Calibration and characterization of trap detector filter radiometers

被引:4
|
作者
Allen, DW [1 ]
Eppeldauer, GP [1 ]
Brown, SW [1 ]
Early, EA [1 ]
Johnson, BC [1 ]
Lykke, KR [1 ]
机构
[1] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
来源
EARTH OBSERVING SYSTEMS VIII | 2003年 / 5151卷
关键词
trap detector; radiometer; remote sensing;
D O I
10.1117/12.507995
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
We describe the development of a mechanically simple, radiometrically stable transfer radiometer designed for both radiance and irradiance measurements. The filter radiometer consists of a six-element Si trap detector, a temperature stabilized filter wheel with up to 5 filters, and two precision apertures in a Gershun tube arrangement. With the Gershun tube installed, the instrument operates in radiance mode; with the front aperture removed, in irradiance mode. Two trap detector filter radiometers have been designed and built by the National Institute of Standards and Technology (NIST) for use in remote sensing applications. The filter radiometers have been characterized for optical and electrical performance, and have been calibrated for responsivity using both narrow-band, tunable-laser-illuminated and broad-band, lamp-illuminated integrating sphere sources. This paper describes the filter radiometer design, characterization, and deployments for two remote sensing projects.
引用
收藏
页码:471 / 479
页数:9
相关论文
共 50 条
  • [1] Accurate calibration of filter radiometers against a cryogenic radiometer using a trap detector
    Friedrich, R
    Fischer, J
    Stock, M
    METROLOGIA, 1996, 32 (06) : 509 - 513
  • [2] Novel calibration method for filter radiometers
    Schrama, CA
    Reijn, H
    METROLOGIA, 1999, 36 (03) : 179 - 182
  • [3] Method for characterization of filter radiometers
    Toivanen, P
    Manoochehri, F
    Kärhä, P
    Ikonen, E
    Lassila, A
    APPLIED OPTICS, 1999, 38 (09) : 1709 - 1713
  • [4] Characterization and calibration of broadband ultraviolet radiometers
    Xu, G
    Huang, X
    METROLOGIA, 2000, 37 (03) : 235 - 242
  • [5] Research on calibration of radiance detector based on trap detector
    Zou, Peng
    Zheng, Xiaobing
    Wu, Haoyu
    Li, Jianjun
    Liu, Enchao
    Hong, Jin
    Xiong, Wei
    Guangxue Xuebao/Acta Optica Sinica, 2013, 33 (07):
  • [6] Calibration of a trap detector against a cryogenic radiometer
    Nield, KM
    Clare, JF
    Hamlin, JD
    Bittar, A
    METROLOGIA, 1998, 35 (04) : 581 - 586
  • [7] Calibration of a trap detector against a cryogenic radiometer
    Nield, K.M.
    Clare, J.F.
    Hamlin, J.D.
    Bittar, A.
    Metrologia, 35 (04): : 581 - 586
  • [8] Detector based calibration of solar UV radiometers in laboratory environments with high accuracy
    Durak, M
    San, SE
    Aslan, MH
    MEASUREMENT, 2005, 37 (03) : 213 - 217
  • [9] Intercomparison of characterization techniques of filter radiometers in the ultraviolet region
    Kärhä, P
    Harrison, NJ
    Nevas, S
    Hartree, WS
    Abu-Kassem, I
    METROLOGIA, 2003, 40 (01) : S50 - S54
  • [10] Calibration technique of spectral radiation for an infrared trap detector
    Zhang Lei
    Lin Zhi-Qiang
    JOURNAL OF INFRARED AND MILLIMETER WAVES, 2018, 37 (05) : 572 - 577