An anatomic coordinate system of the femoral neck for highly reproducible BMD measurements using 3D QCT

被引:42
|
作者
Kang, Y [1 ]
Engelke, K [1 ]
Fuchs, C [1 ]
Kalender, WA [1 ]
机构
[1] Univ Erlangen Nurnberg, Inst Med Phys, D-91054 Erlangen, Germany
关键词
anatomic 3D coordinate system; proximal femur; femoral neck; bone mineral density; cortical thickness;
D O I
10.1016/j.compmedimag.2005.05.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper, a procedure for the determination of an anatomically oriented coordinate system of the femoral neck (NCS) in 3D spiral CT datasets is described. The origin of the NCS is centered in the minimal cross-sectional area of the neck. Its three axes are defined as follows: the so called neck axis is perpendicular to this area and points towards the femoral head, the second axis is the principal axis of the minimal cross-sectional area and the third axis is perpendicular to the other two. After a semi-automatic 3D segmentation of the proximal femur the NCS is automatically determined in a two-step minimization procedure. Relative to the coordinate system volumes of interest (VOIs) are positioned in which bone mineral density (BMD) and cortical thickness are analyzed. We investigated intra- and inter-operator precision of the position of the NCS, the BMD in cortical and trabecular VOls, and cortical thickness in nine pelvic CT datasets obtained from clinical routine examinations. We further investigated the effect of increased noise by adding Gaussian distributed noise to measured projections before tomographic reconstruction. The mean precision error (averaged form the results of the nine datasets) of the NCS position was less than 0.5 mm and smaller than 2.25 degrees. There were no significant differences between inter- and intra-operator analyses. Precision errors in trabecular BNID were smaller than 3% in a stack of five 1 mm thin slices cut perpendicularly to the neck axis and smaller than 1% in a spherical VOI encompassing the neck. Relative precision errors for cortical BNID were smaller than 3% for both VOIs. An increase of noise up to a factor of 5 caused a maximal displacement of the NCS origin position by less than I mm and a rotation by less than 2 degrees. (c) 2005 Elsevier Ltd. All rights reserved.
引用
收藏
页码:533 / 541
页数:9
相关论文
共 50 条
  • [1] QCT of the proximal femur: 3D segmentation and the definition of a neck coordinate system for reproducible BMD analysis
    Engelke, K
    Kang, Y
    Kalender, WA
    OSTEOPOROSIS INTERNATIONAL, 2000, 11 : 8 - 8
  • [2] Comparison of anatomic coordinate systems with rigid multi-resolution 3D registration for the reproducible positioning of analysis volumes of interest in QCT
    Eisa, Fabian
    Museyko, Oleg
    Hess, Andreas
    Kalender, Willi A.
    Engelke, Klaus
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05): : 1429 - 1439
  • [3] 3D femoral neck anteversion measurements based on the posterior femoral plane in ORTHODOC® system
    Yeon Soo Lee
    Seung Hoon Oh
    Jong Keun Seon
    Eun Kyoo Song
    Taek Rim Yoon
    Medical & Biological Engineering & Computing, 2007, 45 : 325 - 325
  • [4] 3D femoral neck anteversion measurements based on the posterior femoral plane in ORTHODOC® system
    Yeon Soo Lee
    Seung Hoon Oh
    Jong Keun Seon
    Eun Kyoo Song
    Taek Rim Yoon
    Medical and Biological Engineering and Computing, 2006, 44
  • [5] 3D femoral neck anteversion measurements based on the posterior femoral plane in ORTHODOC® system
    Lee, Yeon Soo
    SH, Oh
    Seon, Jong Keun
    Song, Eun Kyoo
    Yoon, Taek Rim
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2006, 44 (10) : 895 - 906
  • [6] 3D QCT BMD of the hip: Regional comparison with DXA.
    Cann, CE
    Sanchez, S
    Arnaud, CD
    Roe, B
    delPuerto, G
    Pierini, E
    JOURNAL OF BONE AND MINERAL RESEARCH, 1996, 11 : T717 - T717
  • [7] A System for Reproducible 3D Ultrasound Measurements of Skeletal Muscles
    Sahrmann, Annika S.
    Handsfield, Geoffrey G.
    Gizzi, Leonardo
    Gerlach, Jenin
    Verl, Alexander
    Besier, Thor F.
    Rohrle, Oliver
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2024, 71 (07) : 2022 - 2032
  • [8] PROSPECTIVE COMPARISON OF QCT BMD MEASUREMENT USING EITHER ASYNCHRONOUS OR SIMULTANEOUS CALIBRATION AT THE FEMORAL NECK AND LUMBAR SPINE
    Bodeen, G.
    Brown, J. K.
    Brett, A.
    OSTEOPOROSIS INTERNATIONAL, 2014, 25 : S294 - S294
  • [9] PROSPECTIVE COMPARISON OF QCT BMD MEASUREMENT USING EITHER ASYNCHRONOUS OR SIMULTANEOUS CALIBRATION AT THE FEMORAL NECK AND LUMBAR SPINE
    Bodeen, Gabriel
    Brown, J. Keenan
    Brett, Alan
    OSTEOPOROSIS INTERNATIONAL, 2014, 25 : S507 - S507
  • [10] Comparison of Femoral Neck BMD Evaluation Obtained Using Lunar DXA and QCT With Asynchronous Calibration From CT Colonography
    Pickhardt, Perry J.
    Bodeen, Gabriel
    Brett, Alan
    Brown, J. Keenan
    Binkley, Neil
    JOURNAL OF CLINICAL DENSITOMETRY, 2015, 18 (01) : 5 - 12