Mercury speciation by coupling cold vapour atomic absorption spectrometry with flow injection on-line preconcentration and liquid chromatographic separation

被引:66
|
作者
Yin, XF
Frech, W
Hoffmann, E
Ludke, C
Skole, J
机构
[1] Inst Berlin, Inst Spektrochem & Angew Spektroskopie, D-12489 Berlin, Germany
[2] Zhejiang Univ, Dept Chem, Hangzhou 310027, Peoples R China
[3] Umea Univ, Dept Analyt Chem, S-90187 Umea, Sweden
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1998年 / 361卷 / 08期
基金
中国国家自然科学基金;
关键词
D O I
10.1007/s002160051012
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A fully automated system for the direct determination of methylmercury (MeHg), ethylmercury (EtHg), phenylmercury (PhHg), and inorganic mercury (Hg(II)) at the ng/L level is described. It is based on solid phase extraction preconcentration incorporated in a flow injection (FI) system, high performance liquid chromatography (HPLC) separation, reduction combined with thermolysis and determination by cold vapour atomic absorption spectrometry (CVAAS). For preconcentration a microcolumn of bonded silica with octadecyl functional groups (C18 reversed phase material) was used as a sorbent for the mercury complexes formed on-line with ammonium pyrrolidine dithiocarbamate. Retained mercury species are eluted with a methanol-acetonitrile-water mixture and subjected to separation on an octadecylsilane (ODS) column before determination by CVAAS. The sensitivity of organo-mercury determination could be improved by using NaBH4 as a reductant combined with a thermolysis step. In order to perform on-line measurements the preconcentration microcolumn was mounted in a pressure-tight casing. Limits of detection for MeHg, EtHg, PhHg and Hg(II) employing a sample volume of 58.5 mL were 9, 6, 10 and 5 ng/L, respectively. The relative standard deviation (RSD) calculated from 9 repeated measurements was found to be 3.6%, 5.5%, 10.4% and 7.6% for MeHg, EtHg, PhHg and HE(II), respectively. Finally, the application of this method for speciation of mercury in fish and human urine is described.
引用
收藏
页码:761 / 766
页数:6
相关论文
共 50 条
  • [11] On-line speciation of mercury and methylmercury by cold vapour atomic absorption spectrometry using selective solid phase extraction
    Zachariadis, GA
    Anthemidis, AN
    Daftsis, EI
    Stratis, JA
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2005, 20 (01) : 63 - 65
  • [12] Mercury speciation in environmental samples by cold vapour atomic absorption spectrometry with in situ preconcentration on a gold trap
    Paola Rivaro
    Carmela Ianni
    Francesco Soggia
    Roberto Frache
    Microchimica Acta, 2007, 158 : 345 - 352
  • [13] Mercury speciation in environmental samples by cold vapour atomic absorption spectrometry with in situ preconcentration on a gold trap
    Rivaro, Paola
    Ianni, Carmela
    Soggia, Francesco
    Frache, Roberto
    MICROCHIMICA ACTA, 2007, 158 (3-4) : 345 - 352
  • [14] Flow injection on-line sorption preconcentration and separation coupled with atomic (mass) spectrometry for trace element (speciation) analysis
    Yan, XP
    ACTA CHIMICA SINICA, 2001, 59 (07) : 987 - 993
  • [15] On-line preconcentration and speciation of arsenic by flow injection hydride generation atomic absorption spectrophotometry
    Narcise, CIS
    Coo, LD
    del Mundo, FR
    TALANTA, 2005, 68 (02) : 298 - 304
  • [16] On-line UV photoreduction in a flow-injection/stopped-flow manifold for determination of mercury by cold vapour-atomic absorption spectrometry
    Gil, Sandra
    Costas, Marta
    Pena, Franciso
    De La Calle, Inmaculada
    Cabaleiro, Noelia
    Lavilla, Isela
    Bendicho, Carlos
    ANALYTICAL METHODS, 2010, 2 (11) : 1798 - 1802
  • [17] Speciation and determination of thallium by on-line microcolumn separation/preconcentration by flow injection-flame atomic absorption spectrometry using immobilized oxine as sorbent
    Dadfarnia, S.
    Assadollahi, T.
    Shabani, A. M. Haji
    JOURNAL OF HAZARDOUS MATERIALS, 2007, 148 (1-2) : 446 - 452
  • [18] On-line coupling of flow injection microcolumn separation and preconcentration to electrothermal atomic absorption spectrometry for determination of (ultra)trace selenite and selenate in water
    Yan, XP
    Sperling, M
    Welz, B
    ANALYTICAL CHEMISTRY, 1999, 71 (19) : 4353 - 4360
  • [19] Applications of on-line separation and preconcentration techniques in atomic absorption spectrometry
    Xu, SK
    Fang, ZL
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 27 (07) : 845 - 854
  • [20] Applications of flow injection knotted reactor on-line separation and preconcentration coupled with atomic spectrometry
    Wu Hong
    Jin Yan
    Tian Ye
    Bi Shu-Ping
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2007, 35 (06) : 905 - 911