Laboratory and field validations of a solid-phase microextraction device for the determination of ethylene oxide

被引:14
|
作者
Tsai, SW
Tsai, ST
Wang, VS
Lai, JS
机构
[1] China Med Univ, Dept Occupat Safety & Hlth, Taichung, Taiwan
[2] China Med Univ, Inst Environm Med, Taichung, Taiwan
关键词
sample preparation; air analysis; environmental analysis; instrumentation; ethylene oxide; hydrogen bromide;
D O I
10.1016/j.chroma.2003.10.088
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Laboratory and field evaluations were performed to validate a solid-phase microextraction (SPME) device that was used as a diffusive sampler. Hydrogen bromide (HBr) was loaded onto the carboxen-polydimethylsiloxane (CAR-PDMS) fiber for the determination of ethylene oxide (EtO) with on-fiber derivatization. For laboratory evaluations, known concentrations of ethylene oxide around the threshold limit values (TLV)/time-weighted average and specific relative humidities (RHs) were generated by syringe pumps in a dynamic generation system. The SPME diffusive samplers and the commercially available 3M 3551 passive monitors were placed side-by-side in an exposure chamber which was designed to allow measurement of face velocities, temperatures, exposing vapor concentrations, and RHs. Field validations with both SPME diffusive sampler and 3M 3551 passive monitors were also performed. The correlations between the results from both SPME device and 3M 3551 passive monitor were found to be linear (r > 0.9699) and consistent (slope congruent to 1.12 +/- 0.07). However, the variations of diffusion coefficients at different temperatures needs to be considered and the adjustment of sampling constant was a must when sampling at temperatures different from 25degreesC. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:25 / 30
页数:6
相关论文
共 50 条