Analytical validation of microdialysis analyzer for monitoring glucose, lactate and pyruvate in cerebral microdialysates

被引:14
|
作者
Tholance, Yannick [1 ]
Barcelos, Gleicy [2 ,3 ]
Quadrio, Isabelle [1 ]
Renaud, Bernard [1 ]
Dallier, Frederic [4 ]
Perret-Liaudet, Armand [1 ]
机构
[1] HCL, Ctr Biol & Pathol Est, Neurobiol Lab, Lyon, France
[2] HUG, Dept Anesthesiol Phannacol & Soins Intensifs, Geneva, Switzerland
[3] CNRS, Lab Neurophormacol EAC 5006, Lyon, France
[4] HCL, Hop Neurol & Neurochirurg Pierre Wertheimer, Serv Anesthesie & Reanimat, Lyon, France
关键词
Cerebral microdialysis; Glucose; Lactate/pyruvate ratio; Analytical validation; Brain injury; ANEURYSMAL SUBARACHNOID HEMORRHAGE; CRITICALLY-ILL PATIENTS; INTRACEREBRAL MICRODIALYSIS; BIOLOGICAL VARIATION; NEUROLOGICAL DEFICITS; BEDSIDE MICRODIALYSIS; STATUS EPILEPTICUS; CONTROLLED-TRIAL; AMINO-ACIDS; PHARMACOKINETICS;
D O I
10.1016/j.cca.2010.12.025
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Cerebral microdialysis is a valuable tool for neurochemical monitoring of acute brain injury. We performed an independent analytical validation of glucose, lactate and pyruvate methods on the new ISCUSflex new analyzer developed by CMA Microdialysis. Methods: Evaluation of analytical parameters included limit of detection, limit of quantification, linearity, intra- and inter-assay imprecision expressed as the coefficient of variation (CV), recovery, inter-sample and inter-reagent contamination, drug and bilirubin interferences, sample stability, method comparison. Results: Linearity ranges were 0.1-25 mmol/L, 0.2-12 mmol/L and 19-1500 mu mol/L for glucose, lactate and pyruvate respectively. For critical threshold, intra- and inter-assay CVs were 3.1/4.5% for glucose (1 mmol/L), 3.5/4% for lactate (4 mmol/L) and 3.3/4.3% for pyruvate (100 mu mol/L). Inter-assay CVs for lactate/pyruvate (LPR) and lactate/glucose (LGR) ratios were 5.9% and 6.0% respectively. For glucose, lactate, pyruvate. LPR and LGR, the reference change values (RCV) were 20%, 26%, 20%, 27% and 28% respectively. Practically, variations below 27% between two successive LPR values could not be interpreted as significant. Conclusion: These data prove that ISCUSflex has the qualities required for clinical application in neuro-intensive care. Correct clinical interpretation of data need the implementation of a strict quality control program and strong cooperation between clinicians and biologists. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:647 / 654
页数:8
相关论文
共 50 条
  • [1] Analytical validation of ISCUS Flex analyzer for monitoring Glucose, lactate and pyruvate in cerebral microdialysates
    Tholance, Yannick
    Barcelos, Gleicy
    Renaud, Bernard
    Cartier, Regine
    Dailler, Frederic
    Perret-liaudet, Armand
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2010, 27 : S17 - S18
  • [2] Microdialysis monitoring of glucose, lactate, glycerol, and pyruvate in patients with diabetic ketoacidosis
    Ciechanowska, Anna
    Ladyzynski, Piotr
    Wojcicki, Jan M.
    Sabalinska, Stanislawa
    Krzymien, Janusz
    Pulawska, Elzbieta
    Karnafel, Waldemar
    Foltynski, Piotr
    Kawiak, Jerzy
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2013, 36 (12): : 869 - 877
  • [3] Simultaneous monitoring of extracellular glucose, pyruvate, lactate and glutamate in gerbil cortex during focal cerebral ischemia by dual probe microdialysis
    Yang, DY
    Tsai, TH
    Cheng, CH
    Lee, CW
    Chen, SH
    Cheng, FC
    JOURNAL OF CHROMATOGRAPHY A, 2001, 913 (1-2) : 349 - 354
  • [4] Simultaneously monitoring cerebral blood flow, glucose, pyruvate, lactate and glutamate in gerbils subjected to traumatic brain injury: A microdialysis study
    Wang, Yeou-Chih
    Cheng, Fu-Chou
    Pan, Hung-Chuan
    Lee, Shiuh-Dung
    Lee, Chi-Wen
    JOURNAL OF NEUROTRAUMA, 2006, 23 (06) : 1020 - 1020
  • [5] Simultaneous monitoring of extracellular glucose, pyruvate, and lactate in rat testies during ischemia: a microdialysis study
    Chen-Li Cheng
    Chi-Re Yang
    Dar-Yu Yang
    Jon-Son Kuo
    Yi-Fang Huang
    Fu-Chou Cheng
    Urological Research, 2001, 29 : 272 - 277
  • [6] Simultaneous monitoring of extracellular glucose, pyruvate, and lactate in rat testies during ischemia: a microdialysis study
    Cheng, CL
    Yang, CR
    Yang, DY
    Kuo, JS
    Huang, YF
    Cheng, FC
    UROLOGICAL RESEARCH, 2001, 29 (04): : 272 - 277
  • [7] Monitoring of extracellular pyruvate, lactate, and ascorbic acid during cerebral ischemia: a microdialysis study in awake gerbils
    Cheng, FC
    Yang, LL
    Yang, DY
    Tsai, TH
    Lee, CW
    Chen, SH
    JOURNAL OF CHROMATOGRAPHY A, 2000, 870 (1-2) : 389 - 394
  • [8] Monitoring of lactate and glucose levels in engineered cartilage construct by microdialysis
    Boubriak, OA
    Urban, JPG
    Cui, ZF
    JOURNAL OF MEMBRANE SCIENCE, 2006, 273 (1-2) : 77 - 83
  • [9] Intraperitoneal microdialysis: Postoperative monitoring of splanchnic ischemia by measurements of the lactate pyruvate ratio
    Jansson, M.
    Strand, I.
    Jansson, K.
    CRITICAL CARE MEDICINE, 2006, 34 (10) : 2695 - 2697
  • [10] Is There Any Association Between Cerebral Vasoconstriction/Vasodilatation and Microdialysis Lactate to Pyruvate Ratio Increase?
    Shadnaz Asgari
    Paul Vespa
    Xiao Hu
    Neurocritical Care, 2013, 19 : 56 - 64