Validation of the 3-lead tele-ECG versus the 12-lead tele-ECG and the conventional 12-lead ECG method in older people

被引:13
|
作者
Antonicelli, Roberto [1 ]
Ripa, Chiara [1 ]
Abbatecola, Angela M. [2 ]
Capparuccia, Carlo A. [1 ]
Ferrara, Letizia [3 ]
Spazzafumo, Liana [4 ]
机构
[1] Italian Natl Res Ctr Aging, Telemed Ctr, Dept Cardiol, Ancona, Italy
[2] Italian Natl Res Ctr Aging, Sci Direct Dept, Ancona, Italy
[3] Italian Natl Res Ctr Aging, Med Direct Dept, Ancona, Italy
[4] Italian Natl Res Ctr Aging, Dept Gerontol Res, Stat & Biometry Ctr, Ancona, Italy
关键词
CONGESTIVE-HEART-FAILURE; ATRIAL-FIBRILLATION; ELDERLY-PATIENTS; ELECTROCARDIOGRAM; PALPITATIONS; TRANSMISSION; ARRHYTHMIAS; GUIDELINES; MANAGEMENT; AGREEMENT;
D O I
10.1258/jtt.2011.110613
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
We compared the diagnostic accuracy of 3-lead tele-ECGs to both 12-lead tele-ECGs and a conventional 12-lead ECG device (the gold standard). The subjects were older people (n = 107) with a mean age of 66 years. The overall agreement between two cardiologists on interpreting the 3-lead tele-ECG recordings was 97% (kappa = 0.96, P < 0.001) and it was 91% for the 12-lead tele-ECGs (kappa = 0.90, P < 0.001). We also found excellent agreement on the interpretation of the 3-lead tele-ECGs compared to the gold standard: 98% (kappa = 0.96, P < 0.001) and for the 12-lead tele-ECG compared to the gold standard: 98% (kappa = 0.96, P < 0.001). Bland-Altman plots showed that the apparent differences between the techniques were not clinically relevant. The use of a 3-lead tele-ECG device may be useful for reducing the delay in treating specific heart disease conditions, e.g. in older people affected by chronic heart disease who need frequent ECG monitoring.
引用
收藏
页码:104 / 108
页数:5
相关论文
共 50 条
  • [21] The 12-lead ECG in patients with Mahaim fibers
    Sternick, EB
    Gerken, LM
    [J]. ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2006, 11 (01) : 63 - 83
  • [22] An intelligent framework for the classification of the 12-lead ECG
    Nugent, CD
    Webb, JAC
    Black, ND
    Wright, GTH
    McIntyre, M
    [J]. ARTIFICIAL INTELLIGENCE IN MEDICINE, 1999, 16 (03) : 205 - 222
  • [23] The 12-lead ECG: a continuous reference for the cardiologist
    Boriani, Giuseppe
    Vitolo, Marco
    [J]. JOURNAL OF CARDIOVASCULAR MEDICINE, 2019, 20 (07) : 459 - 463
  • [24] Prehospital 12-lead ECG: Efficacy or effectiveness?
    Swor, Robert
    Hegerberg, Stacey
    McHugh-McNally, Ann
    Goldstein, Mark
    McEachin, Christine C.
    [J]. PREHOSPITAL EMERGENCY CARE, 2006, 10 (03) : 374 - 377
  • [25] A New 12-Lead ECG Prognostic Score
    Soofi, Muhammad
    Jain, Nikhil A.
    Myers, Jonathan
    Froelicher, V. F.
    [J]. ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2015, 20 (06) : 554 - 560
  • [26] Prehospital 12-lead ECG diagnostic programs
    Urban, MJ
    Edmondson, DA
    Aufderheide, TP
    [J]. EMERGENCY MEDICINE CLINICS OF NORTH AMERICA, 2002, 20 (04) : 825 - +
  • [27] A Portable 12-Lead ECG Acquisition System
    Sun, Cheng
    Liao, Jingsheng
    Wang, Gang
    Li, Baopu
    Meng, Max Q-H
    [J]. 2013 IEEE INTERNATIONAL CONFERENCE ON INFORMATION AND AUTOMATION (ICIA), 2013, : 368 - 373
  • [28] Validation of an artificial intelligence model for 12-lead ECG interpretation
    Demolder, A.
    Herman, R.
    Vavrik, B.
    Martonak, M.
    Boza, V.
    Herman, M.
    Palus, T.
    Kresnakova, V.
    Bahyl, J.
    Iring, A.
    Nelis, O.
    Fabbricatore, D.
    Perl, L.
    Hatala, R.
    Bartunek, J.
    [J]. EUROPEAN HEART JOURNAL, 2023, 44
  • [29] Accurate and reliable 3-lead to 12-lead ECG reconstruction methodology for remote health monitoring applications
    Maheshwari, S.
    Acharyya, A.
    Rajalakshmi, P.
    Puddu, P. E.
    Schiariti, M.
    [J]. IRBM, 2014, 35 (06) : 341 - 350
  • [30] Detection of Unicolor ECG Electrode Reversals in Standard 12-Lead ECG
    Jekova, Irena
    Leber, Remo
    Krasteva, Vessela
    Schmid, Ramun
    [J]. 2018 COMPUTING IN CARDIOLOGY CONFERENCE (CINC), 2018, 45