Absolute quantification of myocardial blood flow with 13N-ammonia and 3-dimensional PET

被引:41
|
作者
Schepis, Tiziano [1 ]
Gaemperli, Oliver [1 ]
Treyer, Valerie [1 ]
Valenta, Ines [1 ]
Burger, Cyrill [1 ]
Koepfli, Pascal [1 ]
Namdar, Mehdi [1 ]
Adachi, Itaru [4 ]
Alkadhi, Hatem [2 ]
Kaufmann, Philipp A. [1 ,3 ]
机构
[1] Univ Zurich Hosp, Ctr Cardiovasc, Clin Nucl Med, CH-8091 Zurich, Switzerland
[2] Univ Zurich Hosp, Inst Diagnos Radiol, CH-8091 Zurich, Switzerland
[3] Univ Zurich, Ctr Integrat Human Physiol, CH-8091 Zurich, Switzerland
[4] Osaka Med Coll, Dept Radiol, Takatsuki, Osaka 569, Japan
关键词
myocardial blood flow; 3-dimensional PET; N-13-NH3; image reconstruction algorithms;
D O I
10.2967/jnumed.107.044099
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to compare 2-dimensional (2D) and 3-dimensional (3D) dynamic PET for the absolute quantification of myocardial blood flow (MBF) with N-13-ammonia (N-13-NH3)Methods: 2D and 3D MBF measurements were collected from 21 patients undergoing cardiac evaluation at rest (n = 14) and during standard adenosine stress (n = 7). A lutetium yttrium oxyorthosilicate-based PET/CT system with retractable septa, enabling the sequential acquisition of 2D and 3D images within the same patient and study, was used. All 2D studies were performed by injecting 700-900 MBq of N-13-NH3. For 14 patients, 3D studies were performed with the same injected N-13-NH3 dose as that used in 2D studies. For the remaining 7 patients, 3D images were acquired with a lower dose of N-13-NH3, that is, 500 MBq. 2D images reconstructed by use of filtered back-projection (FBP) provided the reference standard for MBF measurements. 3D images were reconstructed by use of Fourier rebinning (FORE) with FBP (FORE-FBP), FORE with ordered-subsets expectation maximization (FORE-OSEM), and a reprojection algorithm (RP). Results: Global MBF measurements derived from 3D PET with FORE-FBP (r = 0.97), FORE-OSEM (r = 0.97), and RP (r = 0.97) were well correlated with those derived from 2D FBP (all Ps < 0.0001). The mean +/- SD differences in global MBF measurements between 3D FORE-FBP and 2D FBP and between 3D FORE-OSEM and 2D FBP were 0.01 +/- 0.14 and 0.01 +/- 0.15 mL/min/g, respectively. The mean +/- SD difference in global MBF measurements between 3D RP and 2D FBP was 0.00 +/- 0.16 mL/min/g. The best correlation between 2D PET and 3D PET performed with the lower injected activity was found for the 3D FORE-FBP reconstruction algorithm (r = 0.95,P< 0.001). Conclusion: For this scanner type, quantitative measurements of MBF with 3D PET and N-13-NH3 were in excellent agreement with those obtained with the 2D technique, even when a lower activity was injected.
引用
收藏
页码:1783 / 1789
页数:7
相关论文
共 50 条
  • [1] Absolute quantification of myocardial blood flow with 13-N ammonia and 3-dimensional PET
    Schepis, Tiziano
    Gaemperli, Oliver
    Treyer, Valerie
    Valenta, Ines
    Burger, Cyrill
    Koepfli, Pascal
    Namdar, Mehdi
    Adachi, Itaru
    Alkadhi, Hatern
    Buck, Alfred
    Kaufmann, Philipp A.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (09) : 179A - 179A
  • [2] Normal value of myocardial blood flow and myocardial flow reserve evaluated by 3-dimensional dynamic PET with 13N-ammonia
    Kiso, Keisuke
    Shimosegawa, Eku
    Kanai, Yasukazu
    Fujino, Koichi
    Hatazawa, Jun
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [3] Absolute myocardial blood flow evaluation with 13N-ammonia and 3D PET-CT
    Leccisotti, L.
    Indovina, L.
    Paraggio, L.
    Calcagni, M. L.
    Giordano, A.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2011, 13 (0A) : A80 - A80
  • [4] Quantification of myocardial blood flow using 13N-ammonia and PET:: Comparison of tracer models
    Choi, Y
    Huang, SC
    Hawkins, RA
    Kim, JY
    Kim, BT
    Hoh, CK
    Chen, K
    Phelps, ME
    Schelbert, HR
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (06) : 1045 - 1055
  • [5] Validation of an axially distributed model for quantification of myocardial blood flow using 13N-ammonia PET
    Alessio, Adam M.
    Bassingthwaighte, James B.
    Glenny, Robb
    Caldwell, James H.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2013, 20 (01) : 64 - 75
  • [6] Effect of temporal sampling protocols on myocardial blood flow quantification using 13N-ammonia PET
    Hu, Yingqi
    Wang, Ruonan
    Zhang, Chuxin
    Li, Yuanyuan
    Wang, Yanhui
    Diao, Xin
    Zhang, Jun
    Jia, Yanni
    Li, Sijin
    JOURNAL OF NUCLEAR MEDICINE, 2023, 64
  • [7] Validation of an axially distributed model for quantification of myocardial blood flow using 13N-ammonia PET
    Adam M. Alessio
    James B. Bassingthwaighte
    Robb Glenny
    James H. Caldwell
    Journal of Nuclear Cardiology, 2013, 20 : 64 - 75
  • [8] Evaluation of myocardial blood flow in patients with myocardial bridging: A 13N-ammonia PET study
    Monroy-Gonzalez, A.
    Juarez-Orozco, L.
    Berrios-Barcenas, E.
    Gonzalez-Padilla, C.
    Martinez-Aguilar, M.
    Guinto-Nishimura, G.
    Dierckx, R.
    Tio, R.
    Slart, R. H. J. A.
    Alexanderson, E.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S520 - S521
  • [9] The reproducibility of time-of-flight PET and conventional PET for the quantification of myocardial blood flow and coronary flow reserve with 13N-ammonia
    Masaya Suda
    Masahisa Onoguchi
    Takeshi Tomiyama
    Keiichi Ishihara
    Naoto Takahashi
    Minoru Sakurai
    Keiichi Matsumoto
    Shin-ichiro Kumita
    Journal of Nuclear Cardiology, 2016, 23 : 457 - 472
  • [10] The reproducibility of time-of-flight PET and conventional PET for the quantification of myocardial blood flow and coronary flow reserve with 13N-ammonia
    Suda, Masaya
    Onoguchi, Masahisa
    Tomiyama, Takeshi
    Ishihara, Keiichi
    Takahashi, Naoto
    Sakurai, Minoru
    Matsumoto, Keiichi
    Kumita, Shin-ichiro
    JOURNAL OF NUCLEAR CARDIOLOGY, 2016, 23 (03) : 457 - 472