Optimization of treatment isocenter location in single-isocenter LINAC-based stereotactic radiosurgery for management of multiple brain metastases

被引:10
|
作者
Cui, Taoran [1 ]
Zhou, Yongkang [2 ]
Yue, Ning J. [1 ]
Vergalasova, Irina [1 ]
Zhang, Yin [1 ]
Zhu, Jiahua [1 ]
Nie, Ke [1 ]
机构
[1] Rutgers Canc Inst New Jersey, Dept Radiat Oncol, New Brunswick, NJ 08901 USA
[2] Zhongshan Hosp, Dept Radiat Oncol, Shanghai, Peoples R China
关键词
UNCERTAINTIES; RADIOTHERAPY; NUMBER;
D O I
10.1002/mp.15294
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Single-isocenter linear accelerator (LINAC)-based stereotactic radiosurgery (SRS) has become a promising treatment technique for the management of multiple brain metastases. Because of the high prescription dose and steep dose gradient, SRS plans are sensitive to geometric errors, resulting in loss of target coverage and suboptimal local tumor control. Current planning techniques rely on adding a uniform and isotropic setup margin to all gross tumor volumes (GTVs) to account for rotational uncertainties. However, this setup margin may be insufficient, since the magnitude of rotational uncertainties varies and is dependent upon the distance between a GTV and the isocenter. In this study, we designed a framework to determine the optimal isocenter of a single-isocenter SRS plan for multiple brain metastases using stochastic optimization to mitigate potential errors resulting from rotational uncertainties. Methods Planning target volumes (PTVs), defined as GTVs plus a 1-mm margin following common SRS planning convention, were assumed to be originally treated with a prescription dose and therefore covered by the prescription isodose cloud. The dose distribution, including the prescription isodose, was considered invariant assuming small rotations throughout the study. A stochastic optimization scheme was developed to determine the location of the optimal isocenter, so that the prescription dose coverage of rotated GTVs, equivalent to the intersecting volumes between the rotated GTVs and original PTVs, was maximized for any random small rotations about the isocenter. To evaluate the coverage of GTVs, the expected V100% undergoing random rotations was approximated as the sample average V100% undergoing a predetermined number of rotations. The expected V100% of each individual GTV and total GTVs was then compared between the plans using the optimal isocenter and the center-of-mass (CoM), respectively. Results Twenty-two patients previously treated for multiple brain metastases in a single institute were included in this retrospective study. Each patient was initially treated for more than three brain metastases (mean: 7.6; range: 3-15) with the average GTV volume of 0.89 cc (range: 0.03-11.78 cc). The optimal isocenter found for each patient was significantly different from the CoM, with the average Euclidean distance between the optimal isocenter and the CoM being 4.36 +/- 2.59 cm. The dose coverage to GTVs was also significantly improved (paired t-test; p < 0.001) when the optimal isocenter was used, with the average V100% of total GTVs increasing from 87.1% (standard deviation as std: 11.7%; range: 39.9-98.2%) to 94.2% (std: 5.4%; range: 77.7-99.4%). The volume of a GTV was positively correlated with the expected V100% regardless of the isocenter used (Spearman coefficient: rho=0.66; p < 0.001). The distance between a GTV and the isocenter was negatively correlated with the expected V100% when the CoM was used (rho=-0.21; p = 0.004), however no significant correlation was found when the optimal isocenter was used (rho=-0.11; p = 0.137). Conclusion The proposed framework provides an effective approach to determine the optimal isocenter of single-isocenter LINAC-based SRS plans for multiple brain metastases. The implementation of the optimal isocenter results in SRS plans with consistently higher target coverage despite potential rotational uncertainties, and therefore significantly improves SRS plan robustness against random rotational uncertainties.
引用
收藏
页码:7632 / 7640
页数:9
相关论文
共 50 条
  • [1] Biological Plan Evaluation for Linac-Based Stereotactic Radiosurgery: Comparison Between Single-Isocenter and Multiple-Isocenter Techniques
    Lee, M.
    Wang, H.
