Harmonization and external quality assessment of antithrombin activity assays

被引:4
|
作者
van den Besselaar, Anton M. H. P. [1 ]
Haas, Fred J. L. M. [2 ]
Kuypers, Aldy W. H. M. [3 ]
机构
[1] Leiden Univ, Med Ctr, Dept Thrombosis & Hemostasis, NL-2300 RC Leiden, Netherlands
[2] St Antonius Hosp, Dept Clin Chem, Nieuwegein, Netherlands
[3] Maasziekenhuis Pantein, Dept Clin Chem, Boxmeer, Netherlands
关键词
Antithrombin assays; Harmonization; Commutability; External quality assessment; Calibrators; PROJECT CALIBRATION 2000; FIELD-METHODS; PROTEIN-S; PROGRAM; COMMUTABILITY; THROMBOPHILIA; PATHOLOGISTS; PERFORMANCE; COAGULATION; DEFICIENCY;
D O I
10.1016/j.thromres.2011.06.007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In the framework of a Dutch project named "Calibration 2000" harmonization of antithrombin activity assays was studied. The commutability of potential calibrators for antithrombin was assessed by means of a twin-study design, which is a multicentre, split-patient sample, between-field-methods protocol. The twin-study consisted of simultaneous analysis of fresh-frozen patient plasmas and three potential calibrators for antithrombin by 30 Dutch laboratories forming 15 couples. The state-of-the-art intralaboratory standard deviation (SD(SA)) was used to assess the commutability of the potential calibrators. The regression line residuals for the potential calibrators were normalized by expressing them as multiples of SDSA. All residuals of the potential calibrators were within the 3 x SD(SA) limit. One potential calibrator was used in an attempt to harmonize antithrombin assay results in a Dutch field study. The interlaboratory coefficient of variation (CV) of the antithrombin results for three test samples could be reduced from 6.9 - 13.2% (before harmonization) to 5.6 - 9.8% using the common calibrator. Conclusion: The potential calibrators were commutable. Limited harmonization was achieved by using a common calibrator for all participants. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:187 / 191
页数:5
相关论文
共 50 条
  • [21] Evaluation of assays for aminoglycosides in serum:a comparison of accuracy and precision based on external quality assessment
    LI Meng-ya
    CHEN Chao-yang
    LEI Li-li
    Zhao Xia
    ZHOU Ying
    CUI Yi-min
    [J]. 中国药理学与毒理学杂志, 2019, (10) : 815 - 816
  • [22] Necessity for high-quality reference materials in the harmonization of laboratory assays
    Johnson, A. Myron
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (06) : 747 - 748
  • [23] Investigating the Current Harmonization Status of Tumor Markers Using Global External Quality Assessment Programs: A Feasibility Study
    van Rossum, Huub H.
    Holdenrieder, Stefan
    Ballieux, Bart E. P. B.
    Badrick, Tony C.
    Yun, Yeo-Min
    Zhang, Chuanbao
    Patel, Dina
    Thelen, Marc
    Song, Junghan
    Wojtalewicz, Nathalie
    Unsworth, Nick
    Vesper, Hubert W.
    Cui, Wei
    Ramanathan, Lakshmi, V
    Sturgeon, Catharine
    Meng, Qing H.
    [J]. CLINICAL CHEMISTRY, 2024, 70 (04) : 669 - 679
  • [24] External Quality Assurance of Antithrombin, Protein C, and Protein S Assays Results of the College of American Pathologists Proficiency Testing Program in Thrombophilia
    Cunningham, Mark T.
    Olson, John D.
    Chandler, Wayne L.
    Van Cott, Elizabeth M.
    Eby, Charles S.
    Teruya, Jun
    Hollensead, Sandra C.
    Adcock, Dorothy M.
    Allison, Paul M.
    Kottke-Marchant, Kandice K.
    Smith, Marc D.
    [J]. ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2011, 135 (02) : 227 - 232
  • [25] Evaluation of Assays to Measure Aminoglycosides in Serum: Comparison of Accuracy and Precision Based on External Quality Assessment
    Li, Mengya
    Ma, Lingyun
    Chen, Chaoyang
    Lei, Lili
    Zhai, Xingyu
    Luo, Daohuang
    Zhou, Ying
    Cui, Yimin
    [J]. THERAPEUTIC DRUG MONITORING, 2020, 42 (05) : 710 - 715
  • [26] External quality assessment of bone metabolism marker assays. Initial experiences in a UKNEQAS programme
    Whitham, KM
    Milford-Ward, A
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2000, 38 (11) : 1121 - 1124
  • [27] External quality assessment program for Chlamydia trachomatis diagnostic testing by nucleic acid amplification assays
    Land, S
    Tabrizi, S
    Gust, A
    Johnson, E
    Garland, S
    Dax, EM
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (08) : 2893 - 2896
  • [28] Performance of semiquantitative and quantitative D-dimer assays in the ECAT External Quality Assessment Program
    de Maat, MPM
    Meijer, P
    Nieuwenhuizen, W
    Haverkate, F
    Kluft, C
    [J]. SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2000, 26 (06): : 625 - 630
  • [29] ANTITHROMBIN-III ASSAYS
    MESSMORE, HL
    FAREED, J
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1978, 240 (04): : 345 - 346
  • [30] ANTITHROMBIN-III ASSAYS
    YUE, RH
    GERTLER, MM
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1979, 242 (13): : 1360 - 1361