共 48 条
Use of disposable graduated biopsy forceps improves accuracy of polyp size measurements during endoscopy
被引:23
|作者:
Jin, Hei-Ying
[1
]
Leng, Qiang
[1
]
机构:
[1] Nanjing Univ Tradit Chinese Med, Affiliated Hosp 3, Natl Ctr Colorectal Surg, Jiangsu Integrated Colorectal Oncol Ctr, Nanjing 210001, Jiangsu, Peoples R China
关键词:
Disposable graduated biopsy forceps;
Polyp size measurement;
Colonoscopy;
Accuracy;
LATERALLY SPREADING TUMORS;
COLORECTAL-CANCER;
SURVEILLANCE;
DIAGNOSIS;
D O I:
10.3748/wjg.v21.i2.623
中图分类号:
R57 [消化系及腹部疾病];
学科分类号:
摘要:
AIM: To determine the accuracy of endoscopic polyp size measurements using disposable graduated biopsy forceps (DGBF). METHODS: Gradations accurate to 1 mm were assessed with the wire of disposable graduated biopsy forceps. When a polyp was noted, endoscopists determined the width of the polyp; then, the graduated biopsy forceps was inserted and the largest diameter of the tumor was measured. After excision, during surgery or endoscopy, the polyp was measured using the vernier caliper. RESULTS: One hundred and thirty-three colorectal polyps from 119 patients were studied. The mean diameter, by post-polypectomy measurement, was 0.92 +/- 0.69 cm; 83 were < 1 cm, 36 were between 1 and 2 cm, and 14 were > 2 cm. The mean diameter, by visual estimation, was 1.15 +/- 0.88 cm; compared to the actual size measured using vernier calipers, the difference was statistically significant. The mean diameter measured using the DGBF was 0.93 +/- 0.68 cm; compared to the actual size measured using vernier calipers, this difference was not statistically significant. The ratio between the mean size estimated by visual estimation and the actual size was significantly different from that between the mean size estimated using the DGBF and the actual size (1.26 +/- 0.30 vs 1.02 +/- 0.11). CONCLUSION: The accuracy of polyp size estimation was low by visual assessment; however, it improved when the DGBF was used.
引用
收藏
页码:623 / 628
页数:6
相关论文