The quality of intrapartum uterine activity monitoring

被引:31
|
作者
Bakker, Petra C. A. M. [1 ]
Zikkenheimer, Marjoes [1 ]
van Geijn, Herman P. [1 ]
机构
[1] Vrije Univ Amsterdam, Med Ctr, Dept Obstet & Gynecol, NL-1007 MB Amsterdam, Netherlands
关键词
cardiotocography; quality; uterine activity;
D O I
10.1515/JPM.2008.027
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Aims: To determine the quality of intrapartum uterine activity (UA) monitoring in daily practice during the first and second stage of labor. The total duration of inadequate UA monitoring is quantified in relation to the technique applied: external or internal. Methods: One hundred and ninety-two UA recordings, collected from 1 April 2006 untill 1 October 2006 from consecutive deliveries at the Vrije Universiteit Medical Center in Amsterdam, were analyzed. Included recordings were from singleton, spontaneous, vaginal deliveries. The last two hours of the first stage and the complete second stage had to be monitored. Internal and external UA recordings were judged by their quality: adequate, a recognizable and reliable UA pattern during the complete registration, or inadequate. Recordings labeled as inadequate were divided into two groups: absence of UA recording (non recognizable) or inadequate calibration (unreliable). Statistical analyses were performed with the Mann-Whitney U-test and the Wilcoxon signed ranks test. Results: The percentages of adequate UA recordings in the first stage of labor were much higher for the internal than the external mode. Only 2% of the external recordings were of good quality vs. 40% of the internal recordings. In the second stage, percentages of adequate UA recordings were nearly equal, approximately 30%. Inadequate external registrations were characterized by almost 30% of the time absent UA monitoring, whereas with inadequate internal registrations the primary problem was correct calibration. Conclusion: Intrapartum UA monitoring in daily practice via the direct mode provides a more recognizable UA trace.
引用
收藏
页码:197 / 201
页数:5
相关论文
共 50 条
  • [31] INTRAPARTUM FETAL MONITORING
    FILSHIE, M
    BRITISH JOURNAL OF HOSPITAL MEDICINE, 1974, 12 (01): : 33 - &
  • [32] INTRAPARTUM FETAL MONITORING
    NOCHIMSON, DJ
    CETRULO, CL
    JOURNAL OF FAMILY PRACTICE, 1977, 4 (05): : 845 - 849
  • [33] Intrapartum Fetal Monitoring
    Arnold, James J.
    Gawrys, Breanna L.
    AMERICAN FAMILY PHYSICIAN, 2020, 102 (03) : 158 - 167
  • [34] Intrapartum Fetal Monitoring
    Bailey, R. Eugene
    AMERICAN FAMILY PHYSICIAN, 2009, 80 (12) : 1388 - 1396
  • [35] INTRAPARTUM FETAL MONITORING
    ONG, BY
    BRADFORD, CR
    CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1983, 30 (01) : 91 - 93
  • [36] INTRAPARTUM MONITORING - AFRIKAANS
    不详
    SOUTH AFRICAN MEDICAL JOURNAL, 1978, 54 (16): : 634 - 634
  • [37] INTRAPARTUM FETAL MONITORING
    COCKBURN, JE
    PEARCE, JM
    CHAMBERLAIN, GVPC
    LANCET, 1992, 340 (8819): : 610 - 610
  • [38] INTRAUTERINE PRESSURE CATHETER PERFORMANCE IN AN IN-VITRO UTERINE MODEL - A SIMULATION OF PROBLEMS FOR INTRAPARTUM MONITORING
    DEVOE, LD
    SMITH, RP
    STOKER, R
    OBSTETRICS AND GYNECOLOGY, 1993, 82 (02): : 285 - 289
  • [39] UTERINE ACTIVITY MONITORING - FOR EVALUATION OF PROGRESS IN LABOR
    HUEY, JR
    POSTGRADUATE MEDICINE, 1977, 61 (04) : 149 - 154
  • [40] HOME UTERINE ACTIVITY MONITORING IN A RURAL SETTING
    VRBICKY, K
    HILL, WC
    LAMBERTZ, EL
    JURGENSEN, WW
    OBSTETRICS AND GYNECOLOGY, 1990, 76 (01): : S82 - S84