A High-Frequency Annular-Array Transducer Using an Interdigital Bonded 1-3 Composite

被引:32
|
作者
Chabok, Hamid Reza [1 ]
Cannata, Jonathan M. [1 ]
Kim, Hyung Ham [1 ]
Williams, Jay A. [1 ]
Park, Jinhyoung [1 ]
Shung, K. Kirk [1 ]
机构
[1] Univ So Calif, Dept Biomed Engn, NIH Resource Ctr Med Ultrason Transducer Technol, Los Angeles, CA 90089 USA
关键词
ULTRASOUND TRANSDUCERS; FABRICATION; DESIGN; PULSE;
D O I
10.1109/TUFFC.2011.1787
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper reports the design, fabrication, and characterization of a 1-3 composite annular-array transducer. An interdigital bonded (IB) 1-3 composite was prepared using two IB operations on a fine-grain piezoelectric ceramic. The final composite had 19-mu m-wide posts separated by 6-mu m-wide polymer kerfs. A novel method to remove metal electrodes from polymer portions of the 1-3 composite was established to eliminate the need for patterning and aligning the electrode on the composite to the electrodes on a flexible circuit. Unloaded epoxy was used for both the matching and backing layers and a flexible circuit was used for interconnect. A prototype array was successfully fabricated and tested. The results were in reasonable agreement with those predicted by a circuit-analogous model. The average center frequency estimated from the measured pulse-echo responses of array elements was 33.5 MHz and the -6-dB fractional bandwidth was 57%. The average insertion loss recorded was 14.3 dB, and the maximum crosstalk between the nearest-neighbor elements was less than -37 dB. Images of a wire phantom and excised porcine eye were obtained to show the capabilities of the array for high-frequency ultrasound imaging.
引用
收藏
页码:206 / 214
页数:9
相关论文
共 50 条
  • [21] Integrated circuit for high-frequency ultrasound annular array
    Talman, JR
    Garverick, SL
    Lockwood, GR
    PROCEEDINGS OF THE IEEE 2003 CUSTOM INTEGRATED CIRCUITS CONFERENCE, 2003, : 477 - 480
  • [22] ELECTROACOUSTIC CHARACTERISTICS OF A 1-3 COMPOSITE TRANSDUCER
    THOMPSON, W
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1987, 81 : S88 - S88
  • [23] Ultra-wide operation band of the high-frequency underwater acoustic transducer realized by two-layer 1-3 piezoelectric composite
    Ji, Bocheng
    Hong, Lianjin
    Lan, Yu
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2021, 150 (05): : 3474 - 3484
  • [24] Piezoelectric 1-3 composites for high frequency ultrasonic transducer applications
    Zhang, D
    Su, B
    Button, TW
    Abrar, A
    Kirk, KJ
    Cochran, S
    FERROELECTRICS, 2004, 304 : 1031 - 1035
  • [25] High-Frequency Annular Array Fabrication Using a Flex Circuit Matching Layer
    Lay, Holly S.
    Simpson, Eric A.
    Griffin, Greg
    Lockwood, Geoffrey R.
    ULTRASONIC IMAGING, 2012, 34 (03) : 196 - 204
  • [26] High frequency copolymer annular array ultrasound transducer fabrication technology
    Gottlieb, EJ
    Cannata, JM
    Hu, CH
    Strung, KK
    2005 IEEE Ultrasonics Symposium, Vols 1-4, 2005, : 121 - 124
  • [27] HIGH-FREQUENCY ACOUSTOOPTIC DEVICE USING ULTRASONICALLY BONDED AND SPUTTER-MACHINED LINBO3 TRANSDUCER
    UCHIDA, N
    FUKUNISH.S
    SAITO, S
    IEEE JOURNAL OF QUANTUM ELECTRONICS, 1973, QE 9 (06) : 660 - 660
  • [28] Micromachined High Frequency PMN-PT 1-3 Composite Transducer via Cold Ablation Process
    Lv, Jiabing
    Li, Zhangjian
    Cui, Yaoyao
    Jian, Xiaohua
    INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS 2021), 2021,
  • [29] Photoacoustic imaging of the microvasculature with a high-frequency ultrasound array transducer
    Zemp, Roger J.
    Bitton, Rachel
    Li, Meng-Lin
    Shung, K. Kirk
    Stoica, George
    Wang, Lihong V.
    JOURNAL OF BIOMEDICAL OPTICS, 2007, 12 (01)
  • [30] Piezoelectric Nanotube Array for Broadband High-Frequency Ultrasonic Transducer
    Liew, Weng Heng
    Yao, Kui
    Chen, Shuting
    Tay, Francis Eng Hock
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2018, 65 (03) : 457 - 464