A comparison of CAT Doppler and oscillometric Memoprint machines for non-invasive blood pressure measurement in conscious cats

被引:42
|
作者
Jepson, RE [1 ]
Hartley, V [1 ]
Mendl, M [1 ]
Caney, SM [1 ]
Gould, DJ [1 ]
机构
[1] Univ Bristol, Dept Clin Vet Med, Bristol BS40 5DU, Avon, England
关键词
D O I
10.1016/j.jfms.2004.08.003
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Indirect blood pressure measurements were compared in 28 conscious cats using Doppler and oscillometric blood pressure-measuring devices. Ten cats were used to compare Doppler measurements between two examiners and 18 cats were used to compare Doppler and oscillometric measurements. The Doppler machine obtained systolic and diastolic blood pressure readings in 100% and 51% of attempts, respectively. With the oscillometric machine, systolic and diastolic blood pressure readings were obtained in 52% of the attempts. With the Doppler, measures of mean systolic blood pressure between two examiners were positively correlated, but there was no correlation for diastolic blood pressure measures. When comparing the results obtained by Doppler and oscillometric machines there was no significant difference between mean systolic blood pressure readings, but the oscillometric machine produced significantly higher estimates of diastolic blood pressure. In both cases, the standard deviations for the oscillometric machine were considerably larger than those for the Doppler machine. The first reading of systolic blood pressure obtained with the Doppler machine was an excellent predictor of the mean of five readings, but this was not so for the oscillometric machine. It took less than 5 min to obtain five readings in 37.5% of cases with the Doppler machine but this was true for only 5% of cases with the oscillometric machine. Two cats with ophthalmological lesions consistent with systemic hypertension were identified. In these two patients, systolic blood pressure measurements were between 200 and 225 mmHg when measured by Doppler, and between 140 and 150 mmHg when measured by the oscillometric machine. This suggests that a lower reference range for normal systolic blood pressure values should be used for the oscillometric device. (c) 2004 ESFM and AAFP. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:147 / 152
页数:6
相关论文
共 50 条
  • [31] Mathematical modelling of non-invasive oscillometric finger mean blood pressure measurement by maximum oscillation criterion
    Institute of Physiology, University of Tartu, Tartu, Estonia
    不详
    Med. Biol. Eng. Comput., 6 (784-788):
  • [32] INDIRECT MEASUREMENT OF BLOOD-PRESSURE IN NEONATES AND INFANTS UTILIZING AN AUTOMATIC NON-INVASIVE OSCILLOMETRIC MONITOR
    FRIESEN, RH
    LICHTOR, JL
    ANESTHESIA AND ANALGESIA, 1981, 60 (10): : 742 - 745
  • [33] Non-invasive blood pressure measurement displays
    Scott, D. H. T.
    ANAESTHESIA, 2018, 73 (10) : 1299 - 1299
  • [34] NON-INVASIVE BLOOD-PRESSURE MEASUREMENT
    PAULUS, DA
    MEDICAL INSTRUMENTATION, 1981, 15 (02): : 91 - 94
  • [35] Radial vs coccygeal artery Doppler blood pressure measurement in conscious cats
    Zeugswetter, Florian K.
    Tichy, Alexander
    Weber, Katharina
    JOURNAL OF FELINE MEDICINE AND SURGERY, 2018, 20 (10) : 968 - 972
  • [36] Oscillometric non-invasive blood pressure measurements: The influence of the make of instrument on readings?
    Amoore, JN
    Geake, WB
    Scott, DHT
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (02) : 131 - 134
  • [37] Oscillometric non-invasive blood pressure measurements: The influence of the make of instrument on readings?
    Amoore, J.N.
    Geake, W.B.
    Scott, D.H.T.
    Medical and Biological Engineering and Computing, 1997, 35 (02): : 131 - 134
  • [38] Oscillometric non-invasive blood pressure measurements: the influence of the make of instrument on readings?
    J. N. Amoore
    W. B. Geake
    D. H. T. Scott
    Medical and Biological Engineering and Computing, 1997, 35 : 131 - 134
  • [39] An automated system for the functional evaluation of oscillometric non-invasive blood pressure monitors
    Geake, WB
    Amoore, JN
    Scott, DHT
    JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1995, 19 (05) : 162 - 176
  • [40] Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia
    Lakhal, Karim
    Martin, Maelle
    Ehrmann, Stephan
    Faiz, Sofian
    Rozec, Bertrand
    Boulain, Thierry
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2018, 32 (04) : 707 - 715