摘要
La1-xMgxOCl1-x (x = 0.025, 0.05 and 0.1) and La0.9Ca0.1OCl0.9 samples were prepared by a solid state reaction method. The best preparation conditions were determined by means of the differential thermal analysis and thermogravimetry (DTA&TG). The samples were studied by X-ray diffraction (XRD) analysis. Electrical conductivity of La0.9Mg0.1OCl0.1 was measured using electrical impedance spectroscopy (EIS). The prototype chlorine gas sensor was constructed from La0.9Mg0.1OCl0.1 solid electrolyte. The source voltage (SV) of the constructed device was measured as a function of temperature (500-800 degrees C), Cl-2 gradient partial pressure and time. Similar measurements of SV was carried out in atmosphere of synthetic air in argon.
引用
收藏
页码:171 / 174
页数:4
相关论文
共 50 条
  • [11] Influence of moisture on potentiometric Cl2 gas sensor using a Na+ conducting solid electrolyte
    Aono, H
    Sadaoka, Y
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2002, 149 (06) : H123 - H127
  • [12] A solid electrolyte sensor for trace gas analysis
    Schelter, Matthias
    Zosel, Jens
    Oelssner, Wolfram
    Guth, Ulrich
    Mertig, Michael
    SENSORS AND ACTUATORS B-CHEMICAL, 2013, 187 : 209 - 214
  • [13] Lisicon solid electrolyte electrocatalytic gas sensor
    Jasinski, G
    Jasinski, P
    Chachulski, B
    Nowakowski, A
    JOURNAL OF THE EUROPEAN CERAMIC SOCIETY, 2005, 25 (12) : 2969 - 2972
  • [14] Potentiometric CO2 gas sensor with lithium ion electrolyte
    Lee, CH
    Akbar, S
    Park, CO
    PROCESSING AND FABRICATION OF ADVANCED MATERIALS XI, 2003, : 63 - 77
  • [15] SOLID-STATE ELECTROCHEMICAL CHLORINE GAS SENSOR
    SCHLECHTRIEMEN, GL
    SOHEGE, J
    BERICHTE DER BUNSEN-GESELLSCHAFT-PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 1990, 94 (09): : 991 - 995
  • [16] Influence of the Solid-Electrolyte Ionic Material in a Potentiometric Sensor for Ethylene Detection
    Toldra-Reig, Fidel
    Pastor, Daniel
    Serra, Jose M.
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2019, 166 (14) : B1343 - B1355
  • [18] Potentiometric CO2 gas sensor with lithium phosphorous oxynitride electrolyte
    Lee, C
    Akbar, SA
    Park, CO
    SENSORS AND ACTUATORS B-CHEMICAL, 2001, 80 (03) : 234 - 242
  • [19] AUTOMATED-DETERMINATION OF AMMONIA WITH A POTENTIOMETRIC GAS SENSOR AND FLOWING INTERNAL ELECTROLYTE
    FRATICELLI, YM
    MEYERHOFF, ME
    ANALYTICAL CHEMISTRY, 1981, 53 (07) : 992 - 997
  • [20] Organic gas sensor using BiCuVOx solid electrolyte
    Kida, Tetsuya
    Minami, Takuya
    Yuasa, Masayoshi
    Shimanoe, Kengo
    Yamazoe, Noboru
    ELECTROCHEMISTRY COMMUNICATIONS, 2008, 10 (02) : 311 - 314