT concentrations <5 ng/L at presentation reduced the duration of stay in hospital without compromising safety.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Rule-out myocardial infarction in the high-sensitivity cardiac troponin era
    Del Carlo, Carlos H.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2019, 288 : 17 - 18
  • [2] High-sensitivity troponin assays and the early rule-out of acute myocardial infarction
    Shah, Anoop S. V.
    Newby, David E.
    Mills, Nicholas L.
    HEART, 2013, 99 (21) : 1549 - 1550
  • [3] Evaluating Rapid Rule-out of Acute Myocardial Infarction Using a High-Sensitivity Cardiac Troponin I Assay at Presentation
    Greenslade, Jaimi
    Cho, Elizabeth
    Van Hise, Christopher
    Hawkins, Tracey
    Parsonage, William
    Ungerer, Jacobus
    Tate, Jillian
    Pretorius, Carel
    Than, Martin
    Cullen, Louise
    CLINICAL CHEMISTRY, 2018, 64 (05) : 820 - 829
  • [4] One-Hour Rule-out and Rule-in of Acute Myocardial Infarction Using High-Sensitivity Cardiac Troponin T
    Reichlin, Tobias
    Schindler, Christian
    Drexler, Beatrice
    Twerenbold, Raphael
    Reiter, Miriam
    Zellweger, Christa
    Moehring, Berit
    Ziller, Ronny
    Hoeller, Rebeca
    Gimenez, Maria Rubini
    Haaf, Philip
    Potocki, Mihael
    Wildi, Karin
    Balmelli, Cathrin
    Freese, Michael
    Stelzig, Claudia
    Freidank, Heike
    Osswald, Stefan
    Mueller, Christian
    ARCHIVES OF INTERNAL MEDICINE, 2012, 172 (16) : 1211 - 1218
  • [5] High-sensitivity troponin T for early rule-out of myocardial infarction in recent onset chest pain
    Aldous, Sally
    Pemberton, Chris
    Richards, A. Mark
    Troughton, Richard
    Than, Martin
    EMERGENCY MEDICINE JOURNAL, 2012, 29 (10) : 805 - 810
  • [6] Prospective validation of a 1-hour algorithm to rule-out and rule-in acute myocardial infarction using a high-sensitivity cardiac troponin T assay
    Reichlin, Tobias
    Twerenbold, Raphael
    Wildi, Karin
    Rubini Gimenez, Maria
    Bergsma, Nathalie
    Haaf, Philip
    Druey, Sophie
    Puelacher, Christian
    Moehring, Berit
    Freese, Michael
    Stelzig, Claudia
    Krivoshei, Lian
    Hillinger, Petra
    Jaeger, Cedric
    Herrmann, Thomas
    Kreutzinger, Philip
    Radosavac, Milos
    Weidmann, Zoraida Moreno
    Pershyna, Kateryna
    Honegger, Ursina
    Wagener, Max
    Vuillomenet, Thierry
    Campodarve, Isabel
    Bingisser, Roland
    Miro, Oscar
    Rentsch, Katharina
    Bassetti, Stefano
    Osswald, Stefan
    Mueller, Christian
    CANADIAN MEDICAL ASSOCIATION JOURNAL, 2015, 187 (08) : E243 - E252
  • [7] One-hour rule-in and rule-out of acute myocardial infarction using high-sensitivity cardiac troponin
    Jaeger, Cedric
    Wildi, Karin
    Twerenbold, Raphael
    Reichlin, Tobias
    Gimenez, Maria Rubini
    Neuhaus, Jean-Daniel
    Grimm, Karin
    Boeddinghaus, Jasper
    Hillinger, Petra
    Nestelberger, Thomas
    Singeisen, Helene
    Gugala, Mathias
    Pretre, Gil
    Puelacher, Christian
    Wagener, Max
    Honegger, Ursina
    Schumacher, Carmela
    Weidmann, Zoraida Moreno
    Kreutzinger, Philipp
    Krivoshei, Lian
    Freese, Michael
    Stelzig, Claudia
    Dietsche, Sebastian
    Ernst, Susanne
    Rentsch, Katharina
    Osswald, Stefan
    Mueller, Christian
    AMERICAN HEART JOURNAL, 2016, 171 (01) : 92 - 102
  • [8] One-hour rule-out and rule-in of acute myocardial infarction using Siemens high-sensitivity cardiac troponin T
    Twerenbold, R.
    Reichlin, T.
    Rubini-Gimenez, M.
    Mueller, M.
    Wildi, K.
    Haaf, P.
    Mueller, C.
    EUROPEAN HEART JOURNAL, 2013, 34 : 71 - 71
  • [9] One-hour rule-out and rule-in of acute myocardial infarction using a novel high-sensitivity cardiac troponin I assay
    Twerenbold, R.
    Boeddinghaus, J.
    Nestelberger, T.
    Rubini Gimenez, M.
    Puelacher, C.
    Miro, O.
    Martin-Sanchez, F. J.
    Morawiec, B.
    Osswald, S.
    Reichlin, T.
    Mueller, C.
    EUROPEAN HEART JOURNAL, 2017, 38 : 1006 - 1006
  • [10] High-sensitivity cardiac troponin and the early rule out of myocardial infarction: time for action
    Chapman, Andrew R.
    Mills, Nicholas L.
    HEART, 2020, 106 (13) : 955 - 957