Biomechanical evaluation of the tension band wiring principle. A comparison between two different techniques for transverse patella fracture fixation

被引:61
|
作者
Zderic, Ivan [1 ]
Stoffel, Karl [2 ,3 ]
Sommer, Christoph [4 ]
Hoentzsch, Dankward [5 ]
Gueorguiev, Boyko [1 ]
机构
[1] AO Res Inst Davos, Clavadelerstr 8, CH-7270 Davos, Switzerland
[2] Cantonal Hosp Baselland, Bruderholz, Switzerland
[3] Univ Basel, Basel, Switzerland
[4] Cantonal Hosp Graubuenden, Chur, Switzerland
[5] BG Klin Tubingen, Tubingen, Germany
关键词
Transverse patella fracture; Tension band wiring; Kirschner wire; Cannulated screw; Biomechanics; Interfragmentary pressure; OLECRANON FRACTURES; OSTEOSYNTHESIS;
D O I
10.1016/j.injury.2017.05.037
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Purpose: The aim of this study was to investigate the validity of the dynamic compression principle of tension band wiring in two techniques for patella fracture treatment. Methods: Twelve human cadaveric knees with simulated transverse patella fractures were assigned to two groups for treatment with tension band wiring using either Kirschner (K-) wires or cannulated screws. Biomechanical testing was performed over three knee movement cycles between 90 degrees flexion and 0 degrees full extension. Pressure distribution in the fracture gap and fracture site displacement were evaluated at the 3rd cycle in 15 degrees steps, namely 90 degrees -75 degrees -60 degrees -45 degrees -30 degrees -15 degrees -0 degrees extension phase and 0 degrees -15 degrees -30 degrees -45 degrees 60 degrees -75 degrees -90 degrees flexion phase. Results: Mean anterior / posterior interfragmentary pressure in the groups with K-wires and cannulated screws ranged within 0.16-0.40 MPa / 0.12-0.35 MPa and 0.37-0.59 MPa / 0.10-0.30 MPa, respectively. These changes remained non-significant for both groups and loading phases (P >= 0.171). Mean anterior / posterior fracture site displacement for K-wires and cannulated screws ranged within -0.01-0.53 mm / 0.11-0.74 mm and 0.11-0.55 mm / -0.10-0.50 mm, respectively. Anterior displacement remained without significant changes for both groups and loading phases (P >= 0.112). However, posterior displacement underwent a significant increase in the course of knee extension for K-wires (P <= 0.047), but not for cannulated screws (P >= 0.202). Significantly smaller displacement at the posterior fracture site was detected in the group with cannulated screws compared to K-wires at 60 degrees and 75 degrees extension phase (P <= 0.017), as well as at 45 degrees, 60 degrees and 75 degrees flexion phase (P <= 0.018). The critical value of 2 mm displacement at the posterior fracture site was not reached for any specimen and fixation technique. Knee extension was accompanied by synchronous increase in quadriceps pulling force. Conclusions: Tension band wiring fulfills from a biomechanical perspective the requirements for sufficient stability of transverse patella fracture fixation. It should, however, rather be considered as a static fixation principle than a dynamic one. Tension band wiring with cannulated screws was found advantageous over Kirschner wires in terms of interfragmentary movements at the posterior fracture site. (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1749 / 1757
页数:9
相关论文
共 50 条
  • [21] Medial malleolus fractures: A biomechanical comparison of tension band wiring fixation methods
    Uygur, Esat
    Poyanli, Oguz
    Mutlu, Ibrahim
    Celik, Talip
    Akpinar, Fuat
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2018, 104 (08) : 1259 - 1263
  • [22] Biomechanical evaluation of three fixation methods in a feline transverse patella fracture model
    Lee, Min-Yeong
    Ahn, Tae-Hwan
    Son, Sung-Il
    Kim, Hwi-Yool
    JOURNAL OF FELINE MEDICINE AND SURGERY, 2023, 25 (05)
  • [23] New Intramedullary Locking Nail for Olecranon Fracture Fixation-An In Vitro Biomechanical Comparison With Tension Band Wiring
    Nowak, Tobias E.
    Burkhart, Klaus J.
    Mueller, Lars P.
    Mattyasovszky, Stefan G.
    Andres, Torsten
    Sternstein, Werner
    Rommens, Pol M.
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2010, 69 (05): : E56 - E61
  • [24] Internal Fixation of Transverse Patella Fractures Using Cannulated Cancellous Screws with Anterior Tension Band Wiring
    Khan, I
    Dar, M. Y.
    Rashid, S.
    Butt, M. F.
    MALAYSIAN ORTHOPAEDIC JOURNAL, 2016, 10 (02) : 21 - 26
  • [25] A biomechanical comparison of the Rogers interspinous and the Lovely-Carl tension band wiring techniques for fixation of the cervical spine
    Brasil, AVB
    Coelho, DG
    Eloy, T
    Braga, FM
    JOURNAL OF NEUROSURGERY, 2000, 93 (01) : 109 - 116
  • [26] Transverse patella fracture fixation: A cadaveric biomechanical comparison of cannulated screws and anterior tension band versus low-profile, multiplanar mesh plating
    Garner, Matthew R.
    Homcha, Brittany
    Cowman, Trevin
    Goss, Madison
    Reid, J. Spence
    Lewis, Gregory S.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2024, 55 (06):
  • [27] Tension band wire fixation for anterior cruciate ligament avulsion fracture: biomechanical comparison of four fixation techniques
    Gan, Yudong
    Xu, Dachuan
    Ding, Jing
    Xu, Yongqing
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (05) : 909 - 915
  • [28] Tension band wire fixation for anterior cruciate ligament avulsion fracture: biomechanical comparison of four fixation techniques
    Yudong Gan
    Dachuan Xu
    Jing Ding
    Yongqing Xu
    Knee Surgery, Sports Traumatology, Arthroscopy, 2012, 20 : 909 - 915
  • [29] Biomechanical comparison of different techniques in fixation of simple transverse patella fractures- Is gold standard still valid?
    Veeresh, Vivek
    Sahoo, Bismaya
    Chowdhury, Buddhadev
    Mittal, Samarth
    Trikha, Vivek
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2022, 53 (07): : 2643 - 2643
  • [30] PERCUTANEOUS TENSION BAND WIRING - A NEW METHOD OF INTERNAL-FIXATION FOR MILDLY DISPLACED PATELLA FRACTURE
    LEUNG, PC
    MAK, KH
    LEE, SY
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1983, 23 (01): : 62 - 64