Comparison of film measurements and Monte Carlo simulations of dose delivered with very high-energy electron beams in a polystyrene phantom

被引:40
|
作者
Bazalova-Carter, Magdalena [1 ]
Liu, Michael [1 ]
Palma, Bianey [1 ]
Dunning, Michael [2 ]
McCormick, Doug [2 ]
Hemsing, Erik [2 ]
Nelson, Janice [2 ]
Jobe, Keith [2 ]
Colby, Eric [2 ]
Koong, Albert C. [1 ]
Tantawi, Sami [2 ]
Dolgashev, Valery [2 ]
Maxim, Peter G. [1 ]
Loo, Billy W., Jr. [1 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Stanford, CA 94305 USA
[2] SLAC Natl Accelerator Lab, Menlo Pk, CA 94025 USA
关键词
Monte Carlo; dose calculations; very high-energy electrons; DEPENDENCE; CODE; PHOTON; EBT;
D O I
10.1118/1.4914371
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To measure radiation dose in a water-equivalent medium from very high-energy electron (VHEE) beams and make comparisons to Monte Carlo (MC) simulation results. Methods: Dose in a polystyrene phantom delivered by an experimental VHEE beam line was measured with Gafchromic films for three 50 MeV and two 70 MeV Gaussian beams of 4.0-6.9 mm FWHM and compared to corresponding MC-simulated dose distributions. MC dose in the polystyrene phantom was calculated with the EGSnrc/BEAMnrc and DOSXYZnrc codes based on the experimental setup. Additionally, the effect of 2% beam energy measurement uncertainty and possible non-zero beam angular spread on MC dose distributions was evaluated. Results: MC simulated percentage depth dose (PDD) curves agreed with measurements within 4% for all beam sizes at both 50 and 70 MeV VHEE beams. Central axis PDD at 8 cm depth ranged from 14% to 19% for the 5.4-6.9 mm 50 MeV beams and it ranged from 14% to 18% for the 4.0-4.5 mm 70 MeV beams. MC simulated relative beam profiles of regularly shaped Gaussian beams evaluated at depths of 0.64 to 7.46 cm agreed with measurements to within 5%. A 2% beam energy uncertainty and 0.286. beam angular spread corresponded to a maximum 3.0% and 3.8% difference in depth dose curves of the 50 and 70 MeV electron beams, respectively. Absolute dose differences between MC simulations and film measurements of regularly shaped Gaussian beams were between 10% and 42%. Conclusions: The authors demonstrate that relative dose distributions for VHEE beams of 50-70 MeV can be measured with Gafchromic films and modeled with Monte Carlo simulations to an accuracy of 5%. The reported absolute dose differences likely caused by imperfect beam steering and subsequent charge loss revealed the importance of accurate VHEE beam control and diagnostics. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:1606 / 1613
页数:8
相关论文
共 50 条
  • [31] Experimental verification of a commercial Monte Carlo-based dose calculation module for high-energy photon beams
    Kuenzler, Thomas
    Fotina, Irina
    Stock, Markus
    Georg, Dietmar
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (24): : 7363 - 7377
  • [32] Monte Carlo commissioning for electron beams based on large-field dose measurements
    Schreiber, E
    Faddegon, B
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1353 - 1353
  • [33] Optimizing focused very-high-energy electron beams for radiation therapy based on Monte Carlo simulation
    Chaofan An
    Wei Zhang
    Zeyi Dai
    Jia Li
    Xiong Yang
    Jike Wang
    Yuancun Nie
    [J]. Scientific Reports, 14 (1)
  • [34] Evaluation of Very High-Energy Electron (VHEE) Beams in Comparison to VMAT and PBS Treatment Plans
    Schueler, E.
    Eriksson, K.
    Hynning, E.
    Loo, B.
    Maxim, P.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3766 - 3766
  • [35] Treatment planning for radiotherapy with very high-energy electron beams and comparison of VHEE and VMAT plans
    Bazalova-Carter, Magdalena
    Qu, Bradley
    Palma, Bianey
    Hardemark, Bjorn
    Hynning, Elin
    Jensen, Christopher
    Maxim, Peter G.
    Loo, Billy W., Jr.
    [J]. MEDICAL PHYSICS, 2015, 42 (05) : 2615 - 2625
  • [36] Improved Electron Monte Carlo Dose Calculation for Low Energy Electron Beams Using EMC
    Fix, M. K.
    Frei, D.
    Volken, W.
    Neuenschwander, H.
    Born, E. J.
    Manser, P.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [37] VARIATIONS IN IONOMETRIC DOSE MEASUREMENTS IN HIGH-ENERGY X-RAY AND ELECTRON-BEAMS
    HOLT, JG
    FLEISCHMAN, RC
    BUFFA, A
    [J]. MEDICAL PHYSICS, 1981, 8 (04) : 560 - 561
  • [38] Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams
    Ravichandran, Ramamoorthy
    Binukumar, John Pichy
    Al Amri, Iqbal
    Davis, Cheriyathmanjiyil Antony
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 291 - 303
  • [39] ABSORBED DOSE AND CAVITY IONIZATION FOR HIGH-ENERGY ELECTRON BEAMS
    KESSARIS, ND
    [J]. RADIATION RESEARCH, 1970, 43 (02) : 288 - +
  • [40] The response of RadioChromic EBT film in high-energy electron beams
    Gerbi, B. J.
    Han, E. Y.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2144 - 2144