Comparing two Power Supplies for Fast-Field Cycling Nuclear Magnetic Resonance Relaxometers Power Losses and Performance

被引:0
|
作者
Lima, Marco [1 ,2 ]
Pereira, Bruno [1 ,2 ]
Sousa, Duarte M. [1 ,2 ]
Roque, Antonio [3 ,4 ]
Margato, Elmano [5 ,6 ]
机构
[1] Univ Lisbon, Inst Super Tecn, Dept Elect & Comp Engn, Lisbon, Portugal
[2] Univ Lisbon, INESC ID, Lisbon, Portugal
[3] ESTSetubal Inst Politecn Setubal, Dept Elect Engn, Setubal, Portugal
[4] INESC ID, Setubal, Portugal
[5] ISEL, Ctr Electrotecnia & Elect Ind, Lisbon, Portugal
[6] INESC ID, Lisbon, Portugal
关键词
power supply; power losses; fast field cycling NMR; DC/DC converter; NMR;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
The main power supply of a Fast Field-Cycling Nuclear Magnetic Resonance (FFC-NMR) is the key element comparing the performance of different solutions. The power supply is a current source that supplies a magnet being the current controlled in order to perform adjustable and repetitive current cycles. This power supply can be based on different topologies, operating principles and controlled using distinct techniques. If for the final users of this experimental technique the current cycles of the equipment is the core feature, for the developers also the power losses distribution needs to be analyzed in order to develop efficient solutions. In this paper, the power losses and the dynamic behavior of two solutions for the FFC-NMR power supply are compared and discussed.
引用
收藏
页码:546 / 551
页数:6
相关论文
共 33 条