Repeatability of Retinal Macular Thickness Measurements in Healthy Subjects and Diabetic Patients with Clinically Significant Macular Edema: Evaluation of the Follow-Up System of Spectralis Optical Coherence Tomography

被引:6
|
作者
Fiore, Tito [1 ]
Lupidi, Marco [1 ]
Androudi, Sofia [4 ]
Giansanti, Fabrizio [3 ]
Fruttini, Daniela [2 ]
Cagini, Carlo [1 ]
机构
[1] Eye Clin, Dept Ophthalmol, Perugia, Italy
[2] Univ Perugia, Dept Econ Finance & Stat, I-06100 Perugia, Italy
[3] Eye Clin, Dept Otoneuroophthalmol Surg Sci, Florence, Italy
[4] Univ Thessalia, Dept Ophthalmol, Larisa, Greece
关键词
Optical coherence tomography; Spectralis; Diabetic macular edema; REPRODUCIBILITY; DOMAIN; DEGENERATION; DIAGNOSIS; VOLUME; HOLES; OCT;
D O I
10.1159/000380832
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: It was the aim of this study to compare the repeatability of Spectralis optical coherence tomography (OCT) retinal thickness measurements in healthy subjects and diabetic patients with clinically significant macular edema (CSME) with or without the use of the follow-up system. Methods: Thirty-eight eyes of 38 healthy subjects (control group) and 68 eyes of 68 diabetic patients with CSME were included in the study. The coefficient of repeatability (CR) and intrasession coefficients of variation were tested with 20 x 15 degree raster scans consisting of 19 high-resolution line scans (15 frames per scan) that were repeated 3 times by 1 experienced examiner.The first scan was set as the reference scan, whereas the second and third scans were the follow-up scans and were performed with and without the use of the follow-up mode, respectively. Results: The means and standard deviations for the central foveal subfield (CSF) in healthy subjects and diabetic patients were 289 +/- 21 and 402 +/- 105 mu m, respectively. Particularly in diabetic patients, examining the CSF, CR was 2.67% (10.73 mu m) and 6.73% (27.01 mu m) with and without using the follow-up mode, respectively, and the difference was statistically significant (p < 0.05). Conclusion: These results support the hypothesis that the follow-up system improves the repeatability either in healthy subjects or in diabetic patients with poor fixation. The wider improvement in repeatability in diabetic patients in the follow-up system group compared to the no follow-up system group are probably related to poor patient fixation or eye movement in patients with CSME. (C) 2015 S. Karger AG, Basel
引用
收藏
页码:186 / 191
页数:6
相关论文
共 50 条
  • [1] Repeatability of Retinal Macular Thickness Measurements in Patients with Clinically Significant Macular Edema Using Two Different Scanning Protocols of Spectralis Optical Coherence Tomography
    Fiore, Tito
    Lupidi, Marco
    Androudi, Sofia
    Giansanti, Fabrizio
    Fruttini, Daniela
    Cagini, Carlo
    [J]. OPHTHALMOLOGICA, 2015, 234 (03) : 167 - 171
  • [2] Optical coherence tomography for retinal thickness measurement in diabetic patients without clinically significant macular edema
    Schaudig, UH
    Glaefke, C
    Scholz, F
    Richard, G
    [J]. OPHTHALMIC SURGERY AND LASERS, 2000, 31 (03): : 182 - 186
  • [3] Repeatability of Spectralis OCT Measurements of Macular Thickness and Volume in Diabetic Macular Edema
    Comyn, Oliver
    Heng, Ling Zhi
    Ikeji, Felicia
    Bibi, Kanom
    Hykin, Philip G.
    Bainbridge, James W.
    Patel, Praveen J.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2012, 53 (12) : 7754 - 7759
  • [4] Quantitative assessment of retinal thickness in diabetic patients with and without clinically significant macular edema using optical coherence tomography
    Yang, CS
    Cheng, CY
    Lee, FL
    Hsu, WM
    Liu, JH
    [J]. ACTA OPHTHALMOLOGICA SCANDINAVICA, 2001, 79 (03): : 266 - 270
  • [5] Repeatability of Macular Thickness Measurements in Patients With Diabetic Macular Edema
    Invernizzi, A.
    Pellegrini, M.
    Cigada, M.
    Peroglio, A.
    Bassani, N.
    Staurenghi, G.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [6] Optical coherence tomography titrated photocoagulation in diabetic clinically significant macular edema
    Das, T
    Simanjuntak, G
    Sumasri, K
    [J]. MACULA: DIAGNOSIS, TREATMENT AND FUTURE TRENDS, 2004, : 185 - 194
  • [7] Repeatability and reproducibility of retinal and choroidal thickness measurements in Diabetic Macular Edema using Swept-source Optical Coherence Tomography
    Sala-Puigdollers, Anna
    Zarranz-Ventura, Javier
    Figueras-Roca, Marc
    Hereu, Mireia
    Hernandez, Teresa
    Morato, Montserrat
    Mesquida, Marina
    Llorens, Victor
    Teresa Sainz de la Maza, Maria
    Adan Civera, Alfredo
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [8] Repeatability and reproducibility of retinal and choroidal thickness measurements in Diabetic Macular Edema using Swept-source Optical Coherence Tomography
    Sala-Puigdollers, Anna
    Figueras-Roca, Marc
    Hereu, Mireia
    Hernandez, Teresa
    Morato, Montse
    Adan, Alfredo
    Zarranz-Ventura, Javier
    [J]. PLOS ONE, 2018, 13 (07):
  • [9] Optical coherence tomography in the diagnosis and follow-up of patients with uveitic macular edema
    Reinthal, EK
    Völker, M
    Freudenthaler, N
    Grüb, M
    Zierhut, M
    Schlote, T
    [J]. OPHTHALMOLOGE, 2004, 101 (12): : 1181 - 1188
  • [10] Reproducibility of Retinal Thickness Measurements in Healthy Subjects Using Spectralis Optical Coherence Tomography
    Menke, Marcel N.
    Dabov, Simeon
    Knecht, Pascal
    Sturm, Veit
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (03) : 467 - 472