Safety of the Minimum Ophthalmic Viscosurgical Device Technique in Phakic Implantable Collamer Lens Implantation

被引:8
|
作者
Chen, Di [1 ]
Cui, Ge [1 ]
Wang, Xuejiao [1 ]
Li, Ying [1 ]
Luo, Yan [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Peking Union Med Coll Hosp, Dept Ophthalmol, Beijing, Peoples R China
关键词
2 VISCOELASTIC AGENTS; INTRAOCULAR-PRESSURE; ANTERIOR SEGMENT; CATARACT-SURGERY; HIGH MYOPIA; CORNEAL ENDOTHELIUM; FOLLOW-UP; CHAMBER; MODERATE; HOLE;
D O I
10.3928/1081597X-20191119-02
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the safety of the minimum ophthalmic viscosurgical device (OVD) technique with the standard procedure in phakic Visian Implantable Collamer Lens (ICL) (STAAR Surgical AG, Nidau, Switzerland) implantation. METHODS: This retrospective cohort study evaluated a total of 147 eyes of 74 patients who underwent ICL implantation with the minimum OVD technique (minimum OVD group) and 154 eyes of 77 patients with the standard procedure (standard OVD group). Intraoperative and postoperative complications were recorded. Preoperative and postoperative visual acuity, intraocular pressure (IOP), aqueous depth (AQD), and central corneal endothelial cell density (ECD) were collected and analyzed over the 12-month follow-up. Lens vault and occurrence of paracentesis after surgery were also recorded. RESULTS: No intraocular complications were observed. No difference was found in visual outcomes, lens vault, and AQD at all time points between the two groups (P > .05). The minimum OVD group had significantly lower IOP than the standard OVD group at 2 hours (17.04 +/- 4.21 vs 19.40 +/- 6.78 mm Hg, P < .001) and 3 hours (15.12 +/- 3.38 vs 17.15 +/- 5.09 mm Hg, P < .001) postoperatively. The IOP gradually returned to the preoperative level after 24 hours postoperatively. The occurrence rate of paracentesis was significantly less in the minimum OVD group compared with the standard group (0.68% [1 of 147] vs 3.2% [5 of 154], P <.001). ECD was not significantly different between groups at all time points (P > .05). CONCLUSIONS: The minimum OVD technique could achieve visual and structural outcomes comparable to the standard procedure without additional damage to the corneal endothelial cells, while reducing the IOP fluctuations after surgery.
引用
收藏
页码:42 / 48
页数:7
相关论文
共 50 条
  • [41] Factors Affecting Vaulting After Implantable Collamer Lens Implantation
    Kamiya, Kazutaka
    Shimizu, Kimiya
    Komatsu, Mari
    JOURNAL OF REFRACTIVE SURGERY, 2009, 25 (03) : 259 - 264
  • [42] Prediction of vaults in eyes with vertical implantable collamer lens implantation
    Shimada, Ryuichi
    Katagiri, Satoshi
    Horiguchi, Hiroshi
    Nakano, Tadashi
    Kitazawa, Yoshihiro
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2025, 51 (01): : 45 - 52
  • [43] The effects of implantable collamer lens implantation on higher order aberrations
    Belma Kayhan
    Efekan Coskunseven
    Onurcan Sahin
    Ioannis Pallikaris
    International Journal of Ophthalmology, 2019, (12) : 1848 - 1852
  • [44] The effects of implantable collamer lens implantation on higher order aberrations
    Kayhan, Belma
    Coskunseven, Efekan
    Sahin, Onurcan
    Pallikaris, Ioannis
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2019, 12 (12) : 1848 - 1852
  • [45] Comment on: Safety of phakic intraocular collamer lens implantation in 95 highly myopic special-needs children
    Ceran, Basak Bostanci
    Yusuf, Imran H.
    Karim, Rushmia
    Tassignon, Marie-Jose
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (12): : 1605 - 1606
  • [46] Time Course of Accommodation After Implantable Collamer Lens Implantation
    Kamiya, Kazutaka
    Shimizu, Kimiya
    Aizawa, Daisuke
    Ishikawa, Hitoshi
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 146 (05) : 674 - 678
  • [47] Case report: Spontaneous lens absorption after the implantation of an implantable collamer lens
    Liu, Xiaojing
    Zeng, Qiulin
    Lu, Ting
    Wang, Min
    Sun, Yong
    Zhang, Jingfa
    FRONTIERS IN MEDICINE, 2023, 10
  • [48] Intraocular Pressure After Implantation of a Phakic Collamer Intraocular Lens With a Central Hole
    Rodriguez-Una, Ignacio
    Pablo Rodriguez-Calvo, Pedro
    Fernandez-Vega Cueto, Luis
    Lisa, Carlos
    Fernandez-Vega Cueto, Andres
    F. Alfonso, Jose
    JOURNAL OF REFRACTIVE SURGERY, 2017, 33 (04) : 244 - +
  • [49] Cataract development after implantation of the Staar Collamer posterior chamber phakic lens
    Fink, AM
    Gore, C
    Rosen, E
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1999, 25 (02): : 278 - 282
  • [50] Influencing factors comparing different vault groups after phakic implantable collamer lens implantation: review and meta-analysis
    Pengcheng Zhang
    Chenjun Guo
    Song Wang
    Wenshan Jiang
    Dan Wang
    Hong Yan
    BMC Ophthalmology, 24