Improvements in CBCT Image Quality Using a Novel Iterative Reconstruction Algorithm: A Clinical Evaluation

被引:54
|
作者
Gardner, Stephen J. [1 ]
Mao, Weihua [1 ]
Liu, Chang [1 ]
Aref, Ibrahim [1 ]
Elshaikh, Mohamed [1 ]
Lee, Joon K. [1 ]
Pradhan, Deepak [1 ]
Movsas, Benjamin [1 ]
Chetty, Indrin J. [1 ]
Siddiqui, Farzan [1 ]
机构
[1] Henry Ford Hlth Syst, Josephine Ford Canc Inst, Dept Radiat Oncol, 2799 W Grand Blvd, Detroit, MI 48202 USA
关键词
BEAM COMPUTED-TOMOGRAPHY; HEAD-AND-NECK; SCATTERED RADIATION; LINEAR-ACCELERATOR; CT IMAGES; DELINEATION; RADIOTHERAPY; SEGMENTATION; VARIABILITY; THERAPY;
D O I
10.1016/j.adro.2018.12.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This study aimed to evaluate the clinical utility of a novel iterative cone beam computed tomography (CBCT) reconstruction algorithm for prostate and head and neck (HN) cancer. Methods and Materials: A total of 10 patients with HN and 10 patients with prostate cancer were analyzed. For each patient, raw CBCT acquisition data were used to reconstruct images with a currently available algorithm (FDK_CBCT) and novel iterative algorithm (Iterative_CBCT). Quantitative contouring variation analysis was performed using structures delineated by several radiation oncologists. For prostate, observers contoured the prostate, proximal 2 cm seminal vesicles, bladder, and rectum. For HN, observers contoured the brain stem, spinal canal, right-left parotid glands, and right-left submandibular glands. Observer contours were combined to form a reference consensus contour using the simultaneous truth and performance level estimation method. All observer contours then were compared with the reference contour to calculate the Dice coefficient, Hausdorff distance, and mean contour distance (prostate contour only). Qualitative image quality analysis was performed using a 5-point scale ranging from 1 (much superior image quality for Iterative_CBCT) to 5 (much inferior image quality for Iterative_CBCT). Results: The Iterative_CBCT data sets resulted in a prostate contour Dice coefficient improvement of approximately 2.4% (P = .029). The average prostate contour Dice coefficient for the Iterative_CBCT data sets was improved for all patients, with improvements up to approximately 10% for 1 patient. The mean contour distance results indicate an approximate 15% reduction in mean contouring error for all prostate regions. For the parotid contours, Iterative_CBCT data sets resulted in a Hausdorff distance improvement of approximately 2 mm (P < .01) and an approximate 2% improvement in Dice coefficient (P = .03). The Iterative_CBCT data sets were scored as equivalent or of better image quality for 97.3% (prostate) and 90.0% (HN) of the patient data sets. Conclusions: Observers noted an improvement in image uniformity, noise level, and overall image quality for Iterative_CBCT data sets. In addition, expert observers displayed an improved ability to consistently delineate soft tissue structures, such as the prostate and parotid glands. Thus, the novel iterative reconstruction algorithm analyzed in this study is capable of improving the visualization for prostate and HN cancer image guided radiation therapy. (C) 2019 The Authors. Published by Elsevier Inc. on behalf of American Society for Radiation Oncology.
引用
收藏
页码:390 / 400
页数:11
相关论文
共 50 条
  • [1] Improvements in HN CBCT Image Quality Using a Novel Iterative Reconstruction Algorithm
    Gardner, S.
    Mao, W.
    Liu, C.
    Aref, I.
    Mohamed, E.
    Lee, J.
    Pradhan, D.
    Siddiqui, F.
    Movsas, B.
    Chetty, I.
    MEDICAL PHYSICS, 2019, 46 (06) : E152 - E153
  • [2] A Novel Iterative Reconstruction Algorithm for Improving CBCT Image Quality
    Mao, W.
    Liu, C.
    Snyder, K.
    Kumarasiri, A.
    Zhao, B.
    Gardner, S.
    Kim, J.
    Wen, N.
    Chetty, I.
    Siddiqui, F.
    MEDICAL PHYSICS, 2017, 44 (06) : 3148 - 3148
  • [3] Evaluation of a Metal Artifact Reduction Algorithm for Image Reconstruction on a Novel CBCT Platform
    Yashayaeva, Abby
    MacDonald, Robert Lee
    Robar, James
    Cherpak, Amanda
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (11):
  • [4] Clinical Evaluation and Implementation of a Commercial Iterative CBCT Algorithm
    Zheng, Y.
    Wong, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E356 - E357
  • [5] Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT
    Mao, Weihua
    Liu, Chang
    Gardner, Stephen J.
    Siddiqui, Farzan
    Snyder, Karen C.
    Kumarasiri, Akila
    Zhao, Bo
    Kim, Joshua
    Wen, Ning Winston
    Movsas, Benjamin
    Chetty, Indrin J.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2019, 18
  • [6] Improved Target Visualization Using a Novel Iterative CBCT Reconstruction Algorithm for Prostate Cancer RT
    Gardner, S.
    Mao, W.
    Liu, C.
    Aref, I.
    Elshaikh, M.
    Lee, J.
    Pradhan, D.
    Siddiqui, F.
    Movsas, B.
    Chetty, I.
    MEDICAL PHYSICS, 2018, 45 (06) : E485 - E486
  • [7] Image Quality and Dose Reduction Evaluation of a New CT Iterative Reconstruction Algorithm Using Model Observers
    Tseng, H.
    Fan, J.
    Kupinski, M.
    Sainath, P.
    Hsieh, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [8] Iterative image reconstruction for CBCT using edge-preserving prior
    Wang, Jing
    Li, Tianfang
    Xing, Lei
    MEDICAL PHYSICS, 2009, 36 (01) : 252 - 260
  • [9] Iterative Image Reconstruction for CBCT Using Edge-Preserving Prior
    Wang, J.
    Li, T.
    Xing, L.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [10] Improvement in Image Quality of Low Dose De-CBCT Scans Using Iterative Reconstruction Algorithms
    Georgesen, M. C.
    Keeler, A.
    Roeske, J. C.
    Luce, J. P.
    Kang, H.
    MEDICAL PHYSICS, 2024, 51 (10) : 7721 - 7721