A high-resolution analyser for the measurement of ammonium in oligotrophic seawater

被引:25
|
作者
Bey, Samer K. Abi Kaed [1 ]
Connelly, Douglas P. [1 ]
Legiret, Francois-Eric [1 ]
Harris, Andy J. K. [1 ]
Mowlem, Matthew C. [1 ]
机构
[1] Univ Southampton Waterfront Campus, Natl Oceanog Ctr, Southampton SO14 3ZH, Hants, England
基金
英国自然环境研究理事会;
关键词
Ammonium; Shipboard; Analyser; Fluorometric; Resolution; ESTUARINE WATERS; IN-SITU; FLUOROMETRIC-DETERMINATION; FLOW-ANALYSIS; FLUORESCENCE; REAGENT; SAMPLES; FRESH; CELL;
D O I
10.1007/s10236-011-0469-5
中图分类号
P7 [海洋学];
学科分类号
0707 ;
摘要
In this work, we describe a high-resolution fluorometric shipboard analyser and an improved method to determine NH (4) (+) in oligotrophic seawater. The limit of detection is < 5 nM, calculated with 95% confidence level using the weighted regression line applied to the standard addition method using real samples prepared with low nutrient seawater from the Atlantic. The results are summarised and cross-compared with spiked artificial seawater (ASW) and spiked Milli-Q water samples. The analyser has a precision of +/- 1-4% with a high performance over a wide range from 5 nM to 25 mu M. The methodology of NH (4) (+) analysis is based on the fluorescent product formed between o-pthaldialdehyde and NH (4) (+) in the presence of sulfite. Due to the high resolution of the developed system, we were able to study in depth the sensitivity of the method to salinity, amines, amino acids and potential interferences from particles/algae. The method was found to be sensitive to salinity variations, reducing the signal by up to 85% at 5 nM; this effect decreased at higher concentrations of ammonium. It was noted that the interference from amines at low concentrations was negligible; however, at either high amino acid or high amine concentrations, the signal was depressed. To test for the effect of particles on the system, the system was tested with samples containing phytoplankton (Dunaliella primolecta) cells at different concentrations prepared with ASW to simulate the effect of a phytoplankton bloom. This experiment assessed the potential impact of both particles and other potential fluorescence interferences from cells and/or ammonium leaching from cells. This experiment showed that a phytoplankton bloom could potentially have an impact of up to 12% on the signal of interest. Thus, we propose that this method is suitable for oligotrophic environments rather than coastal and eutrophic environments. The reagent was found to be stable for 17 days and standards of 1 mu M were stable for 6 days under laboratory conditions. The developed analyser was successfully demonstrated in the North Atlantic Ocean, in an area of oligotrophic, low NH (4) (+) oceanic waters.
引用
收藏
页码:1555 / 1565
页数:11
相关论文
共 50 条
  • [31] High-Resolution Measurement of Data Center Microbursts
    Zhang, Qiao
    Liu, Vincent
    Zeng, Hongyi
    Krishnamurthy, Arvind
    PROCEEDINGS OF THE 2017 INTERNET MEASUREMENT CONFERENCE (IMC'17), 2017, : 78 - 85
  • [32] SIMPLE DISPERSION MEASUREMENT TECHNIQUE WITH HIGH-RESOLUTION
    CHRISTENSEN, B
    MARK, J
    JACOBSEN, G
    BODTKER, E
    ELECTRONICS LETTERS, 1993, 29 (01) : 132 - 134
  • [33] A Portable High-Resolution Surface Measurement Device
    Ihlefeld, Curtis M.
    Burns, Bradley M.
    Youngquist, Robert C.
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2013, 62 (01) : 205 - 209
  • [34] HIGH-RESOLUTION DETERMINATION OF THE PH OF SEAWATER WITH A FLOW-THROUGH SYSTEM
    FUHRMANN, R
    ZIRINO, A
    DEEP-SEA RESEARCH PART A-OCEANOGRAPHIC RESEARCH PAPERS, 1988, 35 (02): : 197 - 208
  • [35] High resolution retarding field analyser
    Johnson, SD
    El-Gomati, MM
    Mahadevan, J
    Enloe, L
    IVMC 2000: PROCEEDINGS OF THE 14TH INTERNATIONAL VACUUM MICROELECTRONICES CONFERENCE, 2001, : 11 - 12
  • [36] High-resolution measurement by a high-Tc superconductor sampler
    Hidaka, M.
    Satoh, T.
    Koike, M.
    Tahara, S.
    IEEE Transactions on Applied Superconductivity, 1999, 9 (2 III): : 4081 - 4086
  • [37] High-resolution measurement by a high-Tc superconductor sampler
    Hidaka, M
    Satoh, T
    Koike, M
    Tahara, S
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) : 4081 - 4086
  • [38] AMMONIUM UPTAKE BY PHYTOPLANKTON CELLS ON A FILTER - A NEW HIGH-RESOLUTION TECHNIQUE
    PARSLOW, JS
    HARRISON, PJ
    THOMPSON, PA
    MARINE ECOLOGY PROGRESS SERIES, 1985, 25 (02) : 121 - 129
  • [39] HIGH-RESOLUTION CARS MEASUREMENT OF RAMAN LINEWIDTHS OF DEUTERIUM
    RUSSELL, DA
    ROH, WB
    JOURNAL OF MOLECULAR SPECTROSCOPY, 1987, 124 (01) : 240 - 242
  • [40] High-resolution RCS measurement using MUSIC algorithm
    Yamazaki, H
    Murasaki, T
    Takahashi, R
    ELECTRONICS AND COMMUNICATIONS IN JAPAN PART I-COMMUNICATIONS, 2006, 89 (06): : 23 - 31