Effects of sodium hypochlorite and ethylenediaminetetraacetic acid on rotary nickel-titanium instruments evaluated using atomic force microscopy

被引:39
|
作者
Ametrano, G. [1 ]
D'Anto, V. [1 ]
Di Caprio, M. P. [1 ]
Simeone, M. [1 ]
Rengo, S. [1 ]
Spagnuolo, G. [1 ]
机构
[1] Univ Naples Federico II, Dept Oral & Maxillofacial Sci, Naples, Italy
关键词
atomic force microscopy; EDTA; NaOCl; NiTi; ENDODONTIC INSTRUMENTS; CYCLIC FATIGUE; FRACTURE PROPERTIES; STAINLESS-STEEL; CORROSION; SURFACE; RESISTANCE; DEFECTS; TORQUE; FILES;
D O I
10.1111/j.1365-2591.2010.01799.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
P>Aim To use atomic force microscopy (AFM) to evaluate the effects of sodium hypochlorite (NaOCl) and ethylenediaminetetraacetic acid (EDTA) on the surface characteristics of ProTaper rotary nickel-titanium instruments. Methodology A total of twenty ProTaper (Dentsply Maillefer, Ballaigues, Switzerland) instruments (S1, S2, F1, F2) were divided into five groups: no immersion, immersion in 5.25% NaOCl for 5 or 10 min and immersion in 17% EDTA for 5 or 10 min. Twenty surface areas along 3- mm sections at the tip of the files (perfect squares of 1 x 1 mu m) were analysed by AFM operating in contact mode under ambient conditions. Three-dimensional images (400 x 400 lines) were processed using Gwyddion software, and the roughness average (Ra) and the root mean square value (RMS) of the scanned surface profiles were recorded. Data were analysed by means of anova and paired samples t-test. Results Three-dimensional AFM images of the surface of ProTaper instruments, including new and those immersed in NaOCl and EDTA solutions, revealed topographic irregularities at the nanometric scale. RMS and Ra values of instruments treated with NaOCl and EDTA solutions were statistically higher than that of the new ones (P < 0.05). Conclusions Atomic force microscopy three-dimensional images and roughness values indicated that short-term contact between NaOCl and EDTA endodontic irrigants and ProTaper instruments caused alterations in the surface of instruments.
引用
收藏
页码:203 / 209
页数:7
相关论文
共 50 条
  • [21] Wear of nickel-titanium endodontic instruments evaluated by scanning electron microscopy: Effect of ion implantation
    Rapisarda, E
    Bonaccorso, A
    Tripi, TR
    Condorelli, GG
    Torrisi, L
    JOURNAL OF ENDODONTICS, 2001, 27 (09) : 588 - 592
  • [22] In vivo assessment of the corrosion of nickel-titanium orthodontic archwires by using scanning electron microscopy and atomic force microscopy
    Ogawa, Celso M.
    Faltin Jr, Kurt
    Maeda, Fernando A.
    Ortolani, Cristina L. F.
    Guare, Renata O.
    Cardoso, Cristiane A. B.
    Costa, Andre L. F.
    MICROSCOPY RESEARCH AND TECHNIQUE, 2020, 83 (08) : 928 - 936
  • [23] Atomic force microscopy and energy dispersive X-ray spectrophotometry analysis of reciprocating and continuous rotary nickel-titanium instruments following root canal retreatment
    Kaya, Bulem Ureyen
    Erik, Cevat Emre
    Kiraz, Guelsen
    MICROSCOPY RESEARCH AND TECHNIQUE, 2019, 82 (07) : 1157 - 1164
  • [24] ROUGHNESS ANALYSIS AND DETERMINING DISPLACEMENT IN NICKEL-TITANIUM ALLOY BY ATOMIC FORCE MICROSCOPY
    Mendoza Carreno, Edwin Fernando
    Plata Gomez, Arturo
    REVISTA INNOVACIENCIA, 2015, 3 (01): : 4 - 8
  • [25] Evaluation of Intracanal Smear Layer Removal after Passive Microvolume Irrigation with Sodium Hypochlorite and Rotary Nickel-Titanium Instruments: An In-vitro Study
    Priyadarshini, M. S.
    Varma, K. Madhu
    Satish, R. Kalyan
    Durga, Indukuri Sai Lakshmi
    Praveen, Dalavai
    Sowmya, M.
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2023, 17 (10) : ZC23 - ZC27
  • [26] An in vitro comparison of apically extruded debris using three rotary nickel-titanium instruments
    Tasdemir, Tamer
    Er, Kursat
    Celik, Davut
    Aydemir, Hikmet
    JOURNAL OF DENTAL SCIENCES, 2010, 5 (03) : 121 - 125
  • [27] Smooth versus active shaft design using nickel-titanium rotary instruments.
    Bergmans, L
    Van Cleynenbreugel, J
    Beullens, M
    Wevers, M
    Van Meerbeck, B
    Lambrechts, P
    JOURNAL OF DENTAL RESEARCH, 2002, 81 : B238 - B238
  • [28] Effects of Simultaneous Liquid or Gel Sodium Hypochlorite Irrigation on the Cyclic Fatigue of Two Single-File Nickel-Titanium Instruments
    Pedulla, Eugenio
    La Rosa, Giusy Rita Maria
    Albani, Marco Sesto
    Isola, Gaetano
    Ozyurek, Taha
    Generali, Luigi
    APPLIED SCIENCES-BASEL, 2020, 10 (19):
  • [29] Cyclic Fatigue Resistance of Heat-treated Nickel-titanium Instruments after Immersion in Sodium Hypochlorite and/or Sterilization
    Pedulla, Eugenio
    Benites, Angela
    La Rosa, Giusy M.
    Plotino, Gianluca
    Grande, Nicola M.
    Rapisarda, Ernesto
    Generali, Luigi
    JOURNAL OF ENDODONTICS, 2018, 44 (04) : 648 - 653
  • [30] Heat treatment of nickel-titanium rotary endodontic instruments: effects on bending properties and shaping abilities
    Ebihara, A.
    Yahata, Y.
    Miyara, K.
    Nakano, K.
    Hayashi, Y.
    Suda, H.
    INTERNATIONAL ENDODONTIC JOURNAL, 2011, 44 (09) : 843 - 849