Radiation doses to patients and cardiologists from permanent cardiac pacemaker implantation procedures

被引:28
|
作者
Tsalafoutas, IA
Spanodimos, SG
Maniatis, PN
Fournarakis, GM
Koulentianos, ED
Tsigas, DL
机构
[1] Konstantopoulio Angia Olga Hosp, Med Phys Unit, Athens, Greece
[2] Konstantopoulio Angia Olga Hosp, Dept Cardiol, Athens, Greece
[3] Konstantopoulio Angia Olga Hosp, Comp Tomog Dept, Athens, Greece
[4] Konstantopoulio Angia Olga Hosp, Dept Radiol, Athens, Greece
来源
关键词
pacemaker; radiation doses; dose-area product; entrance skin dose;
D O I
10.1111/j.1540-8159.2005.00194.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Radiation Doses to Patients and Cardiologists from Permanent Cardiac Pacemaker Implantation Procedures. Background: The purpose of this study was to determine the patient and cardiologist doses during the implantation of permanent cardiac pacemakers under fluoroscopic control. Methods: For 55 procedures concerning three different types of pacemakers (DDD, VDD, and VVI), the dose-area product (DAP) meter readings and fluoroscopy times were recorded. From these data, the dose to the operating cardiologist was estimated. Results: The median values of DAP and fluoroscopy time for all the procedures monitored were 11.4 Gycm(2) and 6.6 minute, respectively. For the 22 DDD, 21 VDD, and 12 VVI pacemakers implanted, the respective DAP median values were 14.7, 9.9, and 7.3 Gycm(2) and the respective median fluoroscopy times were 8.4, 5, and 2.9 minutes. The median doses to the hands, chest, eyes, and legs of the cardiologist conducting the manipulations were estimated to 0.21, 0.06, 0.03, and 0.11 mGy, respectively, per procedure. Conclusions: Compared to the existing literature, the median DAP value of this study is almost identical to the 11.2 Gycm(2) reported from a sample of 627 patients in 17 different x-ray rooms, whereas the fluoroscopy times ore within the range of values reported by other authors. Concerning the cardiologist exposure, the estimated values indicate that the implantation of pacemakers is a procedure that does not involve a severe risk, especially if it is taken into account that lead aprons and collars ore routinely used.
引用
收藏
页码:910 / 916
页数:7
相关论文
共 50 条
  • [1] PREDICTORS OF CARDIAC TAMPONADE FROM PERMANENT PACEMAKER IMPLANTATION: ANALYSIS OF 1,014,713 PROCEDURES
    Dolmatova, Elena
    Moazzami, Kasra
    Mazza, Victor
    Klapholz, Marc
    Waller, Alfonso
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (13) : 706 - 706
  • [2] Pacemaker implantation and permanent cardiac electrostimalation in dogs
    Hervé, D
    Troger, JC
    PRATIQUE MEDICALE ET CHIRURGICALE DE L ANIMAL DE COMPAGNIE, 2003, 38 (02): : 151 - 160
  • [3] PERMANENT CARDIAC-PACEMAKER IMPLANTATION IN THE ELDERLY
    BELOTT, PH
    CARDIOLOGY IN THE ELDERLY, 1994, 2 (02): : 191 - 201
  • [4] Trends in Permanent Pacemaker Implantation in the United States From 1993 to 2009 Increasing Complexity of Patients and Procedures
    Greenspon, Arnold J.
    Patel, Jasmine D.
    Lau, Edmund
    Ochoa, Jorge A.
    Frisch, Daniel R.
    Ho, Reginald T.
    Pavri, Behzad B.
    Kurtz, Steven M.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (16) : 1540 - 1545
  • [5] A study on patients' radiation doses from interventional cardiac procedures in Tabriz, Iran
    Mesbahi, Asghar
    Aslanabadi, Naser
    RADIATION PROTECTION DOSIMETRY, 2008, 132 (04) : 375 - 380
  • [6] Metoprolol Improves Myocardial Remodeling and Cardiac Function in Patients with Permanent Pacemaker Implantation
    Ye, Li
    Hu, Guofen
    Yu, Huamin
    Sun, Jindong
    Yuan, Hong
    JOURNAL OF HEALTHCARE ENGINEERING, 2022, 2022
  • [7] PERMANENT CARDIAC-PACEMAKER IMPLANTATION IN THE FETAL LAMB
    SCAGLIOTTI, D
    PRINGLE, KC
    GALLASTEGUI, JL
    SHIMOKOCHI, DD
    CLINICAL RESEARCH, 1985, 33 (04): : A817 - A817
  • [8] The issue of permanent pacemaker implantation following cardiac surgery
    Toth Roland
    Tahin Tamas
    Riba Adam
    Rashed Aref
    ORVOSI HETILAP, 2020, 161 (31) : 1271 - 1280
  • [9] Risk factors of pacing dependence and cardiac dysfunction in patients with permanent pacemaker implantation
    Yu, Ziqing
    Liang, Yixiu
    Xiao, Zilong
    Wang, Yucheng
    Bao, Pei
    Zhang, Chunyu
    Su, Enyong
    Li, Minghui
    Chen, Xueying
    Qin, Shengmei
    Chen, Ruizhen
    Su, Yangang
    Ge, Junbo
    ESC HEART FAILURE, 2022, 9 (04): : 2325 - 2335
  • [10] Radiation doses in cardiac angiographic procedures
    Rehani, M
    Kunar, B
    Bahl, V
    Manchanda, S
    Rajani, M
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2474 - 2476