    Tsai, H.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E711 - E711
  • [2] Single-Isocenter Linac-Based Radiosurgery for Brain Metastases with Coplanar Arcs: A Dosimetric and Clinical Analysis
    Faccenda, Valeria
    Panizza, Denis
    Pisoni, Valerio
    Trivellato, Sara
    Daniotti, Martina Camilla
    Bianchi, Sofia Paola
    De Ponti, Elena
    Arcangeli, Stefano
    [J]. CANCERS, 2023, 15 (18)
  • [3] Single-Isocenter Multitarget Stereotactic Radiosurgery Is Safe and Effective in the Treatment of Multiple Brain Metastases
    Palmer, Joshua D.
    Sebastian, Nikhil T.
    Chu, Jacquline
    DiCostanzo, Dominic
    Bell, Erica H.
    Grecula, John
    Arnett, Andrea
    Blakaj, Dukagjin M.
    McGregor, John
    Elder, James B.
    Lu, Lanchun
    Zoller, Wesley
    Addington, Mark
    Lonser, Russell
    Chakravarti, Arnab
    Brown, Paul D.
    Raval, Raju
    [J]. ADVANCES IN RADIATION ONCOLOGY, 2020, 5 (01) : 70 - 76
  • [4] Single-Isocenter Multitarget Stereotactic Radiosurgery is Safe and Effective in the Treatment of Multiple Brain Metastases
    Sebastian, N.
    Raval, R.
    Chu, J.
    DiCostanzo, D. J.
    Bell, E. H.
    Grecula, J. C.
    Arnett, A. L. H.
    Blakaj, D. M.
    McGregor, J.
    Elder, J. B.
    Lonser, R. R.
    Chakravarti, A.
    Brown, P. D.
    Palmer, J. D.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E89 - E89
  • [5] Knowledge-Based Planning for Single-Isocenter Stereotactic Radiosurgery to Multiple Brain Metastases
    Ziemer, B.
    Shiraishi, S.
    Hattangadi-Gluth, J.
    Sanghvi, P.
    Moore, K.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3724 - 3725
  • [6] Clinical Outcomes of Single-Isocenter Versus Multiple-Isocenter Stereotactic Radiosurgery Techniques for Multiple Brain Metastases
    Deng, J.
    Li, M. K.
    Savjani, R. R.
    Chu, F. I.
    Tenn, S. E.
    Lee, C.
    Agazaryan, N.
    Yang, I.
    Everson, R.
    Kim, W.
    Pouratian, N.
    Kishan, A. U.
    Chin, R. K.
    Steinberg, M. L.
    Kaprealian, T. B.
    Hegde, J. V.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E566 - E567
  • [7] Stereotactic radiotherapy with single-isocenter in multiple brain metastases
    Matteucci, P.
    Silipigni, S.
    Rinaldi, C. G.
    Ridolfi, I.
    Falco, P.
    Fiore, M.
    Ippolito, E.
    Ramella, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S866 - S867
  • [8] Stereotactic radiosurgery with single-isocenter dynamic arc therapy of multiple brain metastases
    Bodensohn, R.
    Kaempfel, A.
    Fleischmann, D. F.
    Hofmaier, J.
    Hadi, I.
    Garny, S.
    Reiner, M.
    Corradini, S.
    Forbrig, R.
    Thon, N.
    Belka, C.
    Niyazi, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S470 - S471
  • [9] Optimization of isocenter position for multiple brain metastases single-isocenter stereotactic radiosurgery to minimize dosimetric variations due to rotational uncertainty
    Shen, Zhenjiong
    wang, Hao
    Shao, Yan
    Duan, Yanhua
    Gu, Hengle
    Chen, Hua
    Feng, Aihui
    Huang, Ying
    Xu, Zhiyong
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 111
  • [10] Distance to isocenter is not associated with an increased risk for local failure in LINAC-based single-isocenter SRS or SRT for multiple brain metastases
    Kraft, Johannes
    van Timmeren, Janita E.
    Mayinger, Michael
    Frei, Simon
    Borsky, Kim
    Stark, Luisa Sabrina
    Krayenbuehl, Jerome
    Zamburlini, Mariangela
    Guckenberger, Matthias
    Tanadini-Lang, Stephanie
    Andratschke, Nicolaus