共 50 条
Tungsten coil devices in atomic spectrometry: Absorption, fluorescence, and emission
被引:46
|作者:
Hou, XD
Levine, KE
Salido, A
Jones, BT
[1
]
Ezer, M
Elwood, S
Simeonsson, JB
机构:
[1] Wake Forest Univ, Dept Chem, Winston Salem, NC 27109 USA
[2] Univ Iowa, Dept Chem, Iowa City, IA 52242 USA
关键词:
D O I:
10.2116/analsci.17.175
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
In this work, tungsten coil (W-Coil) devices are used as atomizers for electrothermal atomization atomic absorption spectrometry (ETAAS), electrothermal atomization laser excited atomic fluorescence spectrometry (ETA-LEAFS), and electrothermal vaporization inductively coupled plasma atomic emission spectrometry (ETV-ICP-AES). For most cases in ETAAS and ETA-LEAFS, limits of detection (LODs) using the W-Coil are within a factor of ten of those observed with commercial graphite furnace systems. LOD for Cd by W-Coil AAS is 10 pg, while LODs for As, Se, Cr, Sb and Pb by W-Coil LEAFS are 950, 320, 1400, 330, and 160 fg, respectively. The compact W-Coil device makes it an ideal atomizer for portable atomic spectrometry instrumentation, especially when coupled with a miniature charge coupled device spectrometer. Alternatively, the atomizer can be used as an inexpensive, modular add-on to an existing commercial ICP-AES system; and the thermal separation of Pb with interference elements Al, Mn, and Fe is demonstrated.
引用
收藏
页码:175 / 180
页数:6
相关论文