On-line preconcentration and determination of chromium in parenteral solutions by inductively coupled plasma optical emission spectrometry

被引:38
|
作者
Gil, RA
Cerutti, S
Gásquez, JA
Olsina, RA
Martinez, LD
机构
[1] Natl Univ San Luis, Dept Analyt Chem, Fac Chem Biochem & Pharm, RA-5700 San Luis, Argentina
[2] Consejo Nacl Invest Cient & Tecn, Buenos Aires, DF, Argentina
关键词
chromium speciation; on-line preconcentration; activated carbon; ICP-OES; parenteral solutions;
D O I
10.1016/j.sab.2005.02.008
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A method for the preconcentration and speciation of chromium was developed. On-line preconcentration and determination were obtained using inductively coupled plasma optical emission spectrometry (ICP-OES) coupled with flow injection. To determinate the chromium (III) present in parenteral solutions, chromium was retained on activated carbon at pH 5.0. On the other hand, a step of reduction was necessary in order to determine total chromium content. The Cr(VI) concentration was then determined by difference between the total chromium concentration and that of Cr(III). A sensitivity enrichment factor of 70-fold was obtained with respect to the chromium determination by ICP-OES without preconcentration. The detection limit for the preconcentration of 25 ml of sample was 29 ng l(-1). The precision for the 10 replicate determinations at the 5 mu g l(-1) Cr level was 2.3% relative standard deviation, calculated with the peak heights. The calibration graph using the preconcentration method for chromium species was linear with a correlation coefficient of 0.9995 at levels near the detection limits up to at least 60 mu g l(-1). The method can be applied to the determination and speciation of chromium in parenteral solutions. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:531 / 535
页数:5
相关论文
共 50 条
  • [41] Rapid determination of plutonium in urine using flow injection on-line preconcentration and inductively coupled plasma mass spectrometry
    Epov, VN
    Benkhedda, K
    Cornett, RJ
    Evans, RD
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2005, 20 (05) : 424 - 430
  • [42] On-line-sorption preconcentration and inductively coupled plasma atomic emission spectrometry determination of rare earth elements
    Grebneva, ON
    Kuzmin, NM
    Tsysin, GI
    Zolotov, YA
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (11) : 1417 - 1423
  • [43] Inductively coupled plasma optical emission spectrometric determination of gadolinium in urine using flow injection on-line sorption preconcentration in a knotted reactor
    Salonia, JA
    Gásquez, JA
    Martinez, LD
    Cerutti, S
    Kaplan, M
    Olsina, RA
    INSTRUMENTATION SCIENCE & TECHNOLOGY, 2006, 34 (03) : 305 - 316
  • [44] Preconcentration of cadmium with Saccharomyces cerevisiae and determination in river water using inductively coupled plasma optical emission spectrometry
    Biscaro, Priscila Andreoli
    Menegario, Amauri Antonio
    Tonello, Paulo Sergio
    Caldorin, Rogerio
    QUIMICA NOVA, 2007, 30 (02): : 323 - 326
  • [45] Validation of a Method for the Determination of Total Chromium in Rat Feces by Inductively Coupled Plasma Optical Emission Spectrometry
    Levine, Keith E.
    Stout, Matthew D.
    Ross, Glenn T.
    Essader, Amal S.
    Perlmutter, Jason M.
    Grohse, Peter M.
    Fernando, Reshan A.
    Lang, Michelle
    Collins, Bradley J.
    ANALYTICAL LETTERS, 2009, 42 (17) : 2729 - 2746
  • [46] Nanometer-size titanium dioxide microcolumn on-line preconcentration of trace metals and their determination by inductively coupled plasma atomic emission spectrometry in water
    Liang, P
    Qin, YC
    Hu, B
    Peng, TY
    Jiang, ZC
    ANALYTICA CHIMICA ACTA, 2001, 440 (02) : 207 - 213
  • [47] New magnetic chelating sorbent for chromium speciation by magnetic solid phase extraction on-line with inductively coupled plasma optical emission spectrometry
    Morales-Benitez, I.
    Montoro-Leal, P.
    Garcia-Mesa, J. C.
    Guerrerp, M. M. Lopez
    Alonso, E. Vereda
    TALANTA, 2023, 256
  • [48] Stripping analysis combined on-line with inductively coupled plasma-atomic emission spectrometry and with inductively coupled plasma-mass spectrometry
    Zhou, FM
    ELECTROANALYSIS, 1996, 8 (10) : 855 - 861
  • [49] PRECONCENTRATION AND DETERMINATION OF TRACE CHROMIUM(III) BY FLOW-INJECTION INDUCTIVELY-COUPLED PLASMA/ATOMIC EMISSION-SPECTROMETRY
    COX, AG
    MCLEOD, CW
    ANALYTICA CHIMICA ACTA, 1986, 179 : 487 - 490
  • [50] On-line preconcentration/determination of lead traces in bee honey by inductively coupled plasma optical emission spectrometry (ICP-OES) using a conical minicolumn packed with activated carbon
    Cerutti, S
    Orsi, RF
    Gásquez, JA
    Olsina, RA
    Martinez, LD
    JOURNAL OF TRACE AND MICROPROBE TECHNIQUES, 2003, 21 (03): : 421 - 432