Characteristic responses of an atmospheric pressure DC micro-plasma detector for gas chromatography to organic functional groups

被引:10
|
作者
Fu, Yu-Mao [1 ]
Chu, Shih-Chi [1 ]
Lu, Chia-Jung [1 ]
机构
[1] Fu Jen Catholic Univ, Dept Chem, Hsingchuang, Taiwan
关键词
micro-plasma; helium discharge detector; gas chromatograph;
D O I
10.1016/j.microc.2007.10.005
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A gas chromatographic detector using direct current micro-plasma discharged a distance of less than 50 gm under helium at atmospheric pressure is described. The micro-plasma device was assembled on glass substrates; the volume of the plasma region was only 0.58 nL. Very low power consumption (15 mW) and only carrier helium flow were needed to maintain micro-plasma stability. Organic compounds with various functional groups were chromatographically separated and fed into the micro-plasma. Fluctuations in plasma current were detected as the sample passed through the micro-plasma. All non-halogen organics showed positive peaks, while simple alkyl halides showed negative peaks. Compounds with both aromatic and halogen groups exhibited hybrid peak shapes, consisting of both positive and negative peaks. The detection limits (3 sigma) ranged from 24.5 pg for m-xylene to 756 pg for 1,2-dichloroethane, depending on differences in ionization and electron-capture efficiency. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:7 / 12
页数:6
相关论文
共 50 条
  • [21] Probe measurements of electron energy spectrum and plasma-wall interaction in Helium/air micro-plasma at atmospheric pressure
    Adams, S. F.
    Demidov, V. I.
    Hensley, A. L.
    Koepke, M. E.
    Kurlyandskaya, I. P.
    Miles, J. A.
    Tolson, B. A.
    7TH INTERNATIONAL WORKSHOP AND SUMMER SCHOOL ON PLASMA PHYSICS (IWSSPP'16), 2018, 982
  • [22] Energy-Controlled Atmospheric-Pressure Micro-Plasma Using Multistep External Ballast Capacitors
    Ha, Chang-Seung
    Seo, Kwon-Sang
    Cha, Ju-Hong
    Kim, Dong-Hyun
    Lee, Hae June
    Lee, Ho-Jun
    IEEE TRANSACTIONS ON PLASMA SCIENCE, 2021, 49 (12) : 3823 - 3831
  • [23] AN ATMOSPHERIC-PRESSURE, ARGON-AFTERGLOW DETECTOR FOR GAS-CHROMATOGRAPHY
    RICE, GW
    DSILVA, AP
    FASSEL, VA
    ANALYTICA CHIMICA ACTA, 1984, 166 (DEC) : 27 - 38
  • [24] ATMOSPHERIC-PRESSURE ACTIVE NITROGEN AFTERGLOW AS A DETECTOR FOR GAS-CHROMATOGRAPHY
    RICE, GW
    RICHARD, JJ
    DSILVA, AP
    FASSEL, VA
    ANALYTICAL CHEMISTRY, 1981, 53 (09) : 1519 - 1522
  • [25] Development of an Atmospheric Pressure Glow Discharge Detector for Capillary Column Gas Chromatography
    Hyo J. Kim
    Young A. Woo
    Jong S. Kang
    Susanna S. Anderson
    Edward H. Piepmeier
    Microchimica Acta, 2000, 134 : 1 - 7
  • [26] Development of an atmospheric pressure glow discharge detector for capillary column gas chromatography
    Kim, HJ
    Woo, YA
    Kang, JS
    Anderson, SS
    Piepmeier, EH
    MIKROCHIMICA ACTA, 2000, 134 (1-2) : 1 - 7
  • [27] A micro gas chromatography column with a micro thermal conductivity detector for volatile organic compound analysis
    Sun, J. H.
    Cui, D. F.
    Chen, X.
    Zhang, L. L.
    Cai, H. Y.
    Li, H.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (02):
  • [28] Sterilization by negative and positive DC plasma with a micro discharge gap at atmospheric pressure
    Li, Hua
    Jiang, Lin-Xiu
    Jiang, Yong-Rong
    Zhu, Jian-Min
    Chen, Zhen-Cheng
    MODERN PHYSICS LETTERS B, 2017, 31 (32):
  • [29] Micro-patterning of Functional Groups onto Polymer Surface Using Capillary Atmospheric Pressure Plasma Jet
    Motrescu, Iuliana
    Ogino, Akihisa
    Nagatsu, Masaaki
    JOURNAL OF PHOTOPOLYMER SCIENCE AND TECHNOLOGY, 2012, 25 (04) : 529 - 534
  • [30] Microstructures of SiO2 and TiOX films deposited by atmospheric pressure inductively coupled micro-plasma jet
    Aita, Tadahiro
    Ogawa, Kazunari
    Saito, Yusuke
    Sumiyoshi, Yuichi
    Higuchi, Takeshi
    Sato, Shimio
    SURFACE & COATINGS TECHNOLOGY, 2010, 205 (03): : 861 - 866