Dual-Energy X-ray Imaging by Simultaneous Integration and Campbelling Readout

被引:0
|
作者
Roessl, Ewald [1 ]
Thran, Axel [1 ]
Martens, Gerhard [1 ]
Istel, Thomas [1 ]
Proksa, Roland [1 ]
Schlomka, Jens-Peter [2 ]
机构
[1] Philips Res Europe Hamburg, Digital Imaging, Rontgenstr 24-26, D-22335 Hamburg, Germany
[2] Philips Res Europe Aachen, X ray Imaging Syst, D-52066 Aachen, Germany
关键词
dual-energy; computed tomography; Campbell-mode readout; DETECTOR CIRCUITS;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Dual-energy imaging is based on the acquisition of two spectrally distinct attenuation measurements. The two most common techniques are the dual-kV technique and the dual-detector layer technique. More recently, another technique was introduced based on the simultaneous acquisition of photon-counting and current integration data (CIX) by means of a dedicated detector readout ASIC. All of the above methods have certain advantages and disadvantages. While depending on the particular realization (e. g., kVp switching, dual-tube systems in computed tomography), the dual-kV technique suffers from a time-lag between the two acquisitions and as a result from susceptibility to motion. The dual-layer technique usually offers less spectral separation and requires a dedicated detection system. The CIX technique suffers from a reduced dynamic range due to the limiting rate performance in the counting channel. In this paper we introduce yet another dual-energy technique, similar to the CIX, however, without the restrictions coming from the limited count rate. We propose dual-energy imaging based on simultaneous acquisition of the mean (DC) and variance (AC) components of the electrical detector signal, i.e., simultaneous integration-mode readout and Campbell-mode readout (fluctuation mode). The method is based on "Campbell's theorem" which states that the variance of the deviations from the mean detector signal is proportional to the second moment of the incoming x-ray energy spectrum. We compare in simulations the dual-energy performance of the proposed new technique with conventional dual-kV and dual-crystal techniques and present first experimental Campbell-mode CT images obtained from a scintillator crystal coupled to a photomultiplier tube. Moreover, we demonstrate the feasibility of separating iodine from calcium with the novel technique and compare it to the separability achieved with a dual-kV technique.
引用
收藏
页码:2112 / 2115
页数:4
相关论文
共 50 条
  • [21] DUAL-ENERGY X-RAY ABSORPTIOMETRY OF THE FOREARM
    ROOS, JC
    TENKROODE, K
    LIPS, P
    TEULE, GJJ
    VANLINGEN, A
    JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (05) : P110 - P111
  • [22] ACRONYM FOR DUAL-ENERGY X-RAY ABSORPTIOMETRY
    WILSON, CR
    COLLIER, BD
    CARRERA, GF
    JACOBSON, DR
    RADIOLOGY, 1990, 176 (03) : 875 - 875
  • [23] LIMITATIONS OF DUAL-ENERGY X-RAY ABSORPTIOMETRY
    TOTHILL, P
    NORD, RH
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1995, 61 (02): : 398 - 399
  • [24] REPRODUCIBILITY OF DUAL-ENERGY X-RAY ABSORPTIOMETRY
    CHILIBECK, P
    CALDER, A
    SALE, DG
    WEBBER, C
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 1994, 45 (04): : 297 - 302
  • [25] Dual-energy x-ray absorptiometry in obesity
    Bazzocchi, Alberto
    Diano, Danila
    CANADIAN MEDICAL ASSOCIATION JOURNAL, 2014, 186 (01) : 48 - 48
  • [26] An Update on Dual-Energy X-Ray Absorptiometry
    Blake, Glen M.
    Fogelman, Ignac
    SEMINARS IN NUCLEAR MEDICINE, 2010, 40 (01) : 62 - 73
  • [27] DUAL-ENERGY X-RAY ABSORPTIOMETRY IN THE NEWBORN
    SALLE, BL
    LAPILLONNE, A
    BRAILLON, P
    ARCHIVES FRANCAISES DE PEDIATRIE, 1993, 50 (07): : 545 - 546
  • [28] Dual-energy x-ray absorptiometry in osteopetrosis
    Adler, IN
    Stine, KC
    Kurtzburg, J
    Kinder, DL
    Cox, KS
    Lensing, SY
    Seibert, JJ
    SOUTHERN MEDICAL JOURNAL, 2000, 93 (05) : 501 - 503
  • [29] Overlapped objects discrimination using dual-energy, high energy X-ray imaging
    Li, Qinghua
    Li, Jianmin
    Wang, Xuewu
    Zhang, Li
    Kang, Kejun
    Zhong, Huaqiang
    Qinghua Daxue Xuebao/Journal of Tsinghua University, 2008, 48 (08): : 1256 - 1259
  • [30] X-RAY-IMAGING SENSOR USING CDTE CRYSTALS FOR DUAL-ENERGY X-RAY ABSORPTIOMETRY
    OHTSUCHI, T
    TSUTSUI, H
    OHMORI, K
    BABA, S
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (05) : 1740 - 1745