Background: Tc-99m is a widely used radioisotope in nuclear medicine centers which is obtained by elution from Mo-99/Tc-99m generators. Usually the generators are either supplied by the Iran Atomic Energy Agency or by private companies from foreign countries. In this study we have measured Mo-99 contamination in Tc-99m elute from different generators in a period of one year. Materials and Methods: The radionuclide impurity of the Tc-99m elute were studied in two types of radionuclide generators (A: produced in Iran and B: Imported from other country). In-vitro measurements were performed using dose calibrator. Direct measurements were made, using a standard canister at the time of milking of the generators and also in subsequent hours after milking. Results: The results showed a mean of Mo-99 impurity in generators A and B to be 0.00932 +/- 0.0043 and 0.0170 +/- 0.0127 respectively. Although the results showed that the Mo-99 contamination in Tc-99m elute was lesser than the maximum accepted activity limit of 0.015%, the difference in these two types may reflect different methods of productions of generator, as well as the quality control procedures. Conclusion: The mean of Mo-99 contamination in generators produced in Iran Atomic Energy Organization was lesser than generators imported from foreign origin. Iran. J. Radiat. Res., 2010; 8 (1): 31-35
引用
收藏
页码:31 / 35
页数:5
相关论文
共 50 条
  • [41] 99Mo/99mTc Radioisotope Generator Based on Zirconium Selenomolybdate as Column Matrix
    El-Amir, M. A.
    Ramadan, H. E.
    El-Said, H.
    Radwan, N. G.
    El-Sonbati, A. Z.
    [J]. RADIOCHEMISTRY, 2019, 61 (01) : 86 - 92
  • [42] Diversification in the Supply Chain of 99Mo Ensures a Future for 99mTc
    Cutler, Cathy S.
    Schwarz, Sally W.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (07) : 1208 - 1213
  • [43] 99Mo/99mTc Generator Shortage: Free, Web-Based Software
    Cecchin, Diego
    Zucchetta, Pietro
    Bolla, Enrico
    Faggin, Paolo
    Bui, Franco
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (08) : 14N - 15N
  • [44] REVIEW OF THE 99MO/99MTC SUPPLY SITUATION AND NEW PROJECTS
    Ponsard, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2012, 102 : S169 - S169
  • [45] An Alumina Based 99Mo–99mTc Generator to Produce 99mTc in Organic Medium Suitable for Industrial Radiotracer Applications
    Ramu Ram
    Rubel Chakravarty
    Yugandhara Pamale
    Asuhtosh Dash
    Meera Venkatesh
    [J]. Chromatographia, 2009, 69 : 497 - 501
  • [46] An Automated 99mTc Generator Device Using Low Specific Activity 99Mo
    Yang, Guogui
    Hu, Ji
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [47] The application of atomic layer deposition in the production of sorbents for 99Mo/99mTc generator
    Moret, J. L. T. M.
    Alkemade, J.
    Upcraft, T. M.
    Oehlke, E.
    Wolterbeek, H. T.
    van Ommen, J. R.
    Denkova, A. G.
    [J]. APPLIED RADIATION AND ISOTOPES, 2020, 164
  • [48] Accelerator based 99Mo/99mTc production activity in Armenia
    Avagyan, R.
    Avetisyan, A.
    Kerobyan, I.
    Dallakyan, R.
    [J]. NUCLEAR MEDICINE AND BIOLOGY, 2014, 41 (07) : 648 - 649
  • [49] Preparation and biological profile of 99mTc-lidocaine as a cardioselective imaging agent using 99mTc eluted from 99Mo/99mTc generator based on Al–Mo gel
    Tamer M. Sakr
    A. B. Ibrahim
    T. W. Fasih
    H. M. Rashed
    [J]. Journal of Radioanalytical and Nuclear Chemistry, 2017, 314 : 2091 - 2098
  • [50] ANALYSES ON PRODUCTION OF 99Mo AND 99mTc ISOTOPES FOR NUCLEAR MEDICINE
    Nikolaichuk, L. I.
    Popov, V. A.
    Tutubalin, A. I.
    Krivchenko, O. V.
    Shepelev, A. G.
    [J]. PROBLEMS OF ATOMIC SCIENCE AND TECHNOLOGY, 2012, (04): : 160 - 162