CHOREOGRAPHING COUCH AND COLLIMATOR IN VOLUMETRIC MODULATED ARC THERAPY

被引:61
|
作者
Yang, Yingli [1 ]
Zhang, Pengpeng [1 ]
Happersett, Laura [1 ]
Xiong, Jianping [1 ]
Yang, Jie [1 ]
Chan, Maria [1 ]
Beal, Kathryn [2 ]
Mageras, Gig [1 ]
Hunt, Margie [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10065 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10065 USA
关键词
Volumetric arc modulated therapy; Trajectory based; Treatment planning; Radiotherapy; CNS tumors; RADIOTHERAPY; DELIVERY; IMRT; OPTIMIZATION; QUALITY; GLIOMAS; CANCER; VMAT;
D O I
10.1016/j.ijrobp.2010.10.016
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To design and optimize trajectory-based, noncoplanar subarcs for volumetric modulated arc therapy (VMAT) deliverable on both Varian TrueBEAM system and traditional accelerators; and to investigate their potential advantages for treating central nervous system (CNS) tumors. Methods and Materials: To guide the computerized selection of beam trajectories consisting of simultaneous couch, gantry, and collimator motion, a score function was implemented to estimate the geometric overlap between targets and organs at risk for each couch/gantry angle combination. An initial set of beam orientations is obtained as a function of couch and gantry angle, according to a minimum search of the score function excluding zones of collision. This set is grouped into multiple continuous and extended subarcs subject to mechanical limitations using a hierarchical clustering algorithm. After determination of couch/gantry trajectories, a principal component analysis finds the collimator angle at each beam orientation that minimizes residual target organ at risk overlaps. An in-house VMAT optimization algorithm determines the optimal multileaf collimator position and monitor units for control points within each subarc. A retrospective study of 10 CNS patients compares the proposed method of VMAT trajectory with dynamic gantry, leaves, couch, and collimator motion (Tra-VMAT); a standard noncoplanar VMAT with no couch/collimator motion within subarcs (Std-VMAT); and noncoplanar intensity-modulated radiotherapy (IMRT) plans that were clinically used. Results: Tra-VMAT provided improved target dose conformality and lowered maximum dose to brainstem, optic nerves, and chiasm by 7.7%, 1.1%, 2.3%, and 1.7%, respectively, compared with Std-VMAT. Tra-VMAT provided higher planning target volume minimum dose and reduced maximum dose to chiasm, optic nerves, and cochlea by 6.2%, 1.3%, 6.3%, and 8.4%, respectively, and reduced cochlea mean dose by 8.7%, compared with IMRT. Tra-VMAT averaged beam-on time was comparable to Std-VMAT but significantly (45%) less than IMRT. Conclusion: Optimized couch, gantry, and collimator trajectories may be integrated into VMAT with improved mechanical flexibility and may provide better dosimetric properties and improved efficiency in the treatment of CNS tumors. (C) 2011 Elsevier Inc.
引用
收藏
页码:1238 / 1247
页数:10
相关论文
共 50 条
  • [1] Choreographing Couch and Collimator in Volumetric Modulated Arc Therapy
    Yang, Y.
    Zhang, P.
    Happersett, L.
    Xiong, J.
    Yang, J.
    Chan, M.
    Mageras, G.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 3369 - +
  • [2] Couch modelling for volumetric modulated arc therapy (VMAT)
    Chyou, Te-yu
    Lorenz, Friedlieb
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (02) : 471 - 480
  • [3] Couch modelling for volumetric modulated arc therapy (VMAT)
    Te-yu Chyou
    Friedlieb Lorenz
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 471 - 480
  • [4] Collimator Angle Optimization in Volumetric Modulated Arc Therapy
    Andersen, A.
    Johnson, C.
    Bartlett, G.
    Das, I.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3627 - 3627
  • [5] Dynamic Collimator Rotation for Volumetric Modulated Arc Therapy
    Fix, M. K.
    Frei, D.
    Volken, W.
    Terribilini, D.
    Manser, P.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3412 - 3412
  • [6] Multileaf Collimator Quality Assurance for Volumetric Modulated Arc Therapy
    Lin, T.
    Hossain, M.
    Ma, C.
    [J]. MEDICAL PHYSICS, 2011, 38 (06) : 3598 - U31
  • [7] Dosimetric Impact of Treatment Couch on Volumetric Modulated Arc Therapy (VMAT)
    Ye, J.
    Liu, H.
    Kim, J. J.
    Deng, J.
    Chen, Z.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S708 - S709
  • [8] The importance of collimator angle error in volumetric-modulated arc therapy
    Puchades-Puchades, Vicente
    Serna-Berna, Alfredo
    Mata-Colodro, Fernando
    Ramos-Amores, David
    Alcaraz-Banos, Miguel
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2015, 14 (03) : 323 - 330
  • [9] Effect of the Collimator Angle on Dosimetric Verification of Volumetric Modulated Arc Therapy
    Kim, Yong Ho
    Park, Ha Ryung
    Kim, Won Taek
    Kim, Dong Won
    Ki, Yongkan
    Lee, Juhye
    Bae, Jinsuk
    Park, Dahl
    Jeon, Hosang
    Nam, Ji Ho
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (01) : 243 - 247
  • [10] Effect of the collimator angle on dosimetric verification of volumetric modulated arc therapy
    Yong Ho Kim
    Ha Ryung Park
    Won Taek Kim
    Dong Won Kim
    Yongkan Ki
    Juhye Lee
    Jinsuk Bae
    Dahl Park
    Hosang Jeon
    Ji Ho Nam
    [J]. Journal of the Korean Physical Society, 2015, 67 : 243 - 247