First-in-human evaluation of a hand-held automated venipuncture device for rapid venous blood draws

被引:26
|
作者
Leipheimer, Josh M. [1 ]
Balter, Max L. [1 ]
Chen, Alvin, I [1 ]
Pantin, Enrique J. [2 ]
Davidovich, Alexander E. [3 ]
Labazzo, Kristen S. [1 ]
Yarmush, Martin L. [1 ]
机构
[1] Rutgers State Univ, Dept Biomed Engn, Piscataway, NJ 08854 USA
[2] Robert Wood Johnson Univ Hosp, 1 Robert Wood Johnson Pl, New Brunswick, NJ 08901 USA
[3] Mt Sinai Hosp, Icahn Sch Med, 1 Gustave L Levy Pl, New York, NY 10029 USA
来源
TECHNOLOGY | 2019年 / 7卷 / 3-4期
关键词
Medical Device; Robotics; Image-Guidance; Ultrasound; Vascular Access; Computer Vision; Machine Learning; NEEDLESTICK INJURIES; RISK-FACTORS; EMERGENCY-DEPARTMENT; ACCESS DIFFICULTY; ULTRASOUND; CATHETERIZATION; PREVALENCE; FAILURE; SKIN;
D O I
10.1142/S2339547819500067
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Obtaining venous access for blood sampling or intravenous (IV) fluid delivery is an essential first step in patient care. However, success rates rely heavily on clinician experience and patient physiology. Difficulties in obtaining venous access result in missed sticks and injury to patients, and typically require alternative access pathways and additional personnel that lengthen procedure times, thereby creating unnecessary costs to healthcare facilities. Here, we present the first-in-human assessment of an automated robotic venipuncture device designed to safely perform blood draws on peripheral forearm veins. The device combines ultrasound imaging and miniaturized robotics to identify suitable vessels for cannulation and robotically guide an attached needle toward the lumen center. The device demonstrated results comparable to or exceeding that of clinical standards, with a success rate of 87% on all participants (n = 31), a 97% success rate on nondifficult venous access participants (n = 25), and an average procedure time of 93 +/- 30 s (n = 31). In the future, this device can be extended to other areas of vascular access such as IV catheterization, central venous access, dialysis, and arterial line placement.
引用
收藏
页码:98 / 107
页数:10
相关论文
共 50 条
  • [1] Fatigue evaluation of a hand-held device for the automated colonoscope
    Sheerer, Corey
    Choi, JungHun
    [J]. Medical Devices and Sensors, 2019, 2 (5-6):
  • [2] Evaluation of a coulometric hand-held blood glucose monitoring device
    Toffaletti, JG
    McDonnell, EH
    [J]. CLINICAL CHEMISTRY, 2005, 51 : A32 - A32
  • [3] A First Look at Mobile Hand-Held Device Traffic
    Maier, Gregor
    Schneider, Fabian
    Feldmann, Anja
    [J]. PASSIVE AND ACTIVE MEASUREMENT, PROCEEDINGS, 2010, 6032 : 161 - 170
  • [4] Diagnostics Hand-held device monitors blood clotting
    Dorey, Emma
    [J]. CHEMISTRY & INDUSTRY, 2011, (06) : 8 - 8
  • [5] POCKET PERMEAMETER - HAND-HELD DEVICE FOR RAPID MEASUREMENT OF PERMEABILITY
    BRADLEY, JS
    DUSCHATK.RW
    HINCH, HH
    [J]. AMERICAN ASSOCIATION OF PETROLEUM GEOLOGISTS BULLETIN, 1972, 56 (03): : 568 - &
  • [6] Evaluation of an electronic hand-held device for measurement of β-hydroxybutyrate in ewes
    Pichler, M.
    Gasteiner, J.
    Schwendenwein, I.
    Damberger, A.
    Drillich, M.
    Iwersen, M.
    [J]. REPRODUCTION IN DOMESTIC ANIMALS, 2012, 47 : 69 - 69
  • [7] EVALUATION OF MECHANIZED FLOSSING DEVICE COMPARED TO HAND-HELD FLOSSING
    BARTON, R
    DIAMOND, B
    [J]. JOURNAL OF DENTAL RESEARCH, 1980, 59 : 537 - 537
  • [8] A HAND-HELD DEVICE FOR THE RAPID MEASUREMENT OF WOOL-STAPLE THICKNESS
    GORDON, AJ
    PALLISTER, E
    [J]. JOURNAL OF THE TEXTILE INSTITUTE, 1981, 72 (03) : 145 - 146
  • [9] HAND-HELD DEVICE FOR RAPID PRODUCTION OF THIN SLICES OF FRESH TISSUE
    RINK, TJ
    SMITH, W
    [J]. JOURNAL OF PHYSIOLOGY-LONDON, 1976, 254 (01): : P5 - P6
  • [10] Rapid, hand-held paper diagnostic for measuring Fibrinogen Concentration in blood
    Bialkower, Marek
    McLiesh, Heather
    Manderson, Clare A.
    Tabor, Rico F.
    Garnier, Gil
    [J]. ANALYTICA CHIMICA ACTA, 2020, 1102 : 72 - 83