Inverse planning anatomy-based dose optimization for HDR-brachytherapy of the prostate using fast simulated annealing algorithm and dedicated objective function

被引:197
|
作者
Lessard, E
Pouliot, J
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA 94143 USA
[2] Univ Laval, Ctr Rech Cancerol, Ctr Hosp Univ Quebec, Hotel Dieu Quebec,Serv Radiooncol, Quebec City, PQ G1R 2J6, Canada
关键词
inverse planning; brachytherapy; HDR; prostate; simulated annealing;
D O I
10.1118/1.1368127
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An anatomy-based dose optimization algorithm is developed to automatically and rapidly produce a highly conformal dose coverage of the target volume while minimizing urethra, bladder, and rectal doses in the delivery of an high dose-rate (HDR) brachytherapy boost for the treatment of prostate cancer. The dwell times are optimized using an inverse planning simulated annealing algorithm (IPSA) governed entirely from the anatomy extracted from a CT and by a dedicated objective function (cost function) reflecting clinical prescription and constraints. With this inverse planning approach, the focus is on the physician's prescription and constraint instead of on the technical limitations. Consequently, the physician's control on the treatment is improved. The capacity of this algorithm to represent the physician's prescription is presented for a clinical prostate case. The computation time (CPU) for IPSA optimization is less than 1 min (41 s for 142 915 iterations) for a typical clinical case, allowing fast and practical dose optimization. The achievement of highly conformal dose coverage to the target volume opens the possibility to deliver a higher dose to the prostate without inducing overdosage of urethra and normal tissues surrounding the prostate. Moreover, using the same concept, it will be possible to deliver a boost dose to a delimited tumor volume within the prostate. Finally, this method can be easily extended to other anatomical sites. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:773 / 779
页数:7
相关论文
共 50 条
  • [21] High-dose-rate prostate brachytherapy inverse planning on dose-volume criteria by simulated annealing
    Deist, T. M.
    Gorissen, B. L.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (03): : 1155 - 1170
  • [22] Feasibility of adequate dose coverage in permanent prostate brachytherapy using a MR guided implant robot and inverse planning based on simulated annealing
    van den Bosch, M. R.
    Lips, I.
    Lagerburg, V.
    van Vulpen, M.
    Lagendijk, J.
    Moerland, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S34 - S34
  • [23] Evaluation of hybrid inverse planning and optimization (HIPO) algorithm for optimization in real-time, high-dose-rate (HDR) brachytherapy for prostate
    Pokharel, Shyam
    Rana, Suresh
    Blikenstaff, Joseph
    Sadeghi, Amir
    Prestidge, Bradley
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (04): : 96 - 107
  • [24] Three-dimension anatomy-based planning optimization for high dose rate vaginal vault brachytherapy
    Bahadur, Yasir A.
    Hassouna, Ashraf H.
    Constantinescu, Camelia T.
    Naga, Adly F.
    Ghasal, Noor M.
    Elsayed, Mohamed E.
    [J]. SAUDI MEDICAL JOURNAL, 2012, 33 (06) : 640 - 647
  • [25] Inverse Planning Simulated Annealing [IPSA] Optimization Versus Manual Optimization in the Treatment of Prostate Cancer Using Ir-192 High Dose Rate Brachytherapy
    Malhotra, H.
    Wang, Z.
    Kuettel, M.
    Podgorsak, M.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [26] Evaluation of Fast Inverse Planning Simulated Annealing for High Dose Rate Brachytherapy Tandem and Ovoid Plans
    Jia, Y.
    Rinker, L.
    Farmer, M.
    Wang, Y.
    Dubal, N.
    Ballo, M.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E497 - E497
  • [27] Anatomy-based three-dimensional dose optimization in brachytherapy using multiobjective genetic algorithms
    Lahanas, M
    Baltas, D
    Zamboglou, N
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 1904 - 1918
  • [28] Validation of Inverse Planning-simulated Annealing Class Solution for Prostate High-dose-rate Brachytherapy
    Sheth, N.
    Mistry, N.
    Sim, S.
    Weiss, M. F.
    Grebler, A.
    Geltzeiler, J.
    Litvin, Y. S.
    Keselman, I.
    Waldman, I.
    Yang, C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S372 - S373
  • [29] Dose Escalation in LogoRegionally Advanced Cervical Cancer Using Hybrid Inverse Planning Optimization in HDR Brachytherapy
    Andavar, Senni
    Roshan, Vikas
    Dhiman, Arashdeep
    Singh, Jagmohan
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3107 - 3107
  • [30] Prostate Brachytherapy Optimization Using GPU Accelerated Simulated Annealing and Monte Carlo Dose Simulation
    Mountris, Konstantinos A.
    Bert, Julien
    Visvikis, Dimitris
    [J]. 2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,