On the practical use of calorimetry for routine absolute dosimetry in the radiotherapy clinic

被引:1
|
作者
Renaud, J. [1 ]
Sarfehnia, A. [1 ,2 ]
Seuntjens, J. [1 ]
机构
[1] McGill Univ, Med Phys Unit, Montreal, PQ, Canada
[2] Sunnybrook Res Inst, Odette Canc Res Program, Toronto, ON, Canada
关键词
Medical devices; radiotherapy; dosimetry; calorimetry; instrumentation; PHOTON;
D O I
10.1007/978-3-319-19387-8_163
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The aim of this work is to present the development and modes of operation of a small-scale graphite calorimeter probe (GPC) conceived for absolute dosimetry in the radiotherapy clinic (Patent No. PCT/CA2013/000523). Based on a numerically optimized design, a functioning prototype with embedded active thermal control capable of quasiadiabatic and isothermal operation was constructed in-house. In quasi-adiabatic mode, the radiation-induced temperature rise is measured in the sensitive volume while the device is only partially controlled. In isothermal mode, the entire device is subject to active thermal control and the quantity of interest in the electrical power necessary to maintain a stable temperature while irradiated. An experimental characterization of the GPC repeatability, linearity, and dose rate dependence while operated in quasi-adiabatic mode was carried out in a 6 MV photon field. Similarly, a characterization study of the GPC's isothermal mode was conducted by simulating various radiation fields by means of electrical power dissipation. For the quasi-adiabatic mode, a repeatability of 0.4% was established (n = 64), linearity was characterized by an adjusted R-2 value of 0.9996, and no statistically-significant dose rate dependence was observed. For isothermal mode, differences between expected and actual GPC response ranged between 0.3 - 1.4%, and in general, agreement was superior for longer irradiations with dose rates greater than 2 Gy/min. By establishing the feasibility of constructing a GPC capable of both modes of operation, this work lays the foundation for further device enhancements. The next step is the development of the GPC is to experimentally demonstrate self-consistency between the two independent modes of operation. Following this, a validation study comparing the GPC's absolute dose measurements against calibrated reference class dosimeters in both standard and non-standard radiation fields.
引用
收藏
页码:667 / 670
页数:4
相关论文
共 50 条
  • [1] Absolute dosimetry on a dynamically scanned sample for synchrotron radiotherapy using graphite calorimetry and ionization chambers
    Lye, J. E.
    Harty, P. D.
    Butler, D. J.
    Crosbie, J. C.
    Livingstone, J.
    Poole, C. M.
    Ramanathan, G.
    Wright, T.
    Stevenson, A. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (11): : 4201 - 4222
  • [2] A practical EPR dosimetry system for routine use in radiotherapy: uncertainty analysis of lithium formate dosimeters at the therapeutic dose level
    Hoefel, Sebastian
    Stehle, Michael
    Zwicker, Felix
    Fix, Michael K.
    Drescher, Malte
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (04):
  • [3] Recent developments in absolute dosimetry for FLASH radiotherapy
    Subiel, Anna
    Romano, Francesco
    BRITISH JOURNAL OF RADIOLOGY, 2023, 96 (1148):
  • [4] A PLOTTING OPTICAL DENSITOMETER FOR ROUTINE RADIOTHERAPY DOSIMETRY
    BIRCH, M
    BLOWES, RW
    PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (02): : 463 - 468
  • [5] Practical guidelines for routine intensity-modulated radiotherapy verification: pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry
    Vinall, A. J.
    Williams, A. J.
    Currie, V. E.
    Van Esch, A.
    Huyskens, D.
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (995): : 949 - 957
  • [6] Absolute dosimetry for FLASH proton pencil beam scanning radiotherapy
    Lourenco, Ana
    Subiel, Anna
    Lee, Nigel
    Flynn, Sam
    Cotterill, John
    Shipley, David
    Romano, Francesco
    Speth, Joe
    Lee, Eunsin
    Zhang, Yongbin
    Xiao, Zhiyan
    Mascia, Anthony
    Amos, Richard A.
    Palmans, Hugo
    Thomas, Russell
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [7] Absolute dosimetry for FLASH proton pencil beam scanning radiotherapy
    Ana Lourenço
    Anna Subiel
    Nigel Lee
    Sam Flynn
    John Cotterill
    David Shipley
    Francesco Romano
    Joe Speth
    Eunsin Lee
    Yongbin Zhang
    Zhiyan Xiao
    Anthony Mascia
    Richard A. Amos
    Hugo Palmans
    Russell Thomas
    Scientific Reports, 13
  • [8] BONE MODELS FOR USE IN RADIOTHERAPY DOSIMETRY
    WOODARD, HQ
    WHITE, DR
    BRITISH JOURNAL OF RADIOLOGY, 1982, 55 (652): : 277 - 282
  • [9] Clinical experience with routine diode dosimetry for electron beam radiotherapy
    Yaparpalvi, R
    Fontenla, DP
    Vikram, B
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (04): : 1259 - 1265
  • [10] Uncertainty estimation in intensity-modulated radiotherapy absolute dosimetry verification
    Sanchez-Doblado, Francisco
    Hartmann, Gunther H.
    Pena, Javier
    Capote, Roberto
    Paiusco, Marta
    Rhein, Bernhard
    Leal, Antonio
    Lagares, Juan Ignacio
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (01): : 301 - 310