Set-up errors in radiotherapy for oesophageal cancers - Is electronic portal imaging or conebeam more accurate?

被引:37
|
作者
Hawkins, Maria A. [1 ]
Aitken, Alexandra [1 ]
Hansen, Vibeke N.
McNair, Helen A. [1 ]
Tait, Diana M. [1 ]
机构
[1] Royal Marsden Hosp, NHS Fdn Trust, Dept Radiotherapy, Sutton SM2 5PT, Surrey, England
关键词
Cone beam computed tomography; Image guided radiotherapy; Oesophageal cancer; Set-up errors; Image registration; BEAM COMPUTED-TOMOGRAPHY; CELL LUNG-CANCER; ADAPTIVE RADIOTHERAPY; MARGINS; CARCINOMA; MOTION; TUMOR; CT; CHEMORADIOTHERAPY; POPULATION;
D O I
10.1016/j.radonc.2010.11.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To compare kV computed tomography (CBCT) with electronic portal imaging (EPI) and evaluate set-up variations in the anterior-posterior (AP), right-left (LR) and cranio-caudal (CC) directions and rotational variations: pitch, roll, and yaw, for oesophageal cancer patients treated with radical radiotherapy. Methods and materials: Twenty patients with locally advanced oesophageal cancer treated with chemoradiation were consented for this prospective ethics approved protocol. Patients were positioned using skin marks/tattoos, kV-CBCT scans (XVI) and EPI's were performed prior to treatment and registered to the planning CT scans and digitally reconstructed radiographs, respectively. XVI data was used to adjust patient setups before treatment delivery. A total of 122 EPI pairs and 207 CBCT scans were analysed. The systematic and random errors were calculated. Results: The systematic and random errors (mm) for XVI were 1.3, 1.7, 1.4 and 2.6, 3.9, 2.0 in RL, CC and AP direction, respectively, with EPI of similar magnitude. There was no correlation between the 2 modalities of imaging as 31.7% of all image pairs were discordant >3 mm and 12.5% >5 mm. XVI identified rotations >3 degrees in 44 images. Conclusions: EPI results in different position correction for verification of radiotherapy in oesophageal malignancies when compared with CBCT. CBCT verification offers adequate 3D volumetric image quality to improve the accuracy of treatment delivery for oesophageal malignancies in radiotherapy and should be used for image guidance. (C) 2010 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 98 (2011) 249-254
引用
收藏
页码:249 / 254
页数:6
相关论文
共 50 条
  • [21] Registration of electronic portal images for patient set-up verification
    Matsopoulos, GK
    Asvestas, PA
    Delibasis, KK
    Kouloulias, V
    Uzunoglu, N
    Karaiskos, P
    Sandilos, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (14): : 3279 - 3289
  • [22] Surface imaging, portal imaging, and skin marker set-up vs. CBCT for radiotherapy of the thorax and pelvis
    Pallotta, Stefania
    Vanzi, Eleonora
    Simontacchi, Gabriele
    Marrazzo, Livia
    Ceroti, Marco
    Paiar, Fabiola
    Livi, Lorenzo
    Bucciolini, Marta
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 (09) : 726 - 733
  • [23] EVALUATION OF THREE DIMENSIONAL SET-UP ERRORS IN 3D-CONFORMAL RADIOTHERAPY TO THE PROSTATE AT ST LUKE?S CANCER CENTRE USING ELECTRONIC PORTAL IMAGING.
    Alexander, E.
    Freeman, K.
    Urbano, T. Guerrero
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S432 - S433
  • [24] The Influence of Time of Day On Set-Up Errors in External Beam Radiotherapy
    Valakh, V.
    Day, E.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [25] Comparing Electronic Portal Imaging and Megavoltage Conebeam CT for Positioning Head and Neck Radiotherapy Patients
    Chen, J.
    Lambert, L.
    Weinberg, V.
    Aubin, M.
    Quivey, J.
    Pouliot, J.
    Yom, S. S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S785 - S785
  • [26] Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer
    Erridge, SC
    Seppenwoolde, Y
    Muller, SH
    van Herk, M
    De Jaeger, K
    Belderbos, JSA
    Boersma, LJ
    Lebesque, JV
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 66 (01) : 75 - 85
  • [27] ACCURACY IN TANGENTIAL BREAST TREATMENT SET-UP - A PORTAL IMAGING STUDY
    VANTIENHOVEN, G
    LANSON, JH
    CRABEELS, D
    HEUKELOM, S
    MIJNHEER, BJ
    [J]. RADIOTHERAPY AND ONCOLOGY, 1991, 22 (04) : 317 - 322
  • [28] Construction of Prediction Model of Radiotherapy Set-Up Errors in Patients with Lung Cancer
    Yang, Fan
    Li, Xinxia
    [J]. BIOMED RESEARCH INTERNATIONAL, 2022, 2022
  • [29] The effects of set-up errors on dose distribution in radiotherapy treatment for lung cancer
    Cao, X.
    Liu, M.
    Zhai, F.
    Li, N.
    Bao, C.
    Liu, Y.
    Chen, G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (03): : 515 - 520
  • [30] INFLUENCE OF DAILY SET-UP ERRORS ON DOSE DISTRIBUTION DURING PELVIS RADIOTHERAPY
    Kasabasic, Mladen
    Rajevac, Vedran
    Jurkovic, Slaven
    Ivkovic, Ana
    Sobat, Hrvoje
    Faj, Dario
    [J]. ARHIV ZA HIGIJENU RADA I TOKSIKOLOGIJU-ARCHIVES OF INDUSTRIAL HYGIENE AND TOXICOLOGY, 2011, 62 (03): : 261 - 267