An automated high-performance liquid chromatographic method, benzodiazepines by REMEDi HS, was used to analyze benzodiazepines and their metabolites after β-glucuronidase hydrolysis of 1-mL urine specimens from the following: 924 clinic and hospital patients whose specimens had previously been found to be presumptively positive using either EMIT® or Triage® immunoassay methodologies and 128 individuals whose specimens had screened negative by EMIT d.a.u.(TM). REMEDi analyses did not correlate with the immunoassay results in 136 of the positive and three of the negative urine specimens. Gas chromatographic-mass spectrometric (GC-MS) confirmatory analyses were performed on these discordant specimens using 3 mL β- glucuronidase-hydrolyzed urine followed by extraction with chloroform- isopropanol (9:1) and derivatization with N,O- bis(trimethylsilyl)trifluoroacetamide. Two benzodiazepines, flunitrazepam and clonazepam, and their 7-amino metabolites were analyzed without prior derivatization. The analyses established 87% concordance between REMEDi and GC-MS versus 13% concordance with immunoassay for the subset. GC-MS analysis of these 142 specimens demonstrated two reasons for the nonconcurrence between REMEDi and EMIT: EMIT had given either false-negative or false- positive results and EMIT had given a positive result even though the determined metabolites were below the 200-ng/mL cutoff for the immunoassay and the 80-ng/mL cutoff for REMEDi. A total of 23 specimens were found to contain only lorazepam by REMEDi and GC-MS, 15 of which had been screened by Triage. A reevaluation of these 23 specimens by EMIT d.a.u. demonstrated that 11 were positive. This finding was in contrast to previous reports that EMIT will not detect lorazepam glucuronide in urine. An unexpected finding was the REMEDi identification and subsequent GC-MS confirmation of 7- aminoflunitrazepam, a urinary metabolite of flunitrazepam that is not available in the United States and that represented illicit use by four patients. A distinct advantage of REMEDi proved to be its capability in identifying demoxepam, a major metabolite of chlordiazepoxide; GC-MS analysis could not detect this metabolite because of its thermal decomposition to nordiazepam. To further evaluate the specificity of REMEDi, we conducted GC- MS analyses in a random fashion on 55 additional nondiscordant urine specimens that were identified as either positive or negative, as well as 22 specimens identified as containing 7-aminoclonazepam by REMEDi. Concurrence was observed between the two methods for all specimens, with the exception of one apparent false positive for α-hydroxyalprazolam by REMEDi. The reproducibility of the REMEDi method was found to be excellent; it was assessed by comparing results of 266 specimens that were reprocessed in different batches and for known calibrators and controls also processed with each batch. Study results demonstrated that the automated REMEDi assay for urinary benzodiazepines and their metabolites was comparable with GC-MS but had distinct advantages over GC-MS because of the following reasons: simplicity of the assay, less time required for analyses, and provision of additional information concerning the parent benzodiazepine.
引用
收藏
页码:416 / 424
页数:9
相关论文
共 30 条
  • [11] Serum and urine concentrations of flunitrazepam and metabolites, after a single oral dose, by immunoassay and GC-MS
    Snyder, H
    Schwenzer, KS
    Pearlman, R
    McNally, A
    Tsilimidos, M
    Salamone, SJ
    Brenneisen, R
    ElSohly, MA
    Feng, S
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2001, 25 (08) : 699 - 704
  • [12] Comparison of automated SPDE and SPME GC-MS in a search for African elephant urinary pheromones
    House, Sam J.
    Eggert, Mindy S.
    Weddell, Margaret E.
    Goodwin, Thomas E.
    Rasmussen, L. E. L.
    Schulte, Bruce A.
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2006, 231
  • [13] Determination of urinary organophosphorus pesticide metabolites by automated solid phase extraction, post extraction derivatization, and GC-MS/MS.
    De Alwis, GKH
    Needham, LL
    Barr, DB
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2005, 229 : U85 - U85
  • [14] Screening of clostebol and its metabolites in bovine urine with ELISA and comparison with GC-MS results in an interlaboratory study
    Crabbe, P
    Meyer, UJ
    Zhi, ZL
    Pieraccini, G
    O'Keeffe, M
    Van Peteghem, C
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2003, 27 (04) : 213 - 220
  • [15] Comparison of Immunoassay Screening Tests and LCMS-MS for Urine Detection of Benzodiazepines and Their Metabolites: Results of a National Proficiency Test
    Bertol, Elisabetta
    Vaiano, Fabio
    Borsotti, Maurizio
    Quercioli, Massimo
    Mari, Francesco
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2013, 37 (09) : 659 - 664
  • [16] Determination of metabolites in Uncaria sinensis by HPLC and GC-MS after green solvent microwave-assisted extraction
    Tan, Swee Ngin
    Yong, Jean Wan Hong
    Teo, Chin Chye
    Ge, Liya
    Chan, Yee Wen
    Hew, Choy Sin
    [J]. TALANTA, 2011, 83 (03) : 891 - 898
  • [17] GC-MS characterization of urinary metabolites and changes of ethisterone and testosterone profile after oral administration of danazol in equine
    Kim, JY
    Choi, MH
    Kim, SJ
    Kyong, JB
    Chung, BC
    [J]. JOURNAL OF VETERINARY PHARMACOLOGY AND THERAPEUTICS, 2001, 24 (02) : 147 - 153
  • [18] Identification of 4-methylthioamphetamine and some of its metabolites in mouse urine by GC-MS after acute administration
    Carmo, H
    de Boer, D
    Remiao, F
    Carvalho, F
    dos Reys, LA
    Bastos, MD
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2002, 26 (04) : 228 - 232
  • [19] Fatal opiates overdose.: Toxicological identification of various metabolites in a blood sample by GC-MS after silylation
    Balíková, M
    Maresová, V
    [J]. FORENSIC SCIENCE INTERNATIONAL, 1998, 94 (03) : 201 - 209
  • [20] Comparison of immunoassay and HPLC-MS/MS used to measure urinary metabolites of atrazine, metolachlor, and chlorpyrifos from farmers and non-farmers in Iowa
    Curwin, Brian D.
    Hein, Misty J.
    Barr, Dana B.
    Striley, Cynthia
    [J]. JOURNAL OF EXPOSURE SCIENCE AND ENVIRONMENTAL EPIDEMIOLOGY, 2010, 20 (02) : 205 - 212