Determination of k 0-values for some elements in neutron capture gamma-ray for prompt gamma activation analysis

被引:3
|
作者
Hamid, A. [1 ]
机构
[1] Atom Energy Author, Hot Labs Ctr, Dept Radiated Pollut, Cairo 13759, Egypt
关键词
k(0)-PGNAA technique; Prompt gamma-ray; Prompt k(0)-factors; Neutron flux ratio; Monostandard method; K(0)-FACTORS; SAMPLES;
D O I
10.1007/s10967-011-1388-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Exploration and modification of data for k (0)-PGAA (prompt gamma-ray activation analysis) values used as standards for k (0) standardization are needed. An integrated system has been installed and calibrated in Hot Laboratories Center for PGAA, using Cf-252 isotopic neutron source with neutron flux of 6.16E8 n/cm(2) s. The prompt k (0)-factors of about 12 elements were determined versus the 1951.1 keV gamma-ray of the Cl-35, as well as analyzing crude oil and oil product samples. The concentrations of the elements in crude oil H, B, Mg, Al, Na, Si, P, S, Cl, V and Cd were 1.18E5, 0.084, 5.48E2, 8.45E2 4.88E2, 1.62E2, 3.72E2, 8.2, 144.3, 393 and 209.2 ppm while in oil product samples were 1.32E5, 5.87, 4.56E2, 4.22E2, 7.16E3, 699, 157, 8.74, 492.3, 61 and 198.2 ppm. Validation of the k (0)-PGAA was applied in analyzing standard IAEA reference material (soil-7) which give good agreement with the literature data. The Cf-252 neutron beam in radiation position was characterized by the cadmium-ratio method, and found that the cadmium ratio (f), was 160.
引用
收藏
页码:623 / 628
页数:6
相关论文
共 50 条
  • [41] Prompt gamma-ray activation analysis for determination of boron in aqueous solutions
    Baechler, S
    Kudejova, P
    Jolie, J
    Schenker, JL
    Stritt, N
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 488 (1-2): : 410 - 418
  • [42] Determination of trace boron in steels by prompt gamma-ray activation analysis
    Kim, I. J.
    Cho, K. H.
    Paul, R. L.
    [J]. ANALYTICAL SCIENCE AND TECHNOLOGY, 2009, 22 (04): : 302 - 306
  • [43] k0 Standardization Approach in Neutron-Induced Prompt Gamma-Ray Analysis at JAERI
    H. Matsue
    C. Yonezawa
    [J]. Journal of Radioanalytical and Nuclear Chemistry, 2000, 245 : 189 - 194
  • [44] ASSESSMENT OF NEUTRON CAPTURE GAMMA-RAY ANALYSIS
    GARBRAH, BW
    WHITLEY, JE
    [J]. INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1968, 19 (08): : 605 - &
  • [45] Neutron spectrum correction of k0-factors for k0-based neutron-induced prompt gamma-ray analysis
    Matsue, H
    Yonezawa, C
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2003, 255 (01) : 125 - 129
  • [46] Analytical capability of an explosives detection by a prompt gamma-ray neutron activation analysis
    Im, Hee-Jung
    Cho, Hyun-Je
    Song, Byoung Chul
    Park, Yong Joon
    Chung, Yong-Sam
    Kim, Won-Ho
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 566 (02): : 442 - 447
  • [47] Cold neutron prompt gamma-ray activation analysis at NIST - Recent developments
    Paul, RL
    Lindstrom, RM
    Heald, AE
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 1997, 215 (01) : 63 - 68
  • [48] Cold neutron prompt gamma-ray activation analysis at NIST — Recent developments
    R. L. Paul
    R. M. Lindstrom
    A. E. Heald
    [J]. Journal of Radioanalytical and Nuclear Chemistry, 1997, 215 : 63 - 68
  • [49] Neutron spectrum correction of k0-factors for k0-based neutron-induced prompt gamma-ray analysis
    H. Matsue
    Ch. Yonezawa
    [J]. Journal of Radioanalytical and Nuclear Chemistry, 2003, 255 : 125 - 129
  • [50] Application of the k0 method in neutron activation analysis and in prompt gamma activation analysis
    Revay, Zs
    Kennedy, G.
    [J]. RADIOCHIMICA ACTA, 2012, 100 (8-9) : 687 - 698