QTc interval on 24-hour holter monitor: To trust or not to trust?

被引:4
|
作者
Agha, Mhd Khalil Tamr [1 ]
Fakhri, Ghina [2 ]
Ahmed, Mohamed [3 ]
El Sedawy, Omar [2 ]
Saleh, Bernard Abi [4 ]
Bitar, Fadi [2 ]
Arabi, Mariam [2 ]
机构
[1] Childrens Mercy Kansas City, Div Pulm & Sleep Med, Dept Pediat, Kansas City, MO USA
[2] Amer Univ Beirut, Dept Pediat & Adolescent Med, Div Cardiol, Med Ctr, Beirut, Lebanon
[3] Amer Univ Beirut, Dept Family Med, Med Ctr, Beirut, Lebanon
[4] Amer Univ Beirut, Dept Internal Med, Div Cardiol, Med Ctr, Beirut, Lebanon
关键词
conduction disturbances; event recorder; holter; pediatric electrophysiology; QT dispersion; HEART-RATE; DIAGNOSTIC PERFORMANCE; CORRECTION FORMULAS; LONG; ELECTROCARDIOGRAPHY; DURATION;
D O I
10.1111/anec.12899
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction QT interval represents the ventricular depolarization and repolarization. Its accurate measurement is critical since prolonged QT can lead to sudden cardiac death. QT is affected by heart rate and is corrected to QTc via several formulae. QTc is commonly calculated on the ECG and not the 24-h Holter. Methods 100 patients presenting to our institution were evaluated by an ECG followed by a 24-h Holter. QTc measurement on both platforms using Bazett and Fridericia formulae was recorded and compared. Results Mean age was 14.09 years, with the majority being males. Mean heart rate was 125.87. In the ECG, the mean QTc interval via the Bazett formula was 0.40 s compared with 0.38 s using the Fridericia formula. The mean corrected QT via the Bazett formula was 0.45, 0.39, and 0.42 s for the shortest RR, the longest RR, and the average RR, respectively. In contrast to the Fridericia formula, the corrected QT interval was 0.40, 0.39, and 0.40 s for the shortest RR, the longest RR, and the average RR, respectively. Using the Bazett formula, the highest specificity was reached during the longest RR interval (92.2%), while the highest sensitivity was recorded during the shortest RR interval (40%). As for the Fridericiaformula, sensitivity always reached 0%, while the highest specificity was reached during the average RR interval. Conclusion QTc measured during Holter ECG reached a high specificity regardless of RR interval using the Fridericia and during the longest and the average RR interval for the Bazett formula. The consistently low sensitivity reveals that Holter ECG should not be used to rule out prolonged QT.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] US panel backs 24-hour glucose monitor
    Ault, A
    [J]. LANCET, 1999, 353 (9155): : 822 - 822
  • [22] EFFECT OF NISOLDIPINE ON VARIANT ANGINA - ASSESSMENT BY 24-HOUR HOLTER MONITORING
    MORIMOTO, SI
    MIZUNO, Y
    HIRASAWA, K
    HOSODA, S
    HIRAMORI, K
    HAZE, K
    HAYASAKI, K
    HIROSAWA, K
    [J]. CURRENT THERAPEUTIC RESEARCH-CLINICAL AND EXPERIMENTAL, 1992, 52 (04): : 548 - 555
  • [23] EFFECTS OF EPANOLOL AND METOPROLOL ON THE HEART MEASURED BY 24-HOUR HOLTER MONITORING
    WITHAGEN, AJAM
    CORBEIJ, HMA
    HUIGE, MC
    KRAGTEN, JA
    VERMEULEN, A
    [J]. DRUGS, 1989, 38 : 67 - 69
  • [24] Diagnostic and prognostic efficiency of 24-hour Holter monitoring in patients with syncope
    Barbeito-Caamano, Cayetana
    Sanchez-Fernandez, Gabriel
    Bouzas-Mosquera, Alberto
    Broullón, Francisco J.
    Alvarez-Garcia, Nemesio
    Manuel Vazquez-Rodriguez, Jose
    [J]. MEDICINA CLINICA, 2016, 147 (04): : 148 - 150
  • [25] HOLTER MONITORING OF ANESTHETISTS DURING A 24-HOUR ON-CALL PERIOD
    HARDER, KF
    FAY, J
    SHELLEY, ES
    [J]. CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1985, 32 (03) : S66 - S66
  • [26] Cardiac safety of ribociclib evaluated with 24-hour rhythm Holter electrocardiogram
    Yasin, Ayse Irem
    Uluganyan, Mahmut
    Isleyen, Zehra Sucuoglu
    Topcu, Atakan
    Shbair, Abdallah Tm
    Simsek, Melih
    Besiroglu, Mehmet
    Ersoy, Yeliz Emine
    Turk, Haci Mehmet
    Seker, Mesut
    [J]. SUPPORTIVE CARE IN CANCER, 2024, 32 (08)
  • [27] Establishing 24-hour Holter reference intervals for clinically healthy puppies
    Lee, Pamela M.
    Brown, Rebecca H. T.
    [J]. RESEARCH IN VETERINARY SCIENCE, 2019, 125 : 253 - 255
  • [28] Diagnostic and prognostic performance of 24-hour Holter in patients with unrecorded syncope
    Shihadeh, Leydimar Anmad
    Lorenzo-Gonzalez, Javier
    Fernandez-Rodriguez, Diego
    [J]. MEDICINA CLINICA, 2017, 148 (10): : 479 - 479
  • [29] QT interval and autonomic nerve activity in long QT syndrome: Analysis using 24-hour Holter ECG monitoring
    Fujiki, A
    Shimono, M
    Hayashi, H
    Usui, M
    [J]. CIRCULATION, 1996, 94 (08) : 2963 - 2963
  • [30] Diagnostic and prognostic performance of 24-hour Holter in patients with unrecorded syncope Reply
    Barbeito-Caamano, Cayetana
    Bouzas-Mosquera, Alberto
    Sanchez-Fernandez, Gabriel
    [J]. MEDICINA CLINICA, 2017, 148 (10): : 480 - 480