Design of a video capsule endoscopy system with low-power ASIC for monitoring gastrointestinal tract

被引:21
|
作者
Liu, Gang [1 ]
Yan, Guozheng [1 ]
Zhu, Bingquan [1 ]
Lu, Li [1 ]
机构
[1] Shanghai Jiao Tong Univ, Inst Precise Engn & Intelligent Microsyst, Shanghai 200240, Peoples R China
基金
中国国家自然科学基金;
关键词
Wireless capsule endoscopy (WCE); Application-specific integrated circuit (ASIC); Image compression; Improved JPEG-LS encoder; RF transmitter; In vivo trials; WIRELESS; COMPRESSION;
D O I
10.1007/s11517-016-1472-2
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In recent years, wireless capsule endoscopy (WCE) has been a state-of-the-art tool to examine disorders of the human gastrointestinal tract painlessly. However, system miniaturization, enhancement of the image-data transfer rate and power consumption reduction for the capsule are still key challenges. In this paper, a video capsule endoscopy system with a low-power controlling and processing application-specific integrated circuit (ASIC) is designed and fabricated. In the design, these challenges are resolved by employing a microimage sensor, a novel radio frequency transmitter with an on-off keying modulation rate of 20 Mbps, and an ASIC structure that includes a clock management module, a power-efficient image compression module and a power management unit. An ASIC-based prototype capsule, which measures I broken vertical bar 11 mm x 25 mm, has been developed here. Test results show that the designed ASIC consumes much less power than most of the other WCE systems and that its total power consumption per frame is the least. The image compression module can realize high near-lossless compression rate (3.69) and high image quality (46.2 dB). The proposed system supports multi-spectral imaging, including white light imaging and autofluorescence imaging, at a maximum frame rate of 24 fps and with a resolution of 400 x 400. Tests and in vivo trials in pigs have proved the feasibility of the entire system, but further improvements in capsule control and compression performance inside the ASIC are needed in the future.
引用
收藏
页码:1779 / 1791
页数:13
相关论文
共 50 条
  • [1] Design of a video capsule endoscopy system with low-power ASIC for monitoring gastrointestinal tract
    Gang Liu
    Guozheng Yan
    Bingquan Zhu
    Li Lu
    [J]. Medical & Biological Engineering & Computing, 2016, 54 : 1779 - 1791
  • [2] Integration of Low-Power ASIC and MEMS Sensors for Monitoring Gastrointestinal Tract Using a Wireless Capsule System
    Arefin, Md Shamsul
    Redoute, Jean-Michel
    Yuce, Mehmet Rasit
    [J]. IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2018, 22 (01) : 87 - 97
  • [3] A Wireless Capsule System With ASIC for Monitoring the Physiological Signals of the Human Gastrointestinal Tract
    Xu, Fei
    Yan, Guozheng
    Zhao, Kai
    Lu, Li
    Gao, Jinyang
    Liu, Gang
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2014, 8 (06) : 871 - 880
  • [4] A Wireless Capsule Endoscopic System with a Low-Power Controlling and Processing ASIC
    Chen, Xinkai
    Zhang, Xiaoyu
    Zhang, Lingwei
    Qi, Nan
    Jiang, Hanjun
    Wang, Zhihua
    [J]. 2008 IEEE ASIAN SOLID-STATE CIRCUITS CONFERENCE, 2008, : 317 - +
  • [5] A Wireless Capsule Endoscope System With Low-Power Controlling and Processing ASIC
    Chen, Xinkai
    Zhang, Xiaoyu
    Zhang, Linwei
    Li, Xiaowen
    Qi, Nan
    Jiang, Hanjun
    Wang, Zhihua
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2009, 3 (01) : 11 - 22
  • [6] A Subsample-Based Low-Power Image Compressor for Capsule Gastrointestinal Endoscopy
    Meng-Chun Lin
    Lan-Rong Dung
    [J]. EURASIP Journal on Advances in Signal Processing, 2011
  • [7] A Subsample-Based Low-Power Image Compressor for Capsule Gastrointestinal Endoscopy
    Lin, Meng-Chun
    Dung, Lan-Rong
    [J]. EURASIP JOURNAL ON ADVANCES IN SIGNAL PROCESSING, 2011,
  • [8] Low-Power Wireless Electronic Capsule for Long-Term Gastrointestinal Monitoring
    Kai Zhao
    Guozheng Yan
    Li Lu
    Fei Xu
    [J]. Journal of Medical Systems, 2015, 39
  • [9] Low-Power Wireless Electronic Capsule for Long-Term Gastrointestinal Monitoring
    Zhao, Kai
    Yan, Guozheng
    Lu, Li
    Xu, Fei
    [J]. JOURNAL OF MEDICAL SYSTEMS, 2015, 39 (02)
  • [10] Low-power image compression for wireless capsule endoscopy
    Turcza, Pawel
    Duplaga, Mariusz