Automatic brachytherapy seed placement under MRI guidance

被引:74
|
作者
Patriciu, Alexandru
Petrisor, Doru
Muntener, Michael
Mazilu, Dumitru
Schar, Michael
Stoianovici, Dan [1 ]
机构
[1] Johns Hopkins Univ, Sch Med, URobot Lab, Dept Urol, Baltimore, MD 21224 USA
[2] Johns Hopkins Univ, Sch Med, URobot Lab, Dept Radiol, Baltimore, MD 21224 USA
关键词
brachytherapy; IGI; image-guided robot; MR compatible; GUIDED PROSTATE BRACHYTHERAPY; SYSTEM INTEGRATION; FREE SURVIVAL; CANCER; DESIGN; INTERVENTIONS; MANIPULATOR; DEVICE; ROBOT;
D O I
10.1109/TBME.2007.900816
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The paper presents a robotic method of performing low dose rate prostate brachytherapy under magnetic resonance imaging (MRI) guidance. The design and operation of a fully automated MR compatible seed injector is presented. This is used with the MrBot robot for transperineal percutaneous prostate access. A new image-registration marker and algorithms are also presented. The system is integrated and tested with a 3T MRI scanner. Tests compare three different registration methods, assess the precision of performing automated seed deployment, and use the seeds to assess the accuracy of needle targeting under image guidance. Under the ideal conditions of the in vitro experiments, results show outstanding image-guided needle and seed placement accuracy.
引用
收藏
页码:1499 / 1506
页数:8
相关论文
共 50 条
  • [1] Cooperative Brachytherapy for Prostate Cancer Under MRI Guidance
    Dhaliwal, Sepaldeep Singh
    Chettibi, Taha
    Belarouci, Abdelkader
    Dherbomez, Gerald
    Coelen, Vincent
    Merzouki, Rochdi
    2019 FIFTH INTERNATIONAL CONFERENCE ON ADVANCES IN BIOMEDICAL ENGINEERING (ICABME), 2019, : 136 - 139
  • [2] CoBra: Towards Adaptive Robotized Prostate Brachytherapy under MRI Guidance
    Dhaliwal, Sepaldeep Singh
    Belarouci, Abdelkader
    Lopez, Mario Sanz
    Verbrugghe, Fabien
    Lakhal, Othman
    Dherbomez, Gerald
    Chettibi, Taha
    Merzouki, Rochdi
    2021 SIXTH INTERNATIONAL CONFERENCE ON ADVANCES IN BIOMEDICAL ENGINEERING (ICABME), 2021, : 81 - 84
  • [3] Placement of MammoSite® Brachytherapy Catheter Under Computed-Tomography Scan Guidance
    Mutyala, Subhakar
    Yaparpalvi, Ravindra
    Choi, Walter
    Mehta, Keyur
    Spierer, Marnee
    Kalnicki, Shalom
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2009, 8 (03) : 177 - 180
  • [4] Introducing MRI Guidance Into Skin Brachytherapy
    Harris, T.
    O'Farrell, D.
    Buzurovic, I.
    Friesen, S.
    Devlin, P.
    MEDICAL PHYSICS, 2020, 47 (06) : E631 - E631
  • [5] Dosimetric consequences of seed placement accuracy in permanent breast seed implant brachytherapy
    Guebert, Alexandra
    Roumeliotis, Michael
    Watt, Elizabeth
    Meyer, Tyler
    Quirk, Sarah
    BRACHYTHERAPY, 2021, 20 (03) : 664 - 672
  • [6] The Matador Technique: A technique to improve prostatic brachytherapy seed placement
    Pai, A.
    Rogers, P.
    Jones, A.
    AFRICAN JOURNAL OF UROLOGY, 2015, 21 (01) : 84 - 85
  • [7] Biopsy under MRI guidance
    David, Julie
    Lalonde, Lucie
    Trop, Isabelle
    IMAGERIE DE LA FEMME, 2008, 18 (01) : 11 - 23
  • [8] MRI-BASED PROSTATE BRACHYTHERAPY SEED LOCALIZATION
    Kuo, Nathanael
    Lee, Junghoon
    Tempany, Clare
    Stuber, Matthias
    Prince, Jerry
    2010 7TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, 2010, : 1397 - 1400
  • [9] Seed Placement in Permanent Breast Seed Implant Brachytherapy: Are Concerns Over Accuracy Valid?
    Morton, Daniel
    Hilts, Michelle
    Batchelar, Deidre
    Crook, Juanita
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (03): : 1050 - 1057
  • [10] An Automatic Needle and Seed Planning Algorithm for Head and Neck Seed Implant Brachytherapy
    Xiao, Z.
    Zhou, F.
    Liu, B.
    Ji, Z.
    Sun, H.
    Jiang, Y.
    Wang, J.
    Wu, Q.
    MEDICAL PHYSICS, 2022, 49 (06) : E171 - E171