Effect of Sodium Hypochlorite Concentration on Electronic Apex Locator Reliability

被引:3
|
作者
Diemer, Franck [1 ]
Plews, Emma [1 ]
Georgelin-Gurgel, Marie [1 ]
Mishra, Lora [2 ]
Kim, Hyeon-Cheol [3 ]
机构
[1] CHU Toulouse, Clement Ader Inst, Dept Conservat Dent & Endodont, F-31400 Toulouse, France
[2] Siksha O Anusandhan, Inst Dent Sci, Dept Conservat Dent & Endodont, Bhubaneswar 751003, Odisha, India
[3] Pusan Natl Univ, Dent Res Inst, Sch Dent, Dept Conservat Dent, Yangsan 50612, South Korea
关键词
electronic apex locators; working length; concentration; sodium hypochlorite; WORKING LENGTH DETERMINATION; ROOT-ZX; VIVO EVALUATION; CANAL LENGTH; FILE SIZE; ACCURACY; IRRIGATION; IRRIGANTS; IPEX;
D O I
10.3390/ma15030863
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This ex vivo study aimed to measure the performance of an electronic apex locator (EAL) in the presence of sodium hypochlorite irrigants with different concentrations. Two EALs (Root ZX Mini and Locapex 6) were used to locate the apical foramen in 10 extracted single-rooted teeth in the presence of 0.5%, 2.5% and 5% sodium hypochlorite. Visual working lengths were also determined using #10 K-file under a microscope before the electronic measurements were made. The performance of both EALs was compared for the electronic working lengths determined under the different concentrations of sodium hypochlorite. A multiple-way ANOVA and PLSD Fisher's test with an alpha risk fixed at 5% were conducted. There were no statistical differences in the working lengths determined by both EALs between the three groups with different concentrations of sodium hypochlorite and their visual control measurements. When a +/- 0.5 mm margin was applied, the Root ZX Mini and the Locapex 6 presented 88% and 83% accuracy, respectively. Sodium hypochlorite concentration in irrigants does not affect the accuracy and reliability of either the Root ZX Mini or the Locapex 6. Electronic apex locators are reliable with any concentration of sodium hypochlorite irrigants.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] The influence of sodium hypochlorite irrigation on the accuracy of the Root ZX electronic apex locator
    Meares, WA
    Steiman, HR
    JOURNAL OF ENDODONTICS, 2002, 28 (08) : 595 - 598
  • [2] Evaluation of accuracy of an electronic apex locator in presence of sodium hypochlorite in primary teeth with and without resorption
    Topcuoglu, Gamze
    Kolcakoglu, Kevser
    JOURNAL OF CLINICAL PEDIATRIC DENTISTRY, 2023, 47 (06) : 150 - 154
  • [3] In vitro evaluation of the reliability of the ender electronic apex locator
    Ounsi, HF
    Haddad, G
    JOURNAL OF ENDODONTICS, 1998, 24 (02) : 120 - 121
  • [4] In vitro evaluation of the reliability of the Root ZX electronic apex locator
    Ounsi, HF
    Naaman, A
    INTERNATIONAL ENDODONTIC JOURNAL, 1999, 32 (02) : 120 - 123
  • [5] Accuracy and Reliability of a New Electronic Apex Locator: an In-vivo Study
    Chita, Jair Jatoba
    da Silva, Pedro Gregol
    Souza Pereira, Key Fabiano
    Onoda, Helio Katsuya
    Borba Junior, Jose de Camargo
    Spironelli Ramos, Carlos Alberto
    PESQUISA BRASILEIRA EM ODONTOPEDIATRIA E CLINICA INTEGRADA, 2012, 12 (04): : 456 - 462
  • [7] Evaluation of heated sodium hypochlorite’s effect on the accuracy of contemporary electronic apex locators: an in vitro study
    İkbal Sena Çelebi Keskin
    Turgut Yağmur Yalçın
    BMC Oral Health, 25 (1)
  • [8] The effect of residual calcium hydroxide on the accuracy of a contemporary electronic apex locator
    Ustun, Yakup
    Uzun, Ozgur
    Er, Ozgur
    Canakci, Burhan Can
    Topuz, Ozgur
    ACTA ODONTOLOGICA SCANDINAVICA, 2015, 73 (02) : 132 - 136
  • [9] A simple model to demonstrate the electronic apex locator
    Tinaz, AC
    Alaçam, T
    Topuz, Ö
    INTERNATIONAL ENDODONTIC JOURNAL, 2002, 35 (11) : 940 - 945
  • [10] A Study on the Design and Construction of Electronic Apex Locator
    Boonyagul, S.
    Eadsongkram, T.
    Keawgun, T.
    2016 3RD INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING (BME-HUST), 2016, : 75 - 78