Design and evaluation of a nanometer aerosol differential mobility analyzer (Nano-DMA)

被引:262
|
作者
Chen, DR
Pui, DYH [1 ]
Hummes, D
Fissan, H
Quant, FR
Sem, GJ
机构
[1] Univ Minnesota, Dept Mech Engn, Particle Technol Lab, Minneapolis, MN 55455 USA
[2] Gerhard Mercator Univ, D-47048 Duisburg, Germany
[3] TSI Inc, St Paul, MN 55164 USA
基金
美国国家科学基金会;
关键词
D O I
10.1016/S0021-8502(97)10018-0
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
A nanometer aerosol differential mobility analyzer, Nano-DMA, has been developed for measuring the size distribution of nanometer aerosols in the particle size range of 3-50 nm. The design is based on a cylindrical configuration and is optimized by means of the numerical model of Chen and Pui (1997, J. Aerosol Sci. 28, 985-1004). Important design features include high particle penetration (low loss) through the Nano-DMA and high sizing resolution. For reducing particle loss in the aerosol transport passage, the aerosol residence time in the Nano-DMA is reduced by shortening the inlet transport passage. An optional feature of high inlet flow (up to 16.5 1pm) is designed in order to further reduce the residence lime between the aerosol inlet and the slit in the classifying region of the Nano-DMA. A new entrance slit is designed to have optimal aerosol and sheath flow matching at a flow ratio of 1:10, and has a wide dynamic flow-ratio range (up to aerosol/sheath flow ratio of 1/70) compared with the TSI-standard DMA design. This slit improvement makes the Nano-DMA suitable for high resolution particle sizing and classification. For reducing the effect of Brownian diffusion broadening on the transfer Function of the Nano-DMA, the collector tube length is shortened to 5.0 cm compared to the TSI-Standard DMA of 44.44 cm and TSI-Short DMA of 11.11 cm. At the design flow condition of 1.5 1pm aerosol (or 16.5 1pm aerosol high inlet flow case) and 15.0 1pm sheath flow rates, the measurable size range is from 3-50 nm. The lower detection limit of 3.0 nm coincides with the lower detection limit of the TSI UCPC (Ultrafine Condensation Particle Counter). The base of the Nano-DMA is completely re-designed to avoid particle loss due to the undesirable electrostatic effect observed by Kousaka et al. (1986, J. Chern. Eng. Japan 19, 401), and to obtain an uniform electric field in the entire classifying region. The overall performance of the Nano-DMA is then evaluated by the numerical model of Chen and Pui (1997, J. Aerosol Sci. 28, 985-1004) before its construction and experimental evaluation. By comparing with the experimental results obtained using the Tandem DMA technique described in Hummes et al. (1996, J. Aerosol Sci. 27, S135-S136; Part. Part. Syst. Charact. 5, 327-332), it is concluded that the Nano-DMA is performing well in the designed size range and its transfer function agrees well with the numerical prediction. (C) 1998 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:497 / 509
页数:13
相关论文
共 50 条
  • [1] Slip correction measurements of certified PSL nanoparticles using a nanometer differential mobility analyzer (nano-DMA) for Knudsen number from 0.5 to 83
    Kim, JH
    Mulholland, GW
    Kukuck, SR
    Pui, DYH
    [J]. JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 2005, 110 (01): : 31 - 54
  • [2] Experimental determination of the transfer function of a differential mobility analyzer (DMA) in the nanometer size range
    Hummes, D
    Stratmann, F
    Neumann, S
    Fissan, H
    [J]. PARTICLE & PARTICLE SYSTEMS CHARACTERIZATION, 1996, 13 (05) : 327 - 332
  • [3] Determination of Aerosol Particle Size Distribution using Electrical Differential Mobility Analyzer (DMA)
    Fasih-Ramandi, Fatemeh
    Sadigzadeh, Asghar
    Jafari, Mohammad-Javad
    Khodakarim, Soheila
    [J]. JOURNAL OF HEALTH AND SAFETY AT WORK, 2021, 11 (01) : 92 - 105
  • [4] A differential mobility analyzer (DMA) system for submicron aerosol measurements at ambient relative humidity
    Wang, J
    Flagan, RC
    Seinfeld, JH
    [J]. AEROSOL SCIENCE AND TECHNOLOGY, 2003, 37 (01) : 46 - 52
  • [5] Performance of nano-DMA operated with different gases for sheath and aerosol carrier flows
    Li, WL
    Chen, DR
    [J]. AEROSOL SCIENCE AND TECHNOLOGY, 2005, 39 (10) : 931 - 940
  • [6] Experimental tests of a nano-DMA with no voltage change between aerosol inlet and outlet slits
    Martinez-Lozano, P.
    Labowsky, M.
    de la Mora, J. Fernandez
    [J]. JOURNAL OF AEROSOL SCIENCE, 2006, 37 (11) : 1629 - 1642
  • [8] The Spider DMA: A miniature radial differential mobility analyzer
    Amanatidis, Stavros
    Kim, Changhyuk
    Spielman, Steven R.
    Lewis, Gregory S.
    Hering, Susanne, V
    Flagan, Richard C.
    [J]. AEROSOL SCIENCE AND TECHNOLOGY, 2020, 54 (02) : 175 - 189
  • [9] Differential mobility analyzer for nano particle sizing
    Fissan, H
    Hummes, D
    Pui, D
    [J]. SYNTHESIS AND PROCESSING OF NANOCRYSTALLINE POWDER, 1996, : 289 - 297
  • [10] Radial Differential Mobility Analyzer for One Nanometer Particle Classification
    Brunelli, N. A.
    Flagan, R. C.
    Giapis, K. P.
    [J]. AEROSOL SCIENCE AND TECHNOLOGY, 2009, 43 (01) : 53 - 59