Less than one minute low-pressure gas chromatography - mass spectrometry

被引:18
|
作者
Fialkov, Alexander B. [1 ]
Lehotay, Steven J. [2 ]
Amirav, Aviv [1 ,3 ]
机构
[1] Tel Aviv Univ, Sch Chem, IL-69978 Tel Aviv, Israel
[2] ARS, USDA, Eastern Reg Res Ctr, 600 East Mermaid Lane, Wyndmoor, PA 10938 USA
[3] Aviv Analyt Ltd, 24 Hanagar St, IL-4527713 Hod Hasharon, Israel
基金
以色列科学基金会;
关键词
Fast GC-MS; Low thermal mass gas chromatography; Resistively heated capillary chromatography; High-throughput analysis; FAST GC-MS; OPEN PROBE; COLUMN; OPTIMIZATION; TEMPERATURE; IONIZATION; SYSTEM;
D O I
10.1016/j.chroma.2019.460691
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Conventional gas chromatography - mass spectrometry (GC-MS) takes 20-40 min per sample, which is undesirably slow in any application if speed can be increased while still meeting analytical needs. In this study, we achieved reasonably good separations with full analysis cycle times of less than 1 min by combining for the first time low-pressure (LP) GC-MS with low thermal mass (LTM) resistive-heating for rapid temperature ramping and cooling of the capillary column. The analytical column is threaded into the LTM thin-walled metal tubing in an instrumental device known as "LTM Fast GC" that is mounted at the top of the gas chromatograph in a detector port. The column inlet and outlet are connected to the GC injector and MS transfer line as usual. For LPGC-MS, a 40 cm, 0.1 mm. i.d. uncoated flow restrictor capillary connected at the injector is coupled with a 2.6 m, 0.25 mm i.d., 0.25 gm film thickness analytical column leading to the MS. Thus, the inlet operates at normal GC pressures, but the analytical column is under vacuum, which increases the optimal helium carrier gas flow velocity thereby increasing speed of full range separations while maintaining acceptable quality of chromatography. This column configuration in LTM-LPGC-MS trades a 64-fold gain in speed of analysis vs. standard GC-MS for a 4-fold loss in chromatographic peak capacity, thereby converting analysis time from minutes into seconds in common applications. For example, jet fuel containing fatty acid methyl esters (akin to biofuel) was separated in 25 s with <1 min full analysis cycle time. An EPA Method 8270 mixture of 76 analytes was also analyzed in <1 min full cycle time by LTM-LPGC-MS. Other examples include very fast analysis of heroin in a street drug powder and elucidation of a new organic synthetic compound. In this report, we describe and discuss the several advantageous and practical features of LTM-LPGC-MS, as well as its trade-offs. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Determination of pesticides and some metabolites in different kinds of milk by solid-phase microextraction and low-pressure gas chromatography-tandem mass spectrometry
    M. J. González-Rodríguez
    F. J. Arrebola Liébanas
    A. Garrido Frenich
    J. L. Martínez Vidal
    F. J. Sánchez López
    Analytical and Bioanalytical Chemistry, 2005, 382 : 164 - 172
  • [42] LOW-PRESSURE GAS-CHROMATOGRAPHY WITH SPLIT INJECTION AND PHOTOIONIZATION DETECTION
    PUIG, L
    SACKS, RD
    JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (04) : 158 - 164
  • [43] SYNTHESIS OF PYROXFERROITE AND FE RHODONITE AT LOW-PRESSURE (LESS-THAN 1 ATM)
    RIED, H
    SCHROPFER, L
    KOREKAWA, M
    ZEITSCHRIFT FUR KRISTALLOGRAPHIE, 1979, 149 (1-2): : 121 - 123
  • [44] LOW-PRESSURE GAS DISCHARGES
    POLMAN, J
    VANTONGEREN, H
    VERBEEK, TG
    PHILIPS TECHNICAL REVIEW, 1975, 35 (11-1): : 321 - 330
  • [45] Pyrolysis gas chromatography - mass spectrometry of pressure sensitive adhesive tapes
    Germinario, Giulia
    Mirabile, Antonio
    Sabbatini, Luigia
    JOURNAL OF ANALYTICAL AND APPLIED PYROLYSIS, 2020, 151 (151)
  • [46] Single step determination of fragrances in Cucurbita flowers by coupling headspace solid-phase microextraction low-pressure gas chromatography-tandem mass spectrometry
    Granero, AM
    González, FJE
    Frenich, AG
    Sanz, JMG
    Vidal, JLM
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1045 (1-2) : 173 - 179
  • [47] Evaluation of different sample treatments for determining pesticide residues in fat vegetable matrices like avocado by low-pressure gas chromatography-tandem mass spectrometry
    Moreno, JLF
    Liébanas, FJA
    Frenich, AG
    Vidal, JLM
    JOURNAL OF CHROMATOGRAPHY A, 2006, 1111 (01) : 97 - 105
  • [48] LOW-PRESSURE INDUCTIVELY-COUPLED PLASMA SOURCE FOR MASS-SPECTROMETRY
    EVANS, EH
    CARUSO, JA
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (03) : 427 - 431
  • [49] Radical density measurement at low-pressure discharge denitrification by appearance mass spectrometry
    Ito, K
    Hagiwara, K
    Nakaura, H
    Tanaka, H
    Onda, K
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2001, 40 (3A): : 1472 - 1476
  • [50] Applications of multidimensional gas chromatography mass spectrometry and gas chromatography Fourier transform infrared mass spectrometry
    Tomlinson, MJ
    Sasaki, TA
    Wilkins, CL
    MASS SPECTROMETRY REVIEWS, 1996, 15 (01) : 1 - 14