Orthognathic Surgical Robot With a Workspace Limitation Mechanism

被引:4
|
作者
Hara, Kazuaki [1 ]
Ma, Qingchuan [1 ]
Suenaga, Hideyuki [2 ]
Kobayashi, Etsuko [3 ]
Sakuma, Ichiro [1 ]
Masamune, Ken [3 ]
机构
[1] Univ Tokyo, Grad Sch Engn, Tokyo 1138654, Japan
[2] Univ Tokyo Hosp, Dept Oral Maxillofacial Surg & Orthodont, Tokyo 1138655, Japan
[3] Tokyo Womens Med Coll, Inst Adv BioMed Engn & Sci, Tokyo 1628666, Japan
基金
日本学术振兴会;
关键词
Surgery; Gears; Medical robotics; Kinematics; Robot kinematics; Hardware; Medical robot; motion constraint; orthognathic surgery; osteotomy; surgical robot; workspace limitation; SIMULATION; SYSTEM;
D O I
10.1109/TMECH.2019.2945605
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Impracticable workspaces and cumbersome hardware limit the practical application of the currently available orthognathic surgical robots in a clinical environment. In this study, a compact and lightweight six degree of freedom (DOF) robot with a workspace limitation mechanism (WLM) was developed, and this article reports on the technical specifications of the robot, as well as the details of the three experiments performed to validate it. This robot can be easily installed by a single individual and is able to mechanically protect important organs from unintended injury. The results show that this robot functionally limits the workspace travel of the osteotomy device by an average overshoot of 0.8 mm, with a drilling and cutting accuracy of 1 and 0.5 mm, respectively. This operational accuracy has the potential to be further improved, making this novel robot a potentially precise and safe medical equipment for orthognathic surgery.
引用
收藏
页码:2652 / 2660
页数:9
相关论文
共 50 条
  • [1] Is smaller workspace a limitation for robot performance in laparoscopy?
    Thakre, A. A.
    Bailly, Y.
    Sun, L. W.
    Van Meer, F.
    Yeung, C. K.
    JOURNAL OF UROLOGY, 2008, 179 (03): : 1138 - 1142
  • [2] Is smaller workspace a limitation for robot performance in laparoscopy? Comment
    Casale, Pasquale
    JOURNAL OF UROLOGY, 2008, 179 (03): : 1142 - 1143
  • [3] Mechanism design and workspace analysis of a hexapod robot
    Li, Huayang
    Qi, Chenkun
    Gao, Feng
    Chen, Xianbao
    Zhao, Yue
    Chen, Zhijun
    MECHANISM AND MACHINE THEORY, 2022, 174
  • [4] Workspace searching method of thoracic and celiac surgical robot
    Robotics Institute, Harbin Institute of Technology, Harbin 150080, China
    Nanjing Li Gong Daxue Xuebao, 2008, 4 (476-481): : 476 - 481
  • [5] Research on workspace of a two-arm surgical robot
    Piao, Mingbo
    Fu, Yili
    Wang, Shuguo
    2007 IEEE INTERNATIONAL CONFERENCE ON MECHATRONICS AND AUTOMATION, VOLS I-V, CONFERENCE PROCEEDINGS, 2007, : 1067 - 1072
  • [6] Workspace Mapping and Control of a Teleoperated Endoscopic Surgical Robot
    Khalifa, Alaa
    Ramadan, Ahmed
    Ibrahim, Khalil
    Fanni, Mohamed
    Assal, Samy
    Abo-Ismail, Ahmed
    2014 19TH INTERNATIONAL CONFERENCE ON METHODS AND MODELS IN AUTOMATION AND ROBOTICS (MMAR), 2014, : 675 - 680
  • [7] Mechanism Synthesis and Workspace Analysis of a Spraying Robot for Airfoil
    Zhu, Lizhong
    Wang, Liping
    Zhao, Jingshan
    ADVANCES IN RECONFIGURABLE MECHANISMS AND ROBOTS II, 2016, 36 : 669 - 679
  • [8] Kinematics and Workspace of a New Surgical Robot with Five Degrees of Freedom
    Scurtu, I. L.
    Plitea, N.
    Crisan, N.
    Pisla, D.
    2012 IEEE INTERNATIONAL CONFERENCE ON AUTOMATION, QUALITY AND TESTING, ROBOTICS, THETA 18TH EDITION, 2012, : 406 - +
  • [9] Preoperative optimization of the surgical robot considering internal diversity of workspace
    Yan, Zhiyuan
    Du, Zhijiang
    Zhang, Fan
    Wang, Weidong
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 2018, 232 (06) : 1091 - 1107
  • [10] 3D WORKSPACE CONSTRUCTION OF LARYNGEAL SURGICAL ROBOT
    Chen, H. G.
    Cao, Y.
    Wang, S. L.
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2017, 121 : 12 - 12