Dosimetric properties of equivalent-quality flattening filter-free (FFF) and flattened photon beams of Versa HD linear accelerator

被引:8
|
作者
Meshram, Mukesh N. [1 ]
Pramanik, Srimanta [1 ]
Ranjith, C. P. [1 ]
Gopal, Saravana K. [1 ]
Dobhal, Rishabh [1 ]
机构
[1] CIMS Hosp Pvt Ltd, Ahmadabad 380060, Gujarat, India
来源
关键词
equivalent-quality beam; flattening filter free; dosimetric properties; photon beam data; Versa HD linac; FREE CLINICAL ACCELERATOR; MONTE-CARLO; COLLIMATOR; THERAPY;
D O I
10.1120/jacmp.v17i3.6173
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study presents the basic dosimetric properties of photon beams of a Versa HD linear accelerator (linac), which is capable of delivering flattening filter-free (FFF) beams with a beam quality equivalent to the corresponding flattened beams based on comprehensive beam data measurement. The analyzed data included the PDDs, profiles, penumbra, out-of-field doses, surface doses, output factors, head and phantom scatter factors, and MLC transmissions for both FFF and flattened beams of 6 MV and 10 MV energy from an Elekta Versa HD linac. The 6MVFFF and 10MVFFF beams had an equivalent mean energy to the flattened beams and showed less PDD variations with the field sizes. Compared with their corresponding flattened beams, D-max was deeper for FFF beams for all field sizes; the ionization ratio variations with the field size were lower for FFF beams; the out-of-field doses were lower and the penumbras were sharper for the FFF beams; the off-axis profile variations with the depths were lesser for the FFF beams. Further, the 6MVFFF and 10MVFFF beams had 35.7% and 40.9% less variations in output factor with the field size, respectively. The collimator exchange effect was reduced in the FFF mode. The head scatter factor showed 59.1% and 73.6% less variations, on average, for the 6MVFFF and 10MVFFF beams, respectively; the variations in the phantom scatter factor were also smaller. The surface doses for all beams increased linearly with the field size. The 6MVFFF and 10MVFFF beams had higher surface doses than the corresponding flattened beams for field sizes of up to 10x10cm(2) but had lower surface doses for larger fields. Both FFF beams had lower average MLC transmissions than the flattened beams. The finding that the FFF beams were of equivalent quality to the corresponding flattened beams indicates a significant difference from the data on unmatched FFF beams.
引用
下载
收藏
页码:358 / 370
页数:13
相关论文
共 50 条
  • [1] Dosimetric Characteristics of Equivalent Quality Flattening Filter Free (FFF) and Flattened Photon Beams
    Meshram, M. N.
    Pramanik, S.
    Ranjith, C. P.
    Bhavsar, D.
    Jani, K. R.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E548 - E549
  • [2] Dosimetric Evaluation of Flattening Filter-free (FFF) Photon Beams for Lung SBRT
    Chen, Q.
    Cai, J.
    Wu, Q.
    Yin, F.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S609 - S609
  • [3] Surface dose variations in 6 and 10 MV flattened and flattening filter-free (FFF) photon beams
    Cashmore, Jason
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05): : 293 - 307
  • [4] Surface Dose Evaluation of the Equivalent Quality Flattening Filter-Free Photon Beams
    Meshram, M.
    Pramanik, S.
    Ranjith, C. P.
    Meshram, M.
    MEDICAL PHYSICS, 2015, 42 (06) : 3502 - 3502
  • [5] Dosimetric properties of photon beams from a flattening filter free clinical accelerator
    Vassiliev, ON
    Titt, U
    Pönisch, F
    Kry, SF
    Mohan, R
    Gillin, MT
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (07): : 1907 - 1917
  • [6] Dosimetric properties of photon beams from a flattening filter free clinical accelerator
    Vassiliev, O
    Titt, U
    Kry, S
    Poenisch, F
    Gillin, M
    Mohan, R
    MEDICAL PHYSICS, 2005, 32 (06) : 1896 - 1896
  • [7] Dosimetric Properties of Flattening Filter Free Photon Beams from a New Clinical Accelerator
    Mok, E.
    Kielar, K.
    Hsu, A.
    Maxim, P.
    Xing, L.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [8] Evaluation of the TrueBeam machine performance check (MPC) beam constancy checks for flattened and flattening filter-free (FFF) photon beams
    Barnes, Michael P.
    Greer, Peter B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (01): : 139 - 150
  • [9] Pinnacle3 modeling and end-to-end dosimetric testing of a Versa HD linear accelerator with the Agility head and flattening filter-free modes
    Saenz, Daniel L.
    Narayanasamy, Ganesh
    Cruz, Wilbert
    Papanikolaou, Nikos
    Stathakis, Sotirios
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (01): : 192 - 206
  • [10] Investigation of the bolusing effect of the Varian Exact™ IGRT couch on flattened and flattening filter-free (FFF) photon beams of a Varian TrueBeam linac
    Sdrolia, Athina
    Tambe, Nilesh
    Marsden, Jenny E.
    Wilson, Mark L.
    Colley, William P.
    Beavis, Andrew W.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (01):