A two-dose-rate method for general recombination correction for liquid ionization chambers in pulsed beams

被引:31
|
作者
Tolli, Heikki [1 ]
Sjogren, Rickard [1 ]
Wendelsten, Mikael [1 ]
机构
[1] Umea Univ, Dept Radiat Sci, SE-90185 Umea, Sweden
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2010年 / 55卷 / 15期
关键词
COLLECTION EFFICIENCY; RADIATION; TETRAMETHYLSILANE; ISOOCTANE;
D O I
10.1088/0031-9155/55/15/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The correction for general recombination losses in liquid ionization chambers (LICs) is more complex than that in air-filled ionization chambers. The reason for this is that the saturation charge in LICs, i.e. the charge that escapes initial recombination, depends on the applied voltage. This paper presents a method, based on measurements at two different dose rates in a pulsed beam, for general recombination correction in LICs. The Boag theory for pulsed beams is used and the collection efficiency is determined by numerical methods which are equivalent to the two-voltage method used in dosimetry with air-filled ionization chambers. The method has been tested in experiments in water in a 20 MeV electron beam using two LICs filled with isooctane and tetramethylsilane. The dose per pulse in the electron beam was varied between 0.1 mGy/pulse and 8 mGy/pulse. The relative standard deviations of the collection efficiencies determined with the two-dose-rate method ranged between 0.1% and 1.5%. The dose-rate variations of the general recombination corrected charge measured with the LICs are in excellent agreement with the corresponding values obtained with an air-filled plane parallel ionization chamber.
引用
收藏
页码:4247 / 4260
页数:14
相关论文
共 42 条
  • [31] Experimental determination of the recombination correction factor kS for SNC 125c, SNC 350p and SNC 600c ionization chambers in pulsed photon beams
    Schoenfeld, Ann-Britt
    Schoenfeld, Andreas A.
    Looe, Hui Khee
    Poppe, Bjorn
    de Wilde, Rudy Leon
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2020, 30 (04): : 300 - 304
  • [32] Derivation of a formula describing the saturation correction of plane-parallel ionization chambers in pulsed fields with arbitrary repetition rate
    Karsch, Leonhard
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (08): : 3222 - 3236
  • [33] Use and Choice of Ionization Chambers for the Commissioning of Flattened and Flattening-Filter-Free Photon Beams: Determination of Recombination Correction Factor (ks)
    Stucchi, C.
    Bresolin, A.
    Mongioj, V.
    Carrara, M.
    Pignoli, E.
    Bonfantini, F.
    MEDICAL PHYSICS, 2014, 41 (06) : 371 - 371
  • [34] Ion recombination correction factors (Pion) for Varian TrueBeam high-dose-rate therapy beams
    Kry, Stephen F.
    Popple, Richard
    Molineu, Andrea
    Followill, David S.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (06): : 318 - 325
  • [35] Experimental analysis of general ion recombination in a liquid-filled ionization chamber in high-energy photon beams
    Chung, Eunah
    Davis, Stephen
    Seuntjens, Jan
    MEDICAL PHYSICS, 2013, 40 (06)
  • [36] Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations - perpendicular and parallel
    Casar, Bozidar
    Gershkevitsh, Eduard
    Mendez, Ignasi
    Jurkovic, Slaven
    Huq, M. Saiful
    MEDICAL PHYSICS, 2020, 47 (01) : 242 - 259
  • [37] Experimental derivation of wall correction factors for ionization chambers used in high dose rate 192Ir source calibration
    Maréchal, MH
    de Almeida, CE
    Ferreira, IH
    Sibata, CH
    MEDICAL PHYSICS, 2002, 29 (01) : 1 - 5
  • [38] Ion recombination correction factor in scanned light-ion beams for absolute dose measurement using plane-parallel ionisation chambers
    Rossomme, S.
    Horn, J.
    Brons, S.
    Jaekel, O.
    Mairani, A.
    Ciocca, M.
    Floquet, V.
    Romano, F.
    Garcia, D. Rodriguez
    Vynckier, S.
    Palmans, H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (13): : 5365 - 5382
  • [39] Long-term stability of liquid ionization chambers with regard to their qualification as local reference dosimeters for low dose-rate absorbed dose measurements in water
    Bahar-Gogani, J
    Grindborg, JE
    Johansson, BE
    Wickman, G
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03): : 729 - 740
  • [40] Fluence Beam Monitor for High-Intensity Particle Beams Based on a Multi-Gap Ionization Chamber and a Method for Ion Recombination Correction
    Giordanengo, Simona
    Guarachi, Leslie Fanola
    Braccini, Saverio
    Cirrone, Giuseppe A. P.
    Donetti, Marco
    Fausti, Federico
    Milian, Felix Mas
    Romano, Francesco
    Vignati, Anna
    Monaco, Vincenzo
    Cirio, Roberto
    Sacchi, Roberto
    APPLIED SCIENCES-BASEL, 2022, 12 (23):