Preparation of Commutable Frozen Human Serum Candidate Reference Materials for Standardization of Sodium Measurements

被引:0
|
作者
Zhang, Rui [1 ]
Zhang, Shunli [1 ]
Wang, Qingtao [1 ,2 ]
机构
[1] Capital Med Univ, Beijing Chao Yang Hosp, Dept Clin Lab, 8 Gongtinan Rd, Beijing 100020, Peoples R China
[2] Beijing Ctr Clin Labs, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
sodium; candidate reference materials; homogeneity; stability; commutability; value assignment; ION CHROMATOGRAPHY; METROLOGICAL TRACEABILITY; ELECTROLYTE DISTURBANCES; LABORATORY MEDICINE; POTASSIUM;
D O I
10.7754/Clin.Lab.2017.170205
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The measurement of serum sodium is important in the diagnosis of diseases. However, the accuracy and comparability of results in clinical laboratories are not ideal. The aim of this study was to prepare the candidate reference materials (RMs) to standardize serum sodium measurements. Methods: Fresh sera without hemolysis, lipemia, and choloplania were collected and packed in the cryovials. According to ISO Guide 35, the homogeneity and stability were tested. The value was assigned by transfer from the National Institute of Standards and Technology (NIST) SRM919b using the reference method of ion chromatography (IC) and uncertainty was calculated. The commutability of candidate RMs were observed between the reference method and the two analyzed systems and then distributed to 47 laboratories to apply in routine assays. Results: The F values of the homogeneity test were less than 405; stability can last at least 12 months, 30 days, 15 days, and 7 days at -80 degrees C, 2 - 8 degrees C, room temperature (20 - 24 degrees C), and 37 degrees C, respectively. The results of three levels of candidate RMs for sodium were (159.00 +/- 2.70), (139.16 +/- 2.57), and (124.71 +/- 3.12) mmol/L; coordinate dots of candidate RMs were all within the 95% confidence interval range of a 25-serum regression line. More than 76.6% of laboratories were within a bias of +/- 1.5% from the target values. Conclusions: The homogeneity, stability, and commutability of candidate RMs all meet the requirement, and the target values are assigned accurately. These new RMs of sodium in the human serum pool can be used to set up the traceability chain to improve the comparability of measurement results.
引用
收藏
页码:1243 / 1250
页数:8
相关论文
共 50 条
  • [31] STANDARDIZATION, REFERENCE MATERIALS AND COMPARATIVE MEASUREMENTS FOR SURFACE-AREA AND PORE CHARACTERIZATION
    ROBENS, E
    MULLER, U
    UNGER, KK
    VAKUUM-TECHNIK, 1989, 38 (3-4): : 65 - 75
  • [32] Commutability of Reference Materials for α-Fetoprotein in Human Serum
    Yue, Yuhong
    Zhang, Shunli
    Xu, Zhenzhen
    Chen, Xi
    Wang, Qingtao
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2017, 141 (10) : 1421 - 1427
  • [33] Development of a candidate reference measurement procedure for global standardization of serum/plasma apolipoproteins, including apo(a)
    Ruhaak, R.
    Kuklenyik, Z.
    Dittrich, J.
    Ceglarek, U.
    Cobbaert, C.
    CLINICA CHIMICA ACTA, 2019, 493 : S197 - S198
  • [34] PREPARATION AND CALIBRATION OF A REFERENCE PREPARATION FOR HUMAN-SERUM PROTEINS
    HAMILTON, EJ
    TOSTOWARYK, W
    CARON, JA
    GODARD, A
    GRIFFITHS, BW
    CLINICAL BIOCHEMISTRY, 1984, 17 (01) : 27 - 30
  • [35] An updated protocol based on CLSI document C37 for preparation of off-the-clot serum from individual units for use alone or to prepare commutable pooled serum reference materials
    Danilenko, Uliana
    Vesper, Hubert W.
    Myers, Gary L.
    Clapshaw, Patric A.
    Camara, Johanna E.
    Miller, W. Greg
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (03) : 368 - 374
  • [36] ISOTACHOPHORESIS AS A CANDIDATE REFERENCE METHOD IN ANALYTICAL-CHEMISTRY - DETERMINATION OF SODIUM IN SERUM
    LEMMENS, AAG
    REIJENGA, JC
    EVERAERTS, FM
    JANSSEN, RTP
    HULSMAN, JARJ
    MEIJERS, CAM
    JOURNAL OF CHROMATOGRAPHY, 1985, 320 (01): : 193 - 197
  • [37] Development of a glucose reference material in human serum for clinical assay standardization
    Wang, Zhichao
    Du, Yuanyuan
    Zhang, Qiaoxuan
    Yan, Jun
    Zhan, Min
    Zhang, Pengwei
    Wan, Zemin
    Wang, Jianbing
    Huang, Xianzhang
    Han, Liqiao
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2023, 228
  • [38] Standardization of cardiac troponin I assays: pilot evaluation of ten candidate reference materials.
    Kahn, SE
    Apple, FS
    Bodor, GS
    Panteghini, M
    Welch, MH
    Wu, AHB
    Dalluge, J
    Bunk, D
    Christenson, RH
    CLINICAL CHEMISTRY, 2000, 46 (06) : A89 - A89
  • [39] PREPARATION AND CHARACTERIZATION OF A HUMAN-SERUM MATRIX SUITABLE FOR QUALITY-CONTROL OR REFERENCE MATERIALS
    BAIS, R
    OLOUGHLIN, PD
    PHILCOX, JC
    EDWARDS, JB
    PATHOLOGY, 1983, 15 (01) : 15 - 19
  • [40] Toward standardization of cardiac troponin I measurements part II: Assessing commutability of candidate reference materials and harmonization of cardiac troponin I assays
    Christenson, Robert H.
    Duh, Show Hong
    Apple, Fred S.
    Bodor, Geza S.
    Bunk, David M.
    Panteghini, Mauro
    Welch, Michael J.
    Wu, Alan H. B.
    Kahn, Stephen E.
    CLINICAL CHEMISTRY, 2006, 52 (09) : 1685 - 1